Niraj Ghanwate1, Prashant Thakare2, P R Bhise3, Sonali Gawande1.
Abstract
Rapid and accurate detection of Methicillin Resistant Staphylococcus aureus (MRSA) is an important role of clinical microbiology laboratories to avoid treatment failure. The detection of MRSA is based on phenotypic assays which require at least 24 h to perform. Detection of the mecA gene or of PBP 2a is the "gold standard", but not always available. The aim of this study was to evaluate a rapid method for detection of MRSA by using 3 (4, 5 dimethyl thiazole -2-yl) -2, 5 diphenyl tetrazolium bromide (MTT). Total 126 isolates of MRSA were collected from tertiary healthcare center and were confirmed by oxacillin screening agar test as per CLSI guidelines. Amplification of mecA gene was performed by using PCR. MTT assay was carried out for all the isolates in 96 well Microtitre plate and compared with standard methods of CLSI. Out of 126 isolates, 98 were found to be mecA positive. MTT method was found to be 98.98% sensitive and 96.43% specific. The MTT based colorimetric method is rapid and simple test for screening of oxacillin resistance in Staphylococcus aureus. It significantly shortens the time to just 7 h required to obtained a drug susceptibility test and could be useful to screen MRSA.Entities:
Mesh:
Substances:
Year: 2016 PMID: 26960268 PMCID: PMC4785335 DOI: 10.1038/srep23013
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1Photoplate showing MIC breakpoints with no colour formation.
Comparison of oxacillin MICs by Agar dilution, Easy MIC Test and MTT based method along with errors and agreement with the reference method.
| No. of strains with similar readings. (Total-126) | Easy MIC Test | Agar dilution Method (μg/ml) | MTT based Method (μg/ml) | Errors | Percent agreement | |
|---|---|---|---|---|---|---|
| Minor | Major | |||||
| 21 | 4 | 4 | 4 | 00 | 00 | 100 |
| 29 | 6 | 6 | 6 | 00 | 00 | 100 |
| 03 | 4 | 6 | 4 | 00 | 00 | 100 |
| 16 | 8 | 8 | 10 | >2(0.25) | 00 | 95.83 |
| 31 | 12 | 12 | 12 | 00 | 00 | 100 |
| 07 | 12 | 16 | 14 | >2(0.25) | 00 | 95.83 |
| 12 | 16 | 12 | 14 | >2(0.25) | 00 | 95.83 |
| 03 | 6 | 6 | 4 | >2(0.25) | 00 | 95.83 |
| 04 | 6 | 4 | 6 | 00 | 00 | 100 |
| 1.85% | 00% | 98.14% | ||||
*Reference method.
Evaluation of different methods for determination of oxacillin resistance in S. aureus with their sensitivity and specificity.
| Test | Staphylococcus aureus (126) | Sensitivity % | Specificity % | Positive Likelihood Ratio | Negative Likelihood Ratio | PPV | NPV | |||
|---|---|---|---|---|---|---|---|---|---|---|
| mecA +ve (98) (MRSA) | mecA −ve (28) (MSSA) | |||||||||
| True positive | False negative | True negative | False positive | |||||||
| Oxacillin Screen Agar | 96 | 2 | 21 | 7 | 97.96 | 75 | 3.92 | 0.03 | 93.20 | 91.30 |
| Easy MIC test | 96 | 2 | 25 | 3 | 97.96 | 89.29 | 9.14 | 0.02 | 96.97 | 92.59 |
| Oxacillin Disc Diffusion (1 ug) | 92 | 6 | 7 | 21 | 93.88 | 25.00 | 1.25 | 0.24 | 81.42 | 53.85 |
| Agar Dilution | 95 | 3 | 24 | 4 | 96.94 | 85.71 | 6.79 | 0.04 | 95.96 | 88.89 |
*Reference method.
#Positive Predictive Value.
+Negative Predictive Value.
Technical Comparison between different methods and MTT based method.
| Method | Media requirement per test | Time required | MIC determination | Complexity |
|---|---|---|---|---|
| 300 ml | 24 h | Yes | More tedious with higher media requirement | |
| 20 ml | 24 h | Yes | Required Easy MIC strips which increased the cost | |
| Yes | ||||
| 20 ml | 24 h | No | Cannot detect borderline resistant strains | |
| 20 ml | 24 h | No | Gives false positive results | |
| – | 12–16 h | No | Set up not available in all laboratories |