Literature DB >> 26940470

Modular to Monoblock: Difficulties of Detaching the M(2)a-Magnum(TM) Head Are Common in Metal-on-metal Revisions.

Heikki Mäntymäki1,2, Keijo T Mäkelä3, Tero Vahlberg4, Joni Hirviniemi5, Tuukka Niinimäki6.   

Abstract

BACKGROUND: Modern hip implants typically feature modular heads, which allow for easy exchange and removal from the femoral stem at the time of revision. However, owing to fretting, corrosion, or cold welding, the modular head may be difficult or impossible to separate from the underlying trunnion, especially if the implant has titanium interfaces between the head and the stem. We have repeatedly encountered difficulty removing the titanium sleeve adapter in the M(2)a-Magnum(TM) implant. Although the manufacturer warns about this complication and cases with these difficulties have been reported to the United States FDA, we believed this topic is important to study, because the frequency of difficulties in head removal is unknown and the complications related to this event have not been characterized. QUESTIONS/PURPOSES: We asked: (1) Do revisions of M(2)a-Magnum(TM) implants differ from those of M(2)a-38(TM) implants in terms of ease of removal of the femoral head? (2) In cases where difficulty with M(2)a-Magnum(TM) head removal occurred, was the operative time, bleeding, risk of periprosthetic fracture, or joint infection increased compared with cases where the M(2)a-Magnum(TM) head was removed without difficulties?
METHODS: Between 2004 and 2014, we revised 296 THAs with metal-on-metal implants that involved M(2)a-Magnum(TM) (123) or M(2)a-38(TM) heads (88); of those, 84 were planned to include a femoral stem revision and insufficient data were available for three operations, so they were excluded from this analysis, leaving 124 THAs in the current retrospective study (70 THAs with M(2)a-Magnum(TM) and 54 THAs with M(2)a-38(TM) heads).The method of modular head removal, any difficulties removing the femoral head from the trunnion, operation time, and complications were recorded based on chart review.
RESULTS: All the observed problems of detaching the head or taper adapter were among M(2)a-Magnum(TM) heads; there were no problems detaching the head in revisions of the M(2)a-38(TM) implant. In 29% (20 of 70) of revisions of the M(2)a-Magnum(TM) implant, the modular head could not be detached by knocking it with a punch and a mallet. Seventeen percent (12 of 70) of hips needed an unplanned stem revision owing to difficulties with head removal. In revisions of the M(2)a-Magnum(TM) implant that experienced head-removal problems, the median operative time was longer (144 minutes; range, 75-274 minutes) and bleeding was greater (725 mL; range, 300-2200 mL) compared with revisions of the M(2)a-Magnum(TM) implant without head removal problems (77 minutes, range, 33-197 minutes, p < 0.001; 475 mL, range, 50-1500 mL, p = 0.004). With the numbers available, we did not see differences in terms of the proportion of patients experiencing major complications (periprosthetic fracture or postoperative infections) between the groups (difficult versus easy; 25% [five of 20] versus 8% [four of 50]; odds ratio, 3.8 [95% CI, 0.9-16.2], p = 0.067).
CONCLUSIONS: The titanium-titanium taper junction can be very difficult to separate during revision THAs, and if not anticipated, this problem can result in larger and more complicated revision procedures in patients who have the M(2)a-Magnum(TM) implant. Although the global use of metal-on-metal implants in THAs has decreased dramatically during the last several years, many thousands remain in service and therefore still might require revision. It is crucial to be prepared with special tools, including a femoral head extraction tool and diamond saw. The patient has to be informed of the possibility of a more extensive operation than preoperatively planned. LEVEL OF EVIDENCE: Level III, therapeutic study.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26940470      PMCID: PMC4965364          DOI: 10.1007/s11999-016-4774-7

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  7 in total

1.  A multinational assessment of metal-on-metal bearings in hip replacement.

Authors:  Stephen E Graves; Alastair Rothwell; Keith Tucker; Joshua J Jacobs; Art Sedrakyan
Journal:  J Bone Joint Surg Am       Date:  2011-12-21       Impact factor: 5.284

2.  Ongoing problems with metal-on-metal hip implants.

Authors:  Carl Heneghan; David Langton; M Thompson
Journal:  BMJ       Date:  2012-02-28

3.  Revisions of monoblock metal-on-metal THAs have high early complication rates.

Authors:  Louis S Stryker; Susan M Odum; Thomas K Fehring; Bryan D Springer
Journal:  Clin Orthop Relat Res       Date:  2015-02       Impact factor: 4.176

4.  Accelerating failure rate of the ASR total hip replacement.

Authors:  D J Langton; S S Jameson; T J Joyce; J N Gandhi; R Sidaginamale; P Mereddy; J Lord; A V F Nargol
Journal:  J Bone Joint Surg Br       Date:  2011-08

5.  The use of ceramic-on-ceramic bearings in isolated revision of the acetabular component.

Authors:  C M Jack; D O Molloy; W L Walter; B A Zicat; W K Walter
Journal:  Bone Joint J       Date:  2013-03       Impact factor: 5.082

6.  High rate of infection after aseptic revision of failed metal-on-metal total hip arthroplasty.

Authors:  Cody C Wyles; Robert E Van Demark; Rafael J Sierra; Robert T Trousdale
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

7.  Adverse reaction to metal debris after ReCap-M2A-Magnum large-diameter-head metal-on-metal total hip arthroplasty.

Authors:  Jari Mokka; Mika Junnila; Matti Seppänen; Petri Virolainen; Tuukka Pölönen; Tero Vahlberg; Kimmo Mattila; Esa K J Tuominen; Juho Rantakokko; Ville Aärimaa; Juha Kukkonen; Keijo T Mäkelä
Journal:  Acta Orthop       Date:  2013-10-31       Impact factor: 3.717

  7 in total