Literature DB >> 21768621

Accelerating failure rate of the ASR total hip replacement.

D J Langton1, S S Jameson, T J Joyce, J N Gandhi, R Sidaginamale, P Mereddy, J Lord, A V F Nargol.   

Abstract

There is widespread concern regarding the incidence of adverse soft-tissue reactions after metal-on-metal (MoM) hip replacement. Recent National Joint Registry data have shown clear differences in the rates of failure of different designs of hip resurfacing. Our aim was to update the failure rates related to metal debris for the Articular Surface Replacement (ASR). A total of 505 of these were implanted. Kaplan-Meier analysis showed a failure rate of 25% at six years for the ASR resurfacing and of 48.8% for the ASR total hip replacement (THR). Of 257 patients with a minimum follow-up of two years, 67 (26.1%) had a serum cobalt concentration which was greater than 7 μg/l. Co-ordinate measuring machine analysis of revised components showed that all patients suffering adverse tissue reactions in the resurfacing group had abnormal wear of the bearing surfaces. Six THR patients had relatively low rates of articular wear, but were found to have considerable damage at the trunion-taper interface. Our results suggest that wear at the modular junction is an important factor in the development of adverse tissue reactions after implantation of a large-diameter MoM THR.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21768621     DOI: 10.1302/0301-620X.93B8.26040

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  106 in total

1.  Large-diameter metal-on-metal total hip arthroplasty: dislocation infrequent but survivorship poor.

Authors:  Adolph V Lombardi; Keith R Berend; Michael J Morris; Joanne B Adams; Michael A Sneller
Journal:  Clin Orthop Relat Res       Date:  2014-11-04       Impact factor: 4.176

2.  Early Lessons From a Worldwide, Multicenter, Followup Study of the Recalled Articular Surface Replacement Hip System.

Authors:  Rami Madanat; Daniel K Hussey; Gabrielle S Donahue; Hollis G Potter; Robert Wallace; Charles Bragdon; Orhun Muratoglu; Henrik Malchau
Journal:  Clin Orthop Relat Res       Date:  2015-08-27       Impact factor: 4.176

3.  Is increased modularity associated with increased fretting and corrosion damage in metal-on-metal total hip arthroplasty devices?: a retrieval study.

Authors:  Genymphas B Higgs; Josa A Hanzlik; Daniel W MacDonald; Jeremy L Gilbert; Clare M Rimnac; Steven M Kurtz
Journal:  J Arthroplasty       Date:  2013-08-01       Impact factor: 4.757

4.  Stability and trunnion wear potential in large-diameter metal-on-metal total hips: a finite element analysis.

Authors:  Jacob M Elkins; John J Callaghan; Thomas D Brown
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

Review 5.  Metallic debris from metal-on-metal total hip arthroplasty regulates periprosthetic tissues.

Authors:  Christoph H Lohmann; Gurpal Singh; Hans-Georg Willert; Gottfried H Buchhorn
Journal:  World J Orthop       Date:  2014-11-18

6.  Editorial: arthroplasty devices: registries and beyond.

Authors:  Yelizaveta Torosyan; Steven M Kurtz; William M Mihalko; Danica Marinac-Dabic; Clare M Rimnac
Journal:  Clin Orthop Relat Res       Date:  2014-11-25       Impact factor: 4.176

7.  Editorial Comment: Symposium: Advances in Polyethylene Biomaterials.

Authors:  Steven M Kurtz; Pierangiola Bracco
Journal:  Clin Orthop Relat Res       Date:  2016-11-29       Impact factor: 4.176

8.  Medium term review of the ASR implant system: A single surgeon series.

Authors:  M Curtin; E Murphy; C Bryan; D Jadaan; M Jadaan; D Bergin; C G Murphy; W Curtin
Journal:  J Orthop       Date:  2017-02-03

9.  Revision of ASR hip arthroplasty: analysis of two hundred and ninety six recalled patients at seven years.

Authors:  Giovanni Benelli; Merildo Maritato; Pierpaolo Cerulli Mariani; Francesco Sasso
Journal:  Int Orthop       Date:  2018-09-07       Impact factor: 3.075

10.  Risk of impingement and third-body abrasion with 28-mm metal-on-metal bearings.

Authors:  Ian C Clarke; Jean-Yves Lazennec; Adrien Brusson; Christina Savisaar; John G Bowsher; Michelle Burgett; Thomas K Donaldson
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.