| Literature DB >> 26937098 |
Ashley Chin1, Jeffrey S Healey2, Carlos S Ribas2, Girish M Nair3.
Abstract
Defibrillation testing is no longer routinely performed after automatic implantable cardioverter-defibrillator (AICD) implantation. However, certain subjects undergoing AICD implantation may be at higher risk of undersensing of ventricular arrhythmias resulting in potentially fatal outcomes. We present the case of a 30-year-old woman with hypertrophic cardiomyopathy (HCM; 'asymmetric septal hypertophy' morphologic variant) and prophylactic AICD who experienced an out of hospital cardiac arrest. AICD interrogation revealed undersensing as a result of intermittent high amplitude electrograms during an episode of ventricular fibrillation (VF). The subject underwent replacement and repositioning of the AICD lead along with pulse generator replacement (that utilized a different VF sensing algorithm) with appropriate sensing of VF and successful defibrillation testing. The presence of intermittent high amplitude electrograms during episodes of VF in AICDs using the AGC function should be recognized as a situation that may necessitate interventions to prevent undersensing and consequent delay in therapy.Entities:
Keywords: Automatic implantable cardioverter defibrillator; Cardiac arrest; Hypertrophic cardiomyopathy; Undersensing of ventricular fibrillation
Year: 2015 PMID: 26937098 PMCID: PMC4750165 DOI: 10.1016/j.ipej.2015.07.009
Source DB: PubMed Journal: Indian Pacing Electrophysiol J ISSN: 0972-6292
Fig. 1The device tracing shows 3 recording channels (atrial channel (A), near field ventricular channel (V), and a far-field ventricular channel (shock)). The atrial channel shows sinus arrest with 1 atrial sensed (AS) beat. Close inspection of the near-field and far-field ventricular channels shows VF undersensing. Although the large amplitude ventricular electrograms (at times as high as 14 mV) are sensed appropriately (annotated PVC, VT and occasionally VF), the smaller amplitude ventricular electrograms following the large electrograms are undersensed. These undersensed small ventricular signals frequently correlate with ventricular electrogram signals in the far-field shock-channel confirming their presence. After a sufficient number of sensed beats fell in the “VT” zone (<320 ms), the device administered “VT” therapy.