| Literature DB >> 34386143 |
Adam Lee1, James W Salazar2, Zian H Tseng1.
Abstract
Intracardiac device electrograms demonstrating differing VF sensing performance between two manufacturer ICD generators.Entities:
Year: 2021 PMID: 34386143 PMCID: PMC8339093 DOI: 10.1002/joa3.12554
Source DB: PubMed Journal: J Arrhythm ISSN: 1880-4276
FIGURE 1Intracardiac electrograms during defibrillation threshold testing (Boston Scientific Generator)
FIGURE 2Intracardiac electrograms during defibrillation threshold testing (Medtronic Generator)
Hardware, programming, and outcomes for the four defibrillation threshold test events
| ICD Generator | Attempt no. | Sensitivity (mV) | Detection | VF Detection |
|---|---|---|---|---|
| Boston Scientific Dynagen | 1 | 1.5 (least) | 1 second (nominal) | Failed |
| 2 | 1.0 | 1 second (nominal) | Successful on redetection | |
| Medtronic Evera | 1 | 1.2 (least) | 30/40 (prolonged) | Successful |
| 2 | 0.3 (nominal) | 30/40 (prolonged) | Successful |
Charge diversion occurred on initial detection because of gross VF undersensing that resulted in a significant delay to therapy delivery.