Literature DB >> 21091965

Defibrillation thresholds in hypertrophic cardiomyopathy.

Ernest M Quin1, Frank A Cuoco, Matthew S Forcina, Jason B Coker, Robert H Yoe, William H Spencer, Valerian L Fernandes, Christopher D Nielsen, J Lacy Sturdivant, Robert B Leman, J Marcus Wharton, Michael R Gold.   

Abstract

BACKGROUND: Defibrillation threshold (DFT) testing is performed in part to ensure an adequate safety margin for the termination of spontaneous ventricular arrhythmias. Left ventricular mass is a predictor of high DFTs, so patients with hypertrophic cardiomyopathy (HCM) are often considered to be at risk for increased defibrillation energy requirements. However, there are little prospective data addressing this issue.
OBJECTIVE: To assess DFTs in patients with HCM and evaluate the clinical predictors of elevated DFTs.
METHODS: Eighty-nine consecutive patients with HCM and 600 control patients with ischemic or nonischemic cardiomyopathy underwent a uniform modified step-down DFT testing protocol. DFT was compared between the control and HCM populations. Predictors of elevated DFT were evaluated in the HCM group.
RESULTS: There was no difference in DFT between HCM and control groups (10.4 ± 5.8 J vs 11.2 ± 5.6 J, respectively). Among patients with HCM, clinical parameters such as left ventricular ejection fraction, interventricular septal thickness, left ventricular mass, and QRS duration were not predictive of an elevated DFT. Only 3 patients (3.4%) with HCM had a DFT >20 J.
CONCLUSION: Patients with HCM do not have elevated DFTs as compared to more typical populations undergoing implantable cardioverter-defibrillator implant; high-energy devices or complex lead systems are not needed routinely in this population.
© 2010 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2010        PMID: 21091965     DOI: 10.1111/j.1540-8167.2010.01943.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  2 in total

Review 1.  Implantable cardioverter-defibrillator in hypertrophic cardiomyopathy.

Authors:  Diego Jimenez Sanchez; Ignacio Fernández Lozano
Journal:  Glob Cardiol Sci Pract       Date:  2018-08-12

2.  Delayed AICD therapy and cardiac arrest resulting from undersensing of ventricular fibrillation in a subject with hypertrophic cardiomyopathy-A case report.

Authors:  Ashley Chin; Jeffrey S Healey; Carlos S Ribas; Girish M Nair
Journal:  Indian Pacing Electrophysiol J       Date:  2015-07-29