Interventional procedures, cine acquisitions and operation of fluoroscopic equipment in high-dose fluoroscopic modes, involve long fluoroscopic times which can lead to high staff doses. Also, Coronary angiography (CA) procedures require the cardiologist and assisting personnel to remain close to the patient, which is the main source of scattered radiation. Thus, radiation exposure is a significant concern for radiation workers and it is important to measure the radiation doses received by personnel and evaluate the parameters concerning total radiation burden. In this research, we investigated radiation doses to 10 cardiologists performing 120 CA procedures. Using thermo luminescent dosimeters doses to the wrists, thyroid and eyes per procedure were measured. Based on the measured dose values, maximum doses to the Left wrist, Right wrist, thyroid and eyes of cardiologist were measured 241.45 µSv, 203.17 µSv, 78.21 µSv and 44.58 µSv, respectively. The results of this study indicate that distance from the source, use of protective equipment's, procedure complexity, equipment performance, and cardiologist experience are the principal exposure-determining variables. It can be conclude that if adequate radiation protection approaches have been implemented, occupational dose levels to cardiologists would be within the regulated acceptable dose limits.Entities:
Mesh:
Year: 2015 PMID: 26925906 PMCID: PMC4965688 DOI: 10.5539/gjhs.v8n7p185
Source DB: PubMed Journal: Glob J Health Sci ISSN: 1916-9736
Imaging system features
| Field size(cm2) | 13, 17, 21, 23 (cm2) |
|---|---|
| Frame Rate(F/Sec) | 10,15 (F/Sec) |
| Fluoro(kVp) | 30-80 (kVp) |
| Fluoro (mA) | 4-15 (mA) |
| Cine(kVp) | 120 (kVp) |
| Cine(mA) | 30-170 (mA) |
Field size= radiation field.
Dosimetric parameters for monitored procedures
| Physician | Number of procedure | Mean fluoroscopy time ± SD (min) | Mean DAP ±SD (µGy.m2) |
|---|---|---|---|
| C1 | 12 | 5.57±1.19 | 2085.56±1210.05 |
| C2 | 12 | 2.74±1.97 | 2055.07±921.50 |
| C3 | 12 | 3.22±3.25 | 2389.24±1336.76 |
| C4 | 12 | 2.75±1.03 | 2349.53±801.97 |
| C5 | 12 | 3.08±2.81 | 2469.08±739.74 |
| C6 | 12 | 3.26±1.93 | 2856.83±1646.26 |
| C7 | 12 | 3.24±2.31 | 2377.75±1147.88 |
| C8 | 12 | 3.32±2.07 | 2791.96±1403.13 |
| C9 | 12 | 4.54±5.01 | 2953.91±1540.0 |
| C10 | 12 | 4.30±506 | 2638.4±1544.14 |
Figure 1The average dose in each right and left wrists of 10 cardiologists (C1-C10) from interventional cardiology examinations
Figure 2The measured thyroid and eyes dose of 10 cardiologists (C1-C10) per procedure from interventional cardiology examinations
Summary of cardiologists’ data for CA procedures in the present study
| Organ | Mean | S.D | Med | Min | Max |
|---|---|---|---|---|---|
| Left wrist | 130.9 | 88.83 | 106.94 | 30.55 | 376.01 |
| Right wrist | 93.65 | 82.01 | 57.21 | 19.72 | 385.91 |
| Thyroid | 40.05 | 21.65 | 32.36 | 11.04 | 101.00 |
| Eye | 30.15 | 15.01 | 26.67 | 10.16 | 95.01 |
Thyroid and eye dose reduction (%)
| Thyroid dose reduction (%) | Eye dose reduction (%) | |
|---|---|---|
| C1 | 36 | 34 |
| C2 | 35 | 35 |
| C3 | 27 | 27 |
| C4 | 38 | 34 |
| C5 | 33 | 26 |
| C6 | 39 | 37 |
| C7 | 53 | 38 |
| C8 | 28 | 25 |
| C9 | 41 | 24 |
| C10 | 41 | 30 |
Some of the factors affecting cardiologists dose in CA.
| Category | |
|---|---|
| Clinical problem, lesion characteristic, body habitus | |
| Body size and height, experience and skill, work load, procedural technique, position with respect to patient | |
| Equipment- mounted shielding and personal protective shields | |
Cardiologists demographic data
| Sex | Age (Year) | Height (Cm) | Weight (Kg) | Experience (Year) | Specific disease | |
|---|---|---|---|---|---|---|
| C1 | M | 42 | 181 | 103 | 5 | No |
| C2 | M | 39 | 179 | 83 | 4 | No |
| C3 | F | 45 | 173 | 95 | 4 | No |
| C4 | F | 49 | 159 | 72 | 4 | No |
| C5 | M | 30 | 165 | 79 | 3 | No |
| C6 | M | 37 | 172 | 87 | 2 | No |
| C7 | M | 36 | 184 | 68 | 3 | No |
| C8 | F | 39 | 176 | 70 | 2 | No |
| C9 | F | 35 | 165 | 91 | 2 | No |
| C10 | M | 33 | 157 | 84 | 3 | No |
F: Female; M: Male.
The data on the organ dose (in µSv) of radiation workers for interventional cardiology in three studies
| Organ | Present study (CA) | ||
|---|---|---|---|
| Left wrist | 63.61-241.45 | 20-3684 | 31.8-358.7 |
| Right wrist | 30.33-203.17 | 14-138 | 12.0-124.0 |
| Thyroid | 23.94-78.21 | - | 1.2-100.3 |
| Eyes | 16.37-44.58 | 0-61 | - |
Patient demographic data
| Height[cm] | weight[kg] | age[years] | |
|---|---|---|---|
| Average | 157.6 | 76.90 | 58 |
| Maximum | 178 | 118 | 69 |
| Minimum | 132 | 54 | 47 |
Comparison of this study with literature in CA
| Mean DAP value [Gy.cm2] | |
|---|---|
| 66.5 | |
| 55.9 | |
| 60.6 | |
| 49.0 | |
| This study | 25.23 |