Moon Woo Seong1, Seung Jun Lee1, Sung Im Cho1, Kyungphil Ko1, Mi Na Kim2, Heungsub Sung2, Jae Seok Kim3, Ji Soo Ahn1, Byung Su Yu1, Taek Soo Kim1, Eui Chong Kim1, Sung Sup Park4.
Abstract
BACKGROUND: The largest outbreak of Middle East respiratory syndrome coronavirus (MERS-CoV) infection outside Middle East Asia in 2015 has necessitated the rapid expansion of laboratories that conduct MERS-CoV molecular testing in Korea, together with external quality assessment (EQA) to evaluate the assays used.Entities:
Keywords: External quality assessment; MERS-CoV; Molecular diagnostics; ORF1a; Real-time PCR; upE
Mesh:
Substances:
Year: 2016 PMID: 26915611 PMCID: PMC4773263 DOI: 10.3343/alm.2016.36.3.230
Source DB: PubMed Journal: Ann Lab Med ISSN: 2234-3806 Impact factor: 3.464
Summary of results from the first phase of EQA
| Sample (copies/µL) | Kit | CT (Mean±SD) | N of participants* | |
|---|---|---|---|---|
| upE | ORF1a | |||
| 1,000 | AccuPower | 31.7 ± 1.5 | 32.3 ± 0.6 | 4 |
| PowerChek | 28.6 ± 1.1 | 28.6 ± 1.1 | 24 | |
| DiaPlexQ | 29.6 ± 0.5 | 30.0 ± 1.2 | 5 | |
| Others† | 30.1 ± 1.0 | 29.5 ± 0.9 | 4 | |
| Total | 29.3 ± 1.5 | 29.3 ± 1.6 | 37 | |
| 100 | AccuPower | 35.9 ± 1.0 | 35.9 ± 1.0 | 3 |
| PowerChek | 31.8 ± 1.0 | 32.2 ± 1.2 | 24 | |
| DiaPlexQ | 33.1 ± 0.9 | 32.5 ± 1.1 | 5 | |
| Others‡ | 33.9 ± 1.4 | 33.5 ± 1.9 | 2 | |
| Total | 32.5 ± 1.6 | 32.7 ± 1.6 | 34 | |
| 10 | AccuPower | NA | NA | 0 |
| PowerChek | 35.3 ± 1.7 | 35.9 ± 1.9 | 13 | |
| DiaPlexQ | 38.0 ± 0.0 | 35.0 ± 0.0 | 1 | |
| Others‡ | NA | 36.5 ± 2.0 | 2 | |
| Total | 35.7 ± 1.8 | 36.0 ± 1.8 | 16 | |
| 1 | AccuPower | NA | NA | 0 |
| PowerChek | 37.0 ± 1.5 | 38.1 ± 0.9 | 6 | |
| DiaPlexQ | NA | NA | 0 | |
| Others‡ | 37.8 ± 1.5 | 37.5 ± 2.4 | 2 | |
| Total | 37.3 ± 1.4 | 37.9 ± 1.3 | 8 | |
*The number of participants was different for each concentration because the first EQA phase was based on self-validation; †This includes two laboratory-developed tests, MERS-CoV Detection kit (Nanobiosys, Seoul, Korea) and ModularDx kit MERS-CoV (TIB MOLBIOL, Berlin, Germany); ‡This includes two laboratory-developed tests.
Abbreviations: EQA, external quality assessment; CT, threshold cycle; NA, not applicable.
Summary of results from the second phase of EQA
| Sample | Correct results | Kit | CT (Mean±SD) | N of participants* | |
|---|---|---|---|---|---|
| upE | ORF1a | ||||
| A | 100% | AccuPower | 28.1 ± 0.6 | 29.8 ± 0.2 | 4 |
| PowerChek | 25.8 ± 0.9 | 26.5 ± 0.7 | 31 | ||
| DiaPlexQ | 28.0 ± 1.9 | 26.8 ± 2.2 | 9 | ||
| Others* | 26.5 ± 1.3 | 27.1 ± 1.4 | 3 | ||
| Total | 26.6 ± 1.6 | 27.0 ± 1.6 | 47 | ||
| B | 100% | AccuPower | 31.6 ± 0.7 | 33.3 ± 0.2 | 4 |
| PowerChek | 29.5 ± 1.0 | 30.2 ± 0.9 | 31 | ||
| DiaPlexQ | 30.5 ± 1.8 | 30.6 ± 2.4 | 9 | ||
| Others* | 30.1 ± 1.3 | 30.7 ± 1.4 | 3 | ||
| Total | 30.2 ± 1.5 | 30.7 ± 1.6 | 47 | ||
| C | 100% | AccuPower | Negative | Negative | 4 |
| PowerChek | Negative | Negative | 31 | ||
| DiaPlexQ | Negative | Negative | 9 | ||
| Others* | Negative | Negative | 3 | ||
| Total | Negative | Negative | 47 | ||
*This includes laboratory-developed test, MERS-CoV Detection kit (Nanobiosys, Seoul, Korea) and ModularDx kit MERS-CoV (TIB MOLBIOL, Berlin, Germany).
Abbreviations: EQA, external quality assessment; CT, threshold cycle.
Fig. 1Boxplot of threshold cycle (CT) values of external quality assessment (EQA) materials by target genes. The central line shows the median and the box covers the interquartile range. The four clusters for the upE gene and ORF1a gene represent 1,000, 100, 10, and 1 copies/µL, respectively. Each cluster denotes results from all methods, PowerChek kit (KogeneBiotech), and DiaPlexQ kit (Solgent), respectively, from left to right.
Comparison of CT values between two real-time PCR platforms using PowerChek kit
| EQA phase | Materials | Target | CT (Mean±SD) | ||||
|---|---|---|---|---|---|---|---|
| ABI | N | BioRad | N | ||||
| 1st | 1,000 | upE | 28.2 ± 1.3 | 10 | 29.0±0.8 | 13 | NS |
| ORF1a | 28.3 ± 1.1 | 8 | 28.8±1.1 | 11 | NS | ||
| 100 | upE | 31.5 ± 1.2 | 10 | 32.0±1.0 | 13 | NS | |
| ORF1a | 31.7 ± 0.7 | 10 | 32.4 ±1.4 | 12 | NS | ||
| 10 | upE | 34.5 ± 1.2 | 5 | 36.1±1.2 | 7 | NS | |
| ORF1a | 35.4 ± 1.2 | 5 | 36.9±2.2 | 6 | NS | ||
| 2nd | A | upE | 25.5 ± 0.7 | 14 | 26.3±0.6 | 15 | 0.0011 |
| ORF1a | 26.4 ± 0.7 | 14 | 26.8±0.7 | 15 | 0.0447 | ||
| B | upE | 29.2 ± 0.7 | 14 | 30.0±0.7 | 15 | 0.0035 | |
| ORF1a | 30.0 ± 0.9 | 14 | 30.5±0.8 | 15 | NS | ||
*The comparison of CT values between the two groups was performed by using the Mann-Whitney test because sample sizes of some groups were small.
Abbrviations: EQA, external quality assessment; CT, threshold cycle; NS, not significant.