Literature DB >> 26888222

Movement control during aspiration with different injection systems via video monitoring-an in vitro model.

P W Kämmerer1, D Schneider2, A A Pacyna3, M Daubländer4.   

Abstract

OBJECTIVES: The aim of the present study was an evaluation of movement during double aspiration by different manual syringes and one computer-controlled local anesthesia delivery system (C-CLAD).
MATERIALS AND METHODS: With five different devices (two disposable syringes (2, 5 ml), two aspirating syringes (active, passive), one C-CLAD), simulation of double aspiration in a phantom model was conducted. Two experienced and two inexperienced test persons carried out double aspiration with the injection systems at the right and left phantom mandibles in three different inclination angles (n = 24 × 5 × 2 for each system). 3D divergences of the needle between aspiration procedures (mm) were measured with two video cameras.
RESULTS: An average movement for the 2-ml disposal syringe of 2.85 mm (SD 1.63), for the 5 ml syringe of 2.36 mm (SD 0.86), for the active-aspirating syringe of 2.45 mm (SD 0.9), for the passive-aspirating syringe of 2.01 mm (SD 0.7), and for the C-CLAD, an average movement of 0.91 mm (SD 0.63) was seen. The movement was significantly less for the C-CLAD compared to the other systems (p < 0.001). The movement of the needle in the soft tissue was significantly less for the C-CLAD compared to the other systems (p < 0.001).
CONCLUSIONS: A difference in involuntary movement of the syringe could be seen in comparison between manual and C-CLAD systems. Launching the aspiration by a foot pedal in computer-assisted anesthesia leads to a minor movement. CLINICAL RELEVANCE: To solve the problem of movement during aspiration with possibly increased false-negative results, a C-CLAD seems to be favorable.

Entities:  

Keywords:  Aspiration; C-CLAD; Deviation; Inferior alveolar nerve block; Local anesthesia

Mesh:

Year:  2016        PMID: 26888222     DOI: 10.1007/s00784-016-1756-9

Source DB:  PubMed          Journal:  Clin Oral Investig        ISSN: 1432-6981            Impact factor:   3.573


  24 in total

Review 1.  An evaluation of the closed mouth mandibular block technique.

Authors:  P Donkor; J Wong; A Punnia-Moorthy
Journal:  Int J Oral Maxillofac Surg       Date:  1990-08       Impact factor: 2.789

2.  Interpretation of aspiration tests in local anesthetic injections.

Authors:  J E Watson; R S Colman
Journal:  J Oral Surg       Date:  1976-12

3.  Reasons for local anesthesia failures.

Authors:  M K Wong; P L Jacobsen
Journal:  J Am Dent Assoc       Date:  1992-01       Impact factor: 3.634

4.  Clinical use of an epinephrine-reduced (1/400,000) articaine solution in short-time dental routine treatments--a multicenter study.

Authors:  Monika Daubländer; Peer W Kämmerer; Brita Willershausen; Michael Leckel; Hans-Christoph Lauer; Siegmar Buff; Benita Rösl
Journal:  Clin Oral Investig       Date:  2011-08-23       Impact factor: 3.573

5.  Inferior dental nerve block. Routine aspiration and a modified technique.

Authors:  J P Rood
Journal:  Br Dent J       Date:  1972-02-01       Impact factor: 1.626

6.  Frequency of accidental intravascular injection of local anaesthetics in children.

Authors:  P T Bishop
Journal:  Br Dent J       Date:  1983-02-05       Impact factor: 1.626

7.  Comparative study of different syringes in positive aspiration during inferior alveolar nerve block.

Authors:  E Delgado-Molina; S Bueno-Lafuente; L Berini-Aytés; C Gay-Escoda
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  1999-11

8.  Comparison of 4% articaine with epinephrine (1:100,000) and without epinephrine in inferior alveolar block for tooth extraction: double-blind randomized clinical trial of anesthetic efficacy.

Authors:  Peer W Kämmerer; Victor Palarie; Monika Daubländer; Constantin Bicer; Niloufar Shabazfar; Dan Brüllmann; Bilal Al-Nawas
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol       Date:  2011-08-06

9.  Evaluation and comparison of 2 needle models in terms of blood aspiration during truncal block of the inferior alveolar nerve.

Authors:  Esther Delgado-Molina; Meritxell Tamarit-Borrás; Leonardo Berini-Aytés; Cosme Gay-Escoda
Journal:  J Oral Maxillofac Surg       Date:  2003-09       Impact factor: 1.895

10.  The cardiovascular effect of local anesthesia with articaine plus 1:200,000 adrenalin versus lidocaine plus 1:100,000 adrenalin in medically compromised cardiac patients: a prospective, randomized, double blinded study.

Authors:  Sharon Elad; Dan Admon; Meirav Kedmi; Eitan Naveh; Esti Benzki; Sharon Ayalon; Alex Tuchband; Haim Lutan; Eliezer Kaufman
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2008-06
View more
  1 in total

1.  Evaluation of different mandibular blocks for the removal of lower third molars: a meta-analysis of randomised clinical trials.

Authors:  Carolina-Noemi Cavallini; Jorge Toledano-Serrabona; Cosme Gay-Escoda
Journal:  Clin Oral Investig       Date:  2020-08-25       Impact factor: 3.573

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.