Normal ranges and test-retest reproducibility of flow and velocity parameters in intracranial arteries measured with phase-contrast magnetic resonance imaging.

Maria Correia de Verdier1, Johan Wikström2.   

Abstract

INTRODUCTION: The purpose of the present study was to investigate normal ranges and test-retest reproducibility of phase-contrast MRI (PC-MRI)-measured flow and velocity parameters in intracranial arteries.
METHODS: Highest flow (HF), lowest flow (LF), peak systolic velocity (PSV), and end diastolic velocity (EDV) were measured at two dates in the anterior (ACA), middle (MCA), and posterior (PCA) cerebral arteries of 30 healthy volunteers using two-dimensional PC-MRI at 3 T. Least detectable difference (LDD) was calculated.
RESULTS: In the left ACA, HF was (mean (range, LDD)) 126 ml/min (36-312, 59 %), LF 61 ml/min (0-156, 101 %), PSV 64 cm/s (32-141, 67 %), and EDV 35 cm/s (18-55, 42 %); in the right ACA, HF was 154 ml/min (42-246, 49 %), LF 77 ml/min (0-156, 131 %), PSV 75 cm/s (26-161, 82 %), and EDV 39 cm/s (7-59, 67 %). In the left MCA, HF was 235 ml/min (126-372, 35 %), LF 116 ml/min (42-186, 48 %), PSV 90 cm/s (55-183, 39 %), and EDV 46 cm/s (20-66, 28 %); in the right MCA, HF was 238 ml/min (162-342, 44 %), LF 120 ml/min (72-216, 48 %), PSV 88 cm/s (55-141, 35 %), and EDV 45 cm/s (26-67, 23 %). In the left PCA, HF was 108 ml/min (42-168, 54 %), LF 53 ml/min (18-108, 64 %), PSV 50 cm/s (24-77, 63 %), and EDV 28 cm/s (14-40, 45 %); in the right PCA, HF was 98 ml/min (30-162, 49 %), LF 49 ml/min (12-84, 55 %), PSV 47 cm/s (27-88, 59 %), and EDV 27 cm/s (16-41, 45 %).
CONCLUSION: PC-MRI-measured flow and velocity parameters in the main intracranial arteries have large normal ranges. Reproducibility is highest in MCA.

Entities:  

Keywords:  Blood flow velocity; Cerebral arteries; Hemodynamics; Magnetic resonance imaging; Phase-contrast MRI

Mesh:

Year:  2016        PMID: 26882908     DOI: 10.1007/s00234-016-1661-6

Source DB:  PubMed          Journal:  Neuroradiology        ISSN: 0028-3940            Impact factor:   2.804


  27 in total

1.  Comparison between magnetic resonance phase contrast imaging and transcranial Doppler ultrasound with regard to blood flow velocity in intracranial arteries: work in progress.

Authors:  J Seitz; M Strotzer; J Schlaier; W R Nitz; M Völk; S Feuerbach
Journal:  J Neuroimaging       Date:  2001-04       Impact factor: 2.486

Review 2.  Transcranial color-coded duplex sonography.

Authors:  R W Baumgartner
Journal:  J Neurol       Date:  1999-08       Impact factor: 4.849

3.  In vitro validation of phase-contrast flow measurements at 3 T in comparison to 1.5 T: precision, accuracy, and signal-to-noise ratios.

Authors:  Joachim Lotz; Rolf Döker; Ralph Noeske; Meike Schüttert; Roland Felix; Michael Galanski; Matthias Gutberlet; Gerd Peter Meyer
Journal:  J Magn Reson Imaging       Date:  2005-05       Impact factor: 4.813

4.  Visualization of hemodynamics in intracranial arteries using time-resolved three-dimensional phase-contrast MRI.

Authors:  Shuhei Yamashita; Haruo Isoda; Masaya Hirano; Hiroyasu Takeda; Shoichi Inagawa; Yasuo Takehara; Marcus T Alley; Michael Markl; Norbert J Pelc; Harumi Sakahara
Journal:  J Magn Reson Imaging       Date:  2007-03       Impact factor: 4.813

5.  In vivo assessment and visualization of intracranial arterial hemodynamics with flow-sensitized 4D MR imaging at 3T.

Authors:  S Wetzel; S Meckel; A Frydrychowicz; L Bonati; E-W Radue; K Scheffler; J Hennig; M Markl
Journal:  AJNR Am J Neuroradiol       Date:  2007-03       Impact factor: 3.825

6.  Quantification and visualization of flow in the Circle of Willis: time-resolved three-dimensional phase contrast MRI at 7 T compared with 3 T.

Authors:  P van Ooij; J J M Zwanenburg; F Visser; C B Majoie; E vanBavel; J Hendrikse; A J Nederveen
Journal:  Magn Reson Med       Date:  2012-05-22       Impact factor: 4.668

7.  Feasibility of 4D flow MR imaging of the brain with either Cartesian y-z radial sampling or k-t SENSE: comparison with 4D Flow MR imaging using SENSE.

Authors:  Tetsuro Sekine; Yasuo Amano; Ryo Takagi; Yoshio Matsumura; Yasuo Murai; Shinichiro Kumita
Journal:  Magn Reson Med Sci       Date:  2014-01-31       Impact factor: 2.471

8.  Intraluminal diameters of the intracranial arteries.

Authors:  B A Ring; M M Waddington
Journal:  Vasc Surg       Date:  1967-09

9.  Accuracy of phase-contrast flow measurements in the presence of partial-volume effects.

Authors:  C Tang; D D Blatter; D L Parker
Journal:  J Magn Reson Imaging       Date:  1993 Mar-Apr       Impact factor: 4.813

10.  Anatomical variations of the circulus arteriosus in cadaveric human brains.

Authors:  S A Gunnal; M S Farooqui; R N Wabale
Journal:  Neurol Res Int       Date:  2014-05-07
View more
  6 in total

1.  Response to Letter to the Editor: "Reproducibility of flow and velocity parameters in intracranial arteries measured with phase-contrast magnetic resonance imaging; a methodological issue".

Authors:  Maria Correia de Verdier; Johan Wikström
Journal:  Neuroradiology       Date:  2016-07-22       Impact factor: 2.804

2.  Reproducibility of flow and velocity parameters in intracranial arteries measured with phase-contrast magnetic resonance imaging: a methodological issue.

Authors:  Siamak Sabour
Journal:  Neuroradiology       Date:  2016-06-22       Impact factor: 2.804

3.  Evidence for non-Newtonian behavior of intracranial blood flow from Doppler ultrasonography measurements.

Authors:  Khalid M Saqr; Ossama Mansour; Simon Tupin; Tamer Hassan; Makoto Ohta
Journal:  Med Biol Eng Comput       Date:  2018-12-07       Impact factor: 2.602

4.  Vascular territorial segmentation and volumetric blood flow measurement using dynamic contrast enhanced magnetic resonance angiography of the brain.

Authors:  Oren Geri; Shelly I Shiran; Jonathan Roth; Moran Artzi; Liat Ben-Sira; Dafna Ben Bashat
Journal:  J Cereb Blood Flow Metab       Date:  2017-04-10       Impact factor: 6.200

5.  Identification of intra-individual variation in intracranial arterial flow by MRI and the effect on computed hemodynamic descriptors.

Authors:  Xinke Liu; Evan Kao; Henrik Haraldsson; Megan Ballweber; Alastair Martin; Youxiang Li; Yuting Wang; David Saloner
Journal:  MAGMA       Date:  2021-04-11       Impact factor: 2.533

6.  Accuracy of blood flow assessment in cerebral arteries with 4D flow MRI: Evaluation with three segmentation methods.

Authors:  Tora Dunås; Madelene Holmgren; Anders Wåhlin; Jan Malm; Anders Eklund
Journal:  J Magn Reson Imaging       Date:  2019-01-14       Impact factor: 4.813

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.