MR imaging of healthy knees in varying degrees of flexion using a stretchable coil array provides comparable image quality compared to a standard knee coil array.

Jurek A Nordmeyer-Massner1, Klaas P Pruessmann1, Michael Wyss1, Andrei Manoliu2, Jürg Hodler3, Gustav Andreisek3, Nadja Mamisch-Saupe3.   

Abstract

OBJECTIVE: Stretchable coils allow knee imaging at varying degrees of flexion. Purpose was to compare a new-developed stretchable 8-channel to a standard 8-channel knee coil array by means of quantitative and qualitative image analysis.
MATERIAL AND METHODS: IRB approved prospective study. Knee MR imaging in 10 healthy volunteers was performed at 3T using a standard 8-channel and a new-developed stretchable 8-channel coil array at 0°, 45°, and 60° of flexion and at 0° (standard coil). Image parameters were identical. Signal-to-noise ratio (SNR) was determined by combining the images with separately acquired noise data on a pixel-by-pixel basis using MATLAB routines (Natick, MA, USA). Images were qualitatively analysed by two independent radiologists who graded the visibility of several anatomic structures from 1=not visible to 5=excellent. ANOVA, Wilcoxon and kappa statistics were used.
RESULTS: Mean SNR±standarddeviation of bone was 54.7±10.4 and of muscle 28.0± 4.4 using the stretchable coil array and 54.6±8.2 and 33.4±4.5, respectively, using the standard knee coil array. No statistically significant SNR differences were found between both arrays (bone, p=0.960; muscle, p=0.132). SNR was not degraded at higher degrees of flexion. The qualitative image analysis did not reveal statistically significant differences between the stretchable and standard coil array with regard to the visibility of anatomic structures (p=0.026-1.000). Overall kappa was 0.714.
CONCLUSION: Stretchable 8-channel coil arrays provide similar SNR and visibility of anatomic structures compared to standard 8-channel knee coil arrays. MR imaging with high SNR will now be possible in flexed knees.
Copyright © 2015 Elsevier Ireland Ltd. All rights reserved.

Keywords:  Coil; Coil array; Joint; Kinematics; Knee; MR imaging

Mesh:

Year:  2015        PMID: 26860662     DOI: 10.1016/j.ejrad.2015.12.004

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  3 in total

1.  Magnetic resonance cinematography of the fingers: a 3.0 Tesla feasibility study with comparison of incremental and continuous dynamic protocols.

Authors:  Thomas Bayer; Werner Adler; Rolf Janka; Michael Uder; Frank Roemer
Journal:  Skeletal Radiol       Date:  2017-08-19       Impact factor: 2.199

2.  Conductive Thread-Based Stretchable and Flexible Radiofrequency Coils for Magnetic Resonance Imaging.

Authors:  Jana M Vincent; Joseph V Rispoli
Journal:  IEEE Trans Biomed Eng       Date:  2019-11-28       Impact factor: 4.538

3.  Stitching Stretchable Radiofrequency Coils for MRI: A Conductive Thread and Athletic Fabric Approach.

Authors:  Jana M Vincent; Joseph V Rispoli
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2019-07
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.