Literature DB >> 26816506

Comparison of doses to the rectum derived from treatment planning system with in-vivo dose values in vaginal vault brachytherapy using cylinder applicators.

Emmanuel Oyeyemi Oyekunle1, Rachel Ibhade Obed2, Bidemi Idayat Akinlade3, Atara Ntekim3.   

Abstract

PURPOSE: In-vivo measurements to determine doses to organs-at-risk can be an essential part of brachytherapy quality assurance (QA). This study compares calculated doses to the rectum with measured dose values as a means of QA in vaginal vault brachytherapy using cylinder applicators.
MATERIAL AND METHODS: At the Department of Radiotherapy, University College Hospital (UCH), Ibadan, Nigeria, intracavitary brachytherapy (ICBT) was delivered by a GyneSource high-dose-rate (HDR) unit with (60)Co. Standard 2D treatment plans were created with HDR basic 2.6 software for prescription doses 5-7 Gy at points 5 mm away from the posterior surface of vaginal cylinder applicators (20, 25, and 30 mm diameters). The LiF:Mg, Ti thermoluminescent dosimeter rods (1 x 6 mm) were irradiated to a dose of 7 Gy on Theratron (60)Co machine for calibration purpose prior to clinical use. Measurements in each of 34 insertions involving fourteen patients were performed with 5 TLD-100 rods placed along a re-usable rectal marker positioned in the rectum. The dosimeters were read in Harshaw 3500 TLD reader and compared with doses derived from the treatment planning system (TPS) at 1 cm away from the dose prescription points.
RESULTS: The mean calculated and measured doses ranged from 2.1-3.8 Gy and 1.2-5.6 Gy with averages of 3.0 ± 0.5 Gy and 3.1 ± 1.1 Gy, respectively, for treatment lengths 2-8 cm along the cylinder-applicators. The mean values correspond to 48.9% and 50.8% of the prescribed doses, respectively. The deviations of the mean in-vivo doses from the TPS values ranged from -1.9 to 2.1 Gy with a p-value of 0.427.
CONCLUSIONS: This study was part of efforts to verify rectal dose obtained from the TPS during vaginal vault brachytherapy. There was no significant difference in the dose to the rectum from the two methods of measurements.

Entities:  

Keywords:  brachytherapy; cervical cancer; in-vivo dose; rectum; vaginal cylinder; vaginal vault

Year:  2015        PMID: 26816506      PMCID: PMC4716134          DOI: 10.5114/jcb.2015.56767

Source DB:  PubMed          Journal:  J Contemp Brachytherapy        ISSN: 2081-2841


Purpose

In spite of the fact that intracavitary brachytherapy (ICBT) has been used in the treatment of cervical cancer for more than 20 years, the potential for increased risk of late complications, especially in the rectum is a major concern [1]. Vaginal cylinder is an alternative treatment tool to tandem-ring or tandem-ovoid applicators in cervical cancer patients who had undergone hysterectomy. Similarly, it is used to treat vaginal vault after hysterectomy following endometrial cancer. In these cases, the uterine canal is no longer available for placement of the intra-uterine tandem (IUT). As such, rather than the use of ring or ovoids applicators, which cannot adequately treat the distal part of the vaginal vault, vaginal cylindrical applicators attached to a single central channel or multiple channels are often used. Cervical cancer is the most common indication for brachytherapy in most developing countries and high-dose-rate (HDR) brachytherapy equipment are capable of treating large number of patients. Hence, HDR BT is recommended for developing countries with high incidence of the disease [2]. The HDR brachytherapy technique in Nigeria commenced in 2008 at the University College Hospital (UCH), Ibadan. At UCH, our records indicated that 85% of patients receiving ICBT use the tandem-ring applicator, while the proportion requiring vaginal cylinder is about 15%. Unlike the three-channel ring/ovoid-IUT applicators, in-vivo dosimetry is uncommon for single-channel vaginal cylinders, since dose contribution to organs-at-risk (OARs), such as the rectum, is mainly due to the cylinder. In tandem-cylinder applications, dose prescription points are usually at 5 mm away from the posterior surface of the cylinder. A study by Demanes et al. [3] showed that the multichannel vaginal cylinder allows much better dose control than the single-channel type, as the former achieved lower rectal doses by 15%, when compared to the latter. In-vivo dosimetry is a useful tool for evaluating the doses to OARs, such as the rectum, as part of quality assurance (QA) for assessing the HDR technique including the treatment planning system (TPS). Thermoluminescent dosimeters (TLDs) and semiconductor diodes have been commonly used by many researchers for in-vivo dosimetry [4, 5, 6, 7]. These dosimeters are used for point dose measurements and hence suffer from the usual drawback that the measured dose may not always represent the clinically significant dose to OARs. Therefore, measurements at several points may be needed. This study was designed to compare doses to the rectum derived from the TPS with in-vivo dose values using TLDs, as a means of QA in ICBT with cylinder applicators at the pioneer HDR brachytherapy center in Nigeria.

Material and methods

This was a prospective study carried out among cervical and endometrial cancer patients who had hysterectomy, and were booked to have HDR brachytherapy to the vaginal vault at the University College Hospital, Ibadan, Nigeria. Approval for the study was obtained from the institution's Ethical Review Committee and the patients gave consent for their participation. Vaginal vault brachytherapy was carried out using single-channel vaginal cylinders and in-vivo dosimetry was done using TLD-100 rods. Data analysis in this study was performed using GraphPad Prism 6 statistical software (San Diego, USA).

TLD-100 rods and calibration

Tissue equivalence characteristic of detector, small size, and its affordability informed the choice of TLDs among other possible detectors. The TLDs, Li:Mg, Ti rods (6 mm length, 1 mm diameter) used for this study were procured from Harshaw, Ohio, USA. The dimensions of the rod type were considered more suitable for in-vivo study as in this work. The dosimeters were annealed in a TLD oven at the Center for Energy Research and Development (CERD), Obafemi Awolowo University, Ile-Ife, Nigeria. They were initially irradiated with a dose of 1 Gy using Theratron 780C telecobalt machine (MDS Nordian, Canada, Inc.) at the Radiotherapy Center of EKO Hospital, Lagos, Nigeria. The telecobalt machine was pre-calibrated with a Farmer ionization chamber: NEL 2581, serial no. 837 (0.6 cm3) and the associated electrometer: 2570/1 Farmer, serial no. 967. The TLDs were placed on the surface of a 30 × 30 × 17.6 cm3 polymethylmethacrylate (PMMA) phantom during irradiation with a field size of 20 cm x 20 cm. The dosimeters were kept within the center of the radiation beam. A radiographic film was used to check the dose uniformity of the absorbed dose delivered. The irradiated TLDs were read using a Harshaw 3500 reader. Five calibration (golden) dosimeters with best responses (values ± 2% of 1 Gy) were determined and selected. All the TLDs were re-annealed, with the golden dosimeters separately identified. The entire dosimeters were irradiated with a dose of 7 Gy (typical of brachytherapy dose) using the same external beam radiation therapy (EBRT) unit above. The five golden TLDs were first read and used to calibrate the TLD reader. The other dosimeters were read in turn and the element correction coefficient (ECC) was generated for each according to equation (1). where TLD and (TLD) are the individual reading of TLDs and the mean value, respectively.

Measurements with TLD-100

During the period between June and August 2013, thirty-four ICBT applications using single-channel cylinder applicators were carried out. Fourteen patients with cervical and endometrial cancers who had hysterectomy at UCH were included in this study. Prior to clinical use, each dosimeter was placed in a small black polythene bag properly labelled for identification after annealing in the TLD oven. Five separate dosimeters were placed at 5 mm interval on the posterior surface of a flexible rectal marker, which is of 5 mm thickness. The rods were stabilized using tapes on the surface of the rectal marker. Each ICBT insertion involved the use of five detectors in order to obtain reasonable mean doses, and therefore increase the accuracy of rectal TLD dose measurements. Following the selection and insertion of suitable cylinder applicator size in the vagina, the TLD-100 rods on rectal marker were placed within the rectum. To make it re-useable, the rectal marker was always enclosed in a finger of a hand's glove and held tightly with an elastic rubber (Figure 1). The gloves containing the TLD rods were doubled in order to prevent the TLDs from absorbing body fluids within the cavity. It was then carefully pushed into the rectum such that the TLDs were placed posteriorly at 1 cm distance from the dose prescription points (5 mm rectal wall thickness plus 5 mm diameter of the rectal marker). The entire procedure was under sterile conditions. Before delivery of brachytherapy, a detailed gynecologic examination was performed to evaluate vaginal vault and to determine applicator diameter, as was done in a previous study [8]. Selection of the vaginal cylinder diameter and the treatment length for a given prescribed dose was made after vagina examination. Source loading of dwell points according to the defined treatment length was carried out after the selected cylinder was extracted from the applicator library of our TPS using HDR basic version 2.6. (Bebig, Germany). Standard (2D) plans similar to those shown in Figure 2 were created. In order to derive rectal doses from the TPS, a set of control points (described by coordinates given in Table 1B) was defined at 1 cm (5 mm rectal wall thickness plus 5 mm diameter of the rectal marker) away from dose prescription points. These points correspond to the position of the TLD rods in-vivo. In principle, a control point corresponds to a dwell point on an applicator. The dwell points used on the TPS at UCH have been set at 5 mm intervals (step size) and will normally vary with the treatment length along the straight IUT applicator (tandem) attached to the cylinder. Therefore, the total dwell points are 4, 6, 8, 10, 12, and 16 for treatment lengths 2, 3, 4, 5, 6, and 8 cm, respectively, involved in this study. Dose prescription points were 5 mm away from the posterior surface of the cylinder according to the recommendations of the International Commission on Radiation Units and Measurements [9]. The radius of the selected cylinder is in reference to its center. As such, ICBT fraction dose, which varied depending on the disease type, was prescribed at points 15, 17.5, and 20 mm from applicator midpoints for cylinders 20, 25, and 30 mm diameter, respectively. The prescribed doses were 15, 19.5, and 21 Gy in 3 fractions delivered on weekly basis, following EBRT of 45 Gy in 22 fractions with Theratron 780C telecobalt unit. The values calculated by the TPS at the prescription points and the set rectal dose points are presented in Tables 1A and 1B, respectively.
Fig. 1

Re-usable rectal marker (RM) bearing five TLDs (for rectal dose points, R1-R5) enclosed in a finger of a glove

Fig. 2

Variations in dose distributions in two brachytherapy standard plans using 30 mm central channel vaginal cylinder (CCVC) with different treatment lengths. A) 30 mm treatment length, B) 50 mm treatment length

Table 1

B) Sample of defined dose control (rectal thermoluminescent dosimeters – TLD) points’ details as reported on the treatment planning system (TPS) for a vaginal cylinder 30 mm, treatment length 3 cm and prescribed dose of 5 Gy

IdxX-Pos [cm]Y-Pos [cm]Z-Pos [cm]Dose [Gy]Dose % Rx
13.000.08–3.052.2444.8
23.000.07–2.552.4248.3
33.000.07–2.052.5150.3
43.000.07–1.552.5150.3
53.000.07–1.052.4248.4
63.000.06–0.552.2444.8

Ctrl Point Group Name: LAR01-01 # 3.00 cm

Rx = 5.00 Gy

Min. dose: 2.24 Gy

Average dose: 2.39 Gy (47.80% Rx)

Max. dose: 2.51 Gy

Table 1

A) A sample of dose prescription points’ details as reported on the treatment planning system (TPS) for a vaginal cylinder 30 mm, treatment length 3 cm and prescribed dose of 5 Gy

IdxX-Pos [cm]Y-Pos [cm]Z-Pos [cm]Dose [Gy]Dose % Rx
12.000.08–3.054.7194.2
22.000.07–2.554.9899.6
32.000.07–2.055.02100.4
42.000.07–1.555.03100.6
52.000.07–1.054.9899.6
62.000.06–0.554.7194.2

Ctrl Point Group Name: LAR01-01 # 2.00 cm

Rx = 5.00 Gy

Min. dose: 4.71 Gy

Average dose: 4.91 Gy (98.2% Rx)

Max. dose: 5.03 Gy

Re-usable rectal marker (RM) bearing five TLDs (for rectal dose points, R1-R5) enclosed in a finger of a glove Variations in dose distributions in two brachytherapy standard plans using 30 mm central channel vaginal cylinder (CCVC) with different treatment lengths. A) 30 mm treatment length, B) 50 mm treatment length A) A sample of dose prescription points’ details as reported on the treatment planning system (TPS) for a vaginal cylinder 30 mm, treatment length 3 cm and prescribed dose of 5 Gy Ctrl Point Group Name: LAR01-01 # 2.00 cm Rx = 5.00 Gy Min. dose: 4.71 Gy Average dose: 4.91 Gy (98.2% Rx) Max. dose: 5.03 Gy B) Sample of defined dose control (rectal thermoluminescent dosimeters – TLD) points’ details as reported on the treatment planning system (TPS) for a vaginal cylinder 30 mm, treatment length 3 cm and prescribed dose of 5 Gy Ctrl Point Group Name: LAR01-01 # 3.00 cm Rx = 5.00 Gy Min. dose: 2.24 Gy Average dose: 2.39 Gy (47.80% Rx) Max. dose: 2.51 Gy Brachytherapy was delivered immediately following applicator insertion and placement of rectal TLDs using the standard (2D) treatment plans. As no simulation radiographs were obtained, there was no additional dose contribution (from X-rays) to the dosimeters. Thirty four treatment procedures were performed in fourteen patients. At the end of each treatment, the rectal marker was removed from the patient and the dosimeters were carefully removed. The absorbed dose in the TLD after irradiation, in the form of electrons in the energy traps of the crystal lattice, was retrieved by heating up the dosimeter at a predefined temperature rate of 10°C/s to a maximum temperature (300°C/s) in the Harshaw 3500 TLD reader. The mean rectal dose measured with TLD for each insertion was obtained from the five dosimeters at points R1-R5 on the in-vivo marker.

Results

A total of 34 ICBT insertions was carried out between June and August 2013. The mean calculated doses (TPS) and measured doses (TLD rods) in these brachytherapy applications with the cylinder applicator sizes are presented in Table 2. The deviations of the measured from the mean calculated doses varied from – 1.9 to 2.1 Gy. Table 3 shows the comparisons between calculated TPS rectal doses and doses in-vivo with respect to the minimum and maximum values in the thirty-four ICBT insertions. Figure 3 compares the mean calculated rectal dose values with the mean measured (TLD) rectal doses. Average rectal dose calculated and measured dose per ICBT insertion ranged from 2.1-3.8 Gy and 1.2-5.6 Gy with overall mean doses of 3.0 ± 0.5 Gy and 3.1 ± 1.1 Gy, respectively. The combined (calculated and measured) means yielded 48.9% and 50.8% of the fractional prescription doses of 7 Gy (n = 16), 5 Gy (n = 15) and 6.5 Gy (n = 3). This did not show statistically significant difference (p = 0.427).
Table 2

Comparison of mean in-vivo doses and calculated (treatment planning system – TPS) doses in intracavitary brachytherapy (ICBT) with the cylinder applicator

s/nCylinder, diameter(mm)Treatment, length (cm)Prescribed dose (Gy)Mean calculated (TPS) dose at TLD points (Gy)Mean measured (in-vivo) doses (Gy) (TLD)Deviations (calculated – measured dose)
12047.03.11.91.2
22047.03.12.60.5
32047.03.13.00.1
42047.03.13.2–0.1
52055.02.44.1–1.7
62055.02.42.30.1
72525.02.12.6–0.5
82525.02.12.3–0.2
92537.03.12.70.4
102537.03.11.61.5
112547.03.31.22.1
122547.03.33.7–0.4
132545.02.41.31.1
142545.02.41.80.6
152545.02.42.00.4
162545.02.42.40
172557.03.44.1–0.7
182557.03.44.3–0.9
192557.03.43.30.1
202557.03.44.9–1.5
213035.02.42.9–0.5
223035.02.42.30.1
233057.03.74.1–0.4
243057.03.75.6–1.9
253056.53.44.9–1.5
263056.53.43.30.1
273056.53.44.1–0.7
283055.02.62.30.3
293065.02.73.0–0.3
303065.02.72.8–0.1
313065.02.72.9–0.2
323065.02.73.0–0.3
333087.03.84.2–0.4
343087.03.84.1–0.3
Mean values3.03.1–0.1

TLD – thermoluminescent dosimeters

Table 3

Comparisons between calculated treatment planning system (TPS) rectal doses and doses in-vivo with respect to the minimum and maximum values in the thirty-four intracavitary brachytherapy (ICBT) insertions

Cylinder, diameter (mm)Treatment, length (cm) n Min. calculated dose (TPS) (Gy)Min. in-vivo dose (Gy)Dev. (Gy)Max. calculated dose (Gy)Max. in-vivo dose (Gy)Dev. (Gy)
25222.01.50.52.12.9–0.8
25322.90.42.53.35.1–1.8
30322.20.41.82.55.7–3.2
20442.80.82.03.36.3–3.0
25462.20.51.72.55.3–2.8
20522.00.71.32.46.7–4.3
25543.12.20.93.67.0–3.4
30562.41.21.22.79.7–7.0
30643.21.12.13.77.5–3.8
30823.41.12.34.19.1–5.0
r = 0.30 r = 0.55

TPS – treatment planning system, ICBT – intracavitary brachytherapy, n – number of brachytherapy insertions, dev. – deviation of in-vivo dose from the calculated value

Fig. 3

Comparison of rectal calculated TPS dose values with TLD values in vaginal brachytherapy using cylinders. There is no significant difference between the calculated TPS doses and measured TLD doses (p = 0.427)

Comparison of rectal calculated TPS dose values with TLD values in vaginal brachytherapy using cylinders. There is no significant difference between the calculated TPS doses and measured TLD doses (p = 0.427) Comparison of mean in-vivo doses and calculated (treatment planning system – TPS) doses in intracavitary brachytherapy (ICBT) with the cylinder applicator TLD – thermoluminescent dosimeters Comparisons between calculated treatment planning system (TPS) rectal doses and doses in-vivo with respect to the minimum and maximum values in the thirty-four intracavitary brachytherapy (ICBT) insertions TPS – treatment planning system, ICBT – intracavitary brachytherapy, n – number of brachytherapy insertions, dev. – deviation of in-vivo dose from the calculated value

Discussion

Dose distribution in ICBT clinical target volume (CTV) with cylinder applicators will depend on treatment length, cylinder diameter, prescribed dose, distance from the first dwell position to the tip of the applicator, and the shape of the distal part of the applicator [10]. The most crucial are the first two factors as the treatment length and cylinder diameter. The treatment length, which is determined by the radiation oncologist in the brachytherapy suite will normally vary among patients and sometimes within a given patient from one insertion to another. The applicator diameter will depend on the size of the vaginal cavity such that the cylinder that can be adequately inserted and fitted within the vagina. Physical assessment of the patients by vagina examination during each brachytherapy insertion determined the length of the IUT that was activated with source loading as shown in Figure 2. Treatment lengths along the applicator in this study varied from 2-8 cm. In this study, we adopted a limit of 70% of the fractional prescription doses to assess the rectal doses measured in-vivo by the TLDs. In 31 (91.2%) and 3 (8.8%) insertions (Table 2), the mean measured doses were within and greater than 70% of the prescription doses, respectively. The extent (length) of treatment on the applicator increases the dimension of the isodose longitudinally and this will have impact on rectal dosimetry. In-vivo dosimetry in this study has shown that a larger applicator has the tendency to increase dose contribution within the rectum in vagina vault brachytherapy. In our study, the extreme rectal TLD doses were obtained when 30 mm cylinder applicator was used for 5-8 cm treatment lengths as indicated in Table 3. This confirmed the expected trend obtainable on the TPS where the mean calculated doses increased with cylinder size and the treatment lengths as indicated in Table 2. In a previous study, Demanes et al. [3] examined the use and advantages of a multichannel vaginal cylinder over the single-channel type in HDR brachytherapy. In our study, the vaginal cylinders produced the expected circular isodose distribution in the transverse plane, and the elongated oval shape in the AP and lateral planes as in the study by Demanes et al. The role of in-vivo dosimetry for cervical cancer patients treated with tandem and ovoids was evaluated in a study by Waldhäusl et al. [7]. Flexible probes from PTW/Germany were used. The rectal probes (type 9112) had five semiconductors (separated by 15 mm) and the bladder probes (type 9113) had one. The outer diameters were 7 and 3 mm, respectively. The probes were connected to a computerized control system. For the rectum, the difference between the calculated and the measured dose varied between – 31 and 90%. Forty-four out of fifty-five (80%) applications showed a higher calculated dose and only eleven out fifty five (20%) applications showed a higher measured dose. The observed differences in the previous study were mainly attributed to probe movement during the time interval between radiographs and end of the irradiation. The results of the present work have shown deviations between – 70.8 and 63.6%. In TLD dosimetry, 14/34 (41.2%) applications showed a higher calculated dose, 19/34 (55.9%) indicated a higher measured dose and 1 (2.9%) application showed no difference. In this case, movement of the rectal marker bearing the TLDs is minimal as the procedure did not involve radiographic imaging and the patients were not moved prior to treatment delivery. Table 2 shows that the minimum rectal TLD doses were lower than the corresponding TPS values with a weak correlation of 0.30. Conversely, the maximum in-vivo doses, which were higher than the calculated TPS values showed a better correlation of 0.55. The differences between doses measured in-vivo and the other method could be attributed to the slight intrinsic bend on the flexible rectal marker. Hence, part of the TLDs could be tilted inward or outward of the isodose line describing the dose distribution at the intended points of dose measurements. This would imply greater variations in measured rectal TLD doses for each brachytherapy insertions than as obtainable in TPS calculated values. The differences between the means of the calculated doses and the measured values were however found not to be statistically significant as depicted by Figure 3. This is also evident from the comparable overall mean doses of 3.0 ± 0.5 Gy and 3.1 ± 1.1 Gy corresponding to 48.9 and 50.8% of the prescribed dose, respectively. The marked difference in their standard deviations could be attributed to two reasons. First, for a given application, the calculated doses at predefined points longitudinally around the cylinder on a treatment plan are more reproducible than doses measured in-vivo at R1-R5 along the rectal marker. Also, the mean calculated dose for a given brachytherapy insertion remains unchanged for a patient when the cylinder applicator size and treatment length are the same in subsequent treatment fractions. This is not the case in in-vivo measurements, where the mean dose for each treatment application is unique. The pliability of the re-usable rectal marker engendered a wider range of in-vivo doses as reflected in the mean dose (per insertion) varying from 1.2-5.6 Gy compared with 2.1-3.8 Gy for the TPS values presented in Table 2. In 33 (97.1%) of the applications, the measured doses, which varied considerably across the rectal TLDs were minimum at the innermost TLDs on the rectal marker. In other words, the in-vivo doses were minimum at the superior end (first dosimeter) of the rectal marker and the maximum values were mostly obtained at the central (third) dosimeter. This is in similitude to the pattern of dose distribution observed across the defined control points on the TPS (Table 1B). Maximum rectal TLD doses obtained in this study were associated with the use of 30 mm vaginal cylinder for treatment lengths of 5-8 cm. But it was observed that the effect of the large applicator size on the in-vivo dose was reduced when the treatment length was < 4 cm. Therefore, in ICBT, the choice of the largest cylinder and source-activated length ≥ 5 cm must be clinically justified before use. In this circumstance, wherever possible, individual patient treatment planning for each vaginal vault treatment insertion may be adopted using 3D images. Magnetic resonance imaging guidance allows treatment to be optimized to tumor volumes while sparing the organs at risk [11].

Conclusions

This study, being a part of efforts to conform with good brachytherapy practice in HDR vaginal vault treatment, provides a preliminary evaluation of the current status. There was no significant difference in the mean doses to the rectum between values derived from the planning system and that of in-vivo dosimetry. The study outcome is an indication of the quality of brachytherapy practice at the University College Hospital, Ibadan, Nigeria. However, wherever possible, image-based brachytherapy planning should be adopted for treatment in lieu of 2D standard plans in order to control extreme doses that may arise at some points within the rectum volume. In the modern era, the use of 3D imaging is paramount due to the close proximity of critical organs, which are under-assessed with 2D planning [12].
  10 in total

1.  In-vivo rectal dose measurements with diodes to avoid misadministrations during intracavitary high dose rate brachytherapy for carcinoma of the cervix.

Authors:  R Alecu; M Alecu
Journal:  Med Phys       Date:  1999-05       Impact factor: 4.071

2.  The use and advantages of a multichannel vaginal cylinder in high-dose-rate brachytherapy.

Authors:  D J Demanes; S Rege; R R Rodriquez; K L Schutz; G A Altieri; T Wong
Journal:  Int J Radiat Oncol Biol Phys       Date:  1999-04-01       Impact factor: 7.038

3.  In-vivo dosimetry for gynaecological brachytherapy: physical and clinical considerations.

Authors:  Claudia Waldhäusl; André Wambersie; Richard Pötter; Dietmar Georg
Journal:  Radiother Oncol       Date:  2005-10-14       Impact factor: 6.280

4.  ICRU reference points and maximum doses of rectum and bladder in intracavitary radiotherapy.

Authors:  T Lahtinen; M Tenhunen; M Väyrynen
Journal:  Radiother Oncol       Date:  1993-08       Impact factor: 6.280

5.  High-dose-rate intracavitary brachytherapy: results of analyses of late rectal complications.

Authors:  Koh-ichi Sakata; Hisayasu Nagakura; Atushi Oouchi; Masanori Someya; Kensei Nakata; Mitsuo Shido; Kazumitsu Koito; Satoru Sagae; Ryuichi Kudo; Masato Hareyama
Journal:  Int J Radiat Oncol Biol Phys       Date:  2002-12-01       Impact factor: 7.038

6.  Recommendations for implementation of high dose rate 192Ir brachytherapy in developing countries by the Advisory Group of International Atomic Energy Agency.

Authors:  Subir Nag; Michael Dally; Marcela de la Torre; Hideo Tatsuzaki; Nori Kizilbash; Srichai Kurusun; Luis Pinillos; Boris Pokrajac; Ranjan Sur; Victor Levin
Journal:  Radiother Oncol       Date:  2002-09       Impact factor: 6.280

7.  Rectal complications in patients with carcinoma of the cervix treated with concomitant cisplatin and external beam irradiation with high dose rate brachytherapy: a dosimetric analysis.

Authors:  B G Clark; L Souhami; T N Roman; M D Evans; C Pla
Journal:  Int J Radiat Oncol Biol Phys       Date:  1994-03-30       Impact factor: 7.038

8.  Effects of bladder distension on dose distribution of vaginal vault brachytherapy in patients with endometrial cancer.

Authors:  Ozan C Guler; Cem Onal; Ibrahim Acibuci
Journal:  J Contemp Brachytherapy       Date:  2014-12-31

Review 9.  Brachytherapy for malignancies of the vagina in the 3D era.

Authors:  Scott M Glaser; Sushil Beriwal
Journal:  J Contemp Brachytherapy       Date:  2015-09-14

10.  Dosimetric impact of point A definition on high-dose-rate brachytherapy for cervical cancer: evaluations on conventional point A and MRI-guided, conformal plans.

Authors:  James Anderson; Yunfei Huang; Yusung Kim
Journal:  J Contemp Brachytherapy       Date:  2012-12-28
  10 in total
  3 in total

1.  The effect of tandem-ovoid titanium applicator on points A, B, bladder, and rectum doses in gynecological brachytherapy using 192Ir.

Authors:  Mohammad Hosein Sadeghi; Sedigheh Sina; Amir Mehdizadeh; Reza Faghihi; Vahed Moharramzadeh; Ali Soleimani Meigooni
Journal:  J Contemp Brachytherapy       Date:  2018-02-28

2.  Study of the relationships between brachytherapy plan parameters in an attempt to verify total dwell time in vaginal cylinder applications.

Authors:  Emmanuel O Oyekunle; Bidemi I Akinlade; Mutiu A Jimoh
Journal:  J Appl Clin Med Phys       Date:  2018-05-13       Impact factor: 2.102

Review 3.  Vaginal toxicity after high-dose-rate endovaginal brachytherapy: 20 years of results.

Authors:  Durim Delishaj; Amelia Barcellini; Romerai D'Amico; Stefano Ursino; Francesco Pasqualetti; Ilaria Costanza Fumagalli; Carlo Pietro Soatti
Journal:  J Contemp Brachytherapy       Date:  2018-12-28