| Literature DB >> 26816506 |
Emmanuel Oyeyemi Oyekunle1, Rachel Ibhade Obed2, Bidemi Idayat Akinlade3, Atara Ntekim3.
Abstract
PURPOSE: In-vivo measurements to determine doses to organs-at-risk can be an essential part of brachytherapy quality assurance (QA). This study compares calculated doses to the rectum with measured dose values as a means of QA in vaginal vault brachytherapy using cylinder applicators.Entities:
Keywords: brachytherapy; cervical cancer; in-vivo dose; rectum; vaginal cylinder; vaginal vault
Year: 2015 PMID: 26816506 PMCID: PMC4716134 DOI: 10.5114/jcb.2015.56767
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Fig. 1Re-usable rectal marker (RM) bearing five TLDs (for rectal dose points, R1-R5) enclosed in a finger of a glove
Fig. 2Variations in dose distributions in two brachytherapy standard plans using 30 mm central channel vaginal cylinder (CCVC) with different treatment lengths. A) 30 mm treatment length, B) 50 mm treatment length
B) Sample of defined dose control (rectal thermoluminescent dosimeters – TLD) points’ details as reported on the treatment planning system (TPS) for a vaginal cylinder 30 mm, treatment length 3 cm and prescribed dose of 5 Gy
| Idx | X-Pos [cm] | Y-Pos [cm] | Z-Pos [cm] | Dose [Gy] | Dose % Rx |
|---|---|---|---|---|---|
| 1 | 3.00 | 0.08 | –3.05 | 2.24 | 44.8 |
| 2 | 3.00 | 0.07 | –2.55 | 2.42 | 48.3 |
| 3 | 3.00 | 0.07 | –2.05 | 2.51 | 50.3 |
| 4 | 3.00 | 0.07 | –1.55 | 2.51 | 50.3 |
| 5 | 3.00 | 0.07 | –1.05 | 2.42 | 48.4 |
| 6 | 3.00 | 0.06 | –0.55 | 2.24 | 44.8 |
Ctrl Point Group Name: LAR01-01 # 3.00 cm
Rx = 5.00 Gy
Min. dose: 2.24 Gy
Average dose: 2.39 Gy (47.80% Rx)
Max. dose: 2.51 Gy
A) A sample of dose prescription points’ details as reported on the treatment planning system (TPS) for a vaginal cylinder 30 mm, treatment length 3 cm and prescribed dose of 5 Gy
| Idx | X-Pos [cm] | Y-Pos [cm] | Z-Pos [cm] | Dose [Gy] | Dose % Rx |
|---|---|---|---|---|---|
| 1 | 2.00 | 0.08 | –3.05 | 4.71 | 94.2 |
| 2 | 2.00 | 0.07 | –2.55 | 4.98 | 99.6 |
| 3 | 2.00 | 0.07 | –2.05 | 5.02 | 100.4 |
| 4 | 2.00 | 0.07 | –1.55 | 5.03 | 100.6 |
| 5 | 2.00 | 0.07 | –1.05 | 4.98 | 99.6 |
| 6 | 2.00 | 0.06 | –0.55 | 4.71 | 94.2 |
Ctrl Point Group Name: LAR01-01 # 2.00 cm
Rx = 5.00 Gy
Min. dose: 4.71 Gy
Average dose: 4.91 Gy (98.2% Rx)
Max. dose: 5.03 Gy
Comparison of mean in-vivo doses and calculated (treatment planning system – TPS) doses in intracavitary brachytherapy (ICBT) with the cylinder applicator
| s/n | Cylinder, diameter(mm) | Treatment, length (cm) | Prescribed dose (Gy) | Mean calculated (TPS) dose at TLD points (Gy) | Mean measured (in-vivo) doses (Gy) (TLD) | Deviations (calculated – measured dose) |
|---|---|---|---|---|---|---|
| 1 | 20 | 4 | 7.0 | 3.1 | 1.9 | 1.2 |
| 2 | 20 | 4 | 7.0 | 3.1 | 2.6 | 0.5 |
| 3 | 20 | 4 | 7.0 | 3.1 | 3.0 | 0.1 |
| 4 | 20 | 4 | 7.0 | 3.1 | 3.2 | –0.1 |
| 5 | 20 | 5 | 5.0 | 2.4 | 4.1 | –1.7 |
| 6 | 20 | 5 | 5.0 | 2.4 | 2.3 | 0.1 |
| 7 | 25 | 2 | 5.0 | 2.1 | 2.6 | –0.5 |
| 8 | 25 | 2 | 5.0 | 2.1 | 2.3 | –0.2 |
| 9 | 25 | 3 | 7.0 | 3.1 | 2.7 | 0.4 |
| 10 | 25 | 3 | 7.0 | 3.1 | 1.6 | 1.5 |
| 11 | 25 | 4 | 7.0 | 3.3 | 1.2 | 2.1 |
| 12 | 25 | 4 | 7.0 | 3.3 | 3.7 | –0.4 |
| 13 | 25 | 4 | 5.0 | 2.4 | 1.3 | 1.1 |
| 14 | 25 | 4 | 5.0 | 2.4 | 1.8 | 0.6 |
| 15 | 25 | 4 | 5.0 | 2.4 | 2.0 | 0.4 |
| 16 | 25 | 4 | 5.0 | 2.4 | 2.4 | 0 |
| 17 | 25 | 5 | 7.0 | 3.4 | 4.1 | –0.7 |
| 18 | 25 | 5 | 7.0 | 3.4 | 4.3 | –0.9 |
| 19 | 25 | 5 | 7.0 | 3.4 | 3.3 | 0.1 |
| 20 | 25 | 5 | 7.0 | 3.4 | 4.9 | –1.5 |
| 21 | 30 | 3 | 5.0 | 2.4 | 2.9 | –0.5 |
| 22 | 30 | 3 | 5.0 | 2.4 | 2.3 | 0.1 |
| 23 | 30 | 5 | 7.0 | 3.7 | 4.1 | –0.4 |
| 24 | 30 | 5 | 7.0 | 3.7 | 5.6 | –1.9 |
| 25 | 30 | 5 | 6.5 | 3.4 | 4.9 | –1.5 |
| 26 | 30 | 5 | 6.5 | 3.4 | 3.3 | 0.1 |
| 27 | 30 | 5 | 6.5 | 3.4 | 4.1 | –0.7 |
| 28 | 30 | 5 | 5.0 | 2.6 | 2.3 | 0.3 |
| 29 | 30 | 6 | 5.0 | 2.7 | 3.0 | –0.3 |
| 30 | 30 | 6 | 5.0 | 2.7 | 2.8 | –0.1 |
| 31 | 30 | 6 | 5.0 | 2.7 | 2.9 | –0.2 |
| 32 | 30 | 6 | 5.0 | 2.7 | 3.0 | –0.3 |
| 33 | 30 | 8 | 7.0 | 3.8 | 4.2 | –0.4 |
| 34 | 30 | 8 | 7.0 | 3.8 | 4.1 | –0.3 |
| Mean values | 3.0 | 3.1 | –0.1 |
TLD – thermoluminescent dosimeters
Comparisons between calculated treatment planning system (TPS) rectal doses and doses in-vivo with respect to the minimum and maximum values in the thirty-four intracavitary brachytherapy (ICBT) insertions
| Cylinder, diameter (mm) | Treatment, length (cm) |
| Min. calculated dose (TPS) (Gy) | Min. in-vivo dose (Gy) | Dev. (Gy) | Max. calculated dose (Gy) | Max. in-vivo dose (Gy) | Dev. (Gy) |
|---|---|---|---|---|---|---|---|---|
| 25 | 2 | 2 | 2.0 | 1.5 | 0.5 | 2.1 | 2.9 | –0.8 |
| 25 | 3 | 2 | 2.9 | 0.4 | 2.5 | 3.3 | 5.1 | –1.8 |
| 30 | 3 | 2 | 2.2 | 0.4 | 1.8 | 2.5 | 5.7 | –3.2 |
| 20 | 4 | 4 | 2.8 | 0.8 | 2.0 | 3.3 | 6.3 | –3.0 |
| 25 | 4 | 6 | 2.2 | 0.5 | 1.7 | 2.5 | 5.3 | –2.8 |
| 20 | 5 | 2 | 2.0 | 0.7 | 1.3 | 2.4 | 6.7 | –4.3 |
| 25 | 5 | 4 | 3.1 | 2.2 | 0.9 | 3.6 | 7.0 | –3.4 |
| 30 | 5 | 6 | 2.4 | 1.2 | 1.2 | 2.7 | 9.7 | –7.0 |
| 30 | 6 | 4 | 3.2 | 1.1 | 2.1 | 3.7 | 7.5 | –3.8 |
| 30 | 8 | 2 | 3.4 | 1.1 | 2.3 | 4.1 | 9.1 | –5.0 |
|
|
|
TPS – treatment planning system, ICBT – intracavitary brachytherapy, n – number of brachytherapy insertions, dev. – deviation of in-vivo dose from the calculated value
Fig. 3Comparison of rectal calculated TPS dose values with TLD values in vaginal brachytherapy using cylinders. There is no significant difference between the calculated TPS doses and measured TLD doses (p = 0.427)