| Literature DB >> 26815644 |
Hunter J Sismaet1, Anirban Banerjee2, Sean McNish2, Yongwook Choi3, Manolito Torralba3, Sarah Lucas3, Agnes Chan3, Victoria K Shanmugam2, Edgar D Goluch1.
Abstract
In clinical practice, point-of-care diagnostic testing has progressed rapidly in the last decade. For the field of wound care, there is a compelling need to develop rapid alternatives for bacterial identification in the clinical setting, where it generally takes over 24 hours to receive a positive identification. Even new molecular and biochemical identification methods require an initial incubation period of several hours to obtain a sufficient number of cells prior to performing the analysis. Here we report the use of an inexpensive, disposable electrochemical sensor to detect pyocyanin, a unique, redox-active quorum sensing molecule released by Pseudomonas aeruginosa, in wound fluid from patients with chronic wounds enrolled in the WE-HEAL Study. By measuring the metabolite excreted by the cells, this electrochemical detection strategy eliminates sample preparation, takes less than a minute to complete, and requires only 7.5 μL of sample to complete the analysis. The electrochemical results were compared against 16S rRNA profiling using 454 pyrosequencing. Blind identification yielded 9 correct matches, 2 false negatives, and 3 false positives giving a sensitivity of 71% and specificity of 57% for detection of Pseudomonas. Ongoing enhancement and development of this approach with a view to develop a rapid point-of-care diagnostic tool is planned.Entities:
Mesh:
Substances:
Year: 2016 PMID: 26815644 PMCID: PMC4853203 DOI: 10.1111/wrr.12414
Source DB: PubMed Journal: Wound Repair Regen ISSN: 1067-1927 Impact factor: 3.617