Optimal size of cuff bladder for indirect measurement of arterial pressure in adults.

A E Russell1, L M Wing, S A Smith, P E Aylward, R J McRitchie, R M Hassam, M J West, J P Chalmers.   

Abstract

This study tested the hypothesis that a sphygmomanometer cuff bladder long enough to encircle the arm in most adults ('obese cuff') would provide a more accurate and precise estimate of intra-arterial pressure than the usual 'standard' cuff bladder. In 53 patients undergoing diagnostic coronary angiography (35 males, 18 females, aged 36-79 years), indirect blood pressure, measured in the left arm with a random-zero sphygmomanometer, was compared with simultaneously measured femoral intra-arterial pressure. Duplicate indirect measurements were made with each of two cuffs containing bladders measuring 39 x 15 cm ('obese') and 23 x 12 cm ('standard'). The obese cuff bladder encircled 80% or more of the arm circumference in all subjects, whereas the standard cuff bladder met this requirement in only 19% of the subjects. For both systolic and diastolic pressure there was marked interindividual variability in the differences between indirect and direct measurements with both cuffs. With the obese cuff there was no systematic error in the diastolic blood pressure measurement. The standard cuff consistently overestimated diastolic pressure by 7.7 +/- 8.3 mmHg (mean +/- s.d.). For both cuffs, the difference between indirect and direct diastolic pressure increased with arm size (P less than 0.05). Both cuffs underestimated systolic blood pressure, the obese cuff by 15.5 +/- 11.7 mmHg and the standard cuff by 7.6 +/- 12.1 mmHg. These systolic blood pressure underestimates were greater at higher blood pressures (P less than 0.01) and with smaller arms (P less than 0.05). Age was not related to measurement error with either cuff.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Year:  1989        PMID: 2681407     DOI: 10.1097/00004872-198908000-00002

Source DB:  PubMed          Journal:  J Hypertens        ISSN: 0263-6352            Impact factor:   4.844


  11 in total

Review 1.  Evidence based treatment of hypertension. Measurement of blood pressure: an evidence based review.

Authors:  F A McAlister; S E Straus
Journal:  BMJ       Date:  2001-04-14

2.  Use of a neonatal blood pressure cuff to monitor blood pressure in the adult finger--comparison with a standard adult arm cuff.

Authors:  S Q Khan; J M Wardlaw; R Davenport; J Slattery; S Lewis
Journal:  J Clin Monit Comput       Date:  1998-05       Impact factor: 2.502

Review 3.  Efficacy and toxicity of antihypertensive pharmacotherapy relative to effective dose 50.

Authors:  Simon B Dimmitt; Hans G Stampfer; Jennifer H Martin; Robin E Ferner
Journal:  Br J Clin Pharmacol       Date:  2019-08-19       Impact factor: 4.335

4.  A Meta-analysis to Determine the Validity of Taking Blood Pressure Using the Indirect Cuff Method.

Authors:  Scott J Dankel; Minsoo Kang; Takashi Abe; Jeremy P Loenneke
Journal:  Curr Hypertens Rep       Date:  2019-02-07       Impact factor: 5.369

5.  Agreement between large and small cuffs in sphygmomanometry: a quantitative assessment.

Authors:  Y Iyriboz; C M Hearon; K Edwards
Journal:  J Clin Monit       Date:  1994-03

6.  Use of arterial catheters in the management of acute aortic dissection.

Authors:  Michael W Ruszala; Andrew P Reimer; Ronald L Hickman; John M Clochesy; Fredric M Hustey
Journal:  Air Med J       Date:  2014-11-05

7.  Knowledge, attitude, behavior, and practice of interns toward errors in sphygmomanometer and blood pressure measurement.

Authors:  Ramachandran Meenakshisundaram; Shah Sweni; Govindaraj Dhanalakshmi; Raju Sridevi; Ponniah Thirumalaikolundusubramanian
Journal:  J Clin Hypertens (Greenwich)       Date:  2009-12       Impact factor: 3.738

8.  Influence of sex on the accuracy of oscillometric-derived blood pressures.

Authors:  Skand D Bhatt; Alan L Hinderliter; George A Stouffer
Journal:  J Clin Hypertens (Greenwich)       Date:  2010-11-24       Impact factor: 3.738

9.  Office blood pressure measurement practices among community health providers (medical and paramedical) in northern district of India.

Authors:  Bishav Mohan; Naved Aslam; Upma Ralhan; Sarit Sharma; Naveen Gupta; Vivudh Pratap Singh; Shibba Takkar; G S Wander
Journal:  Indian Heart J       Date:  2014-08-06

10.  Differences in blood pressure measurements obtained using an automatic oscillometric sphygmomanometer depending on clothes-wearing status.

Authors:  Ji Hoon Ki; Mi Kyeong Oh; Soo Hee Lee
Journal:  Korean J Fam Med       Date:  2013-03-20
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.