Myocardial recovery during mechanical circulatory support: weaning and explantation criteria.

Michael Dandel1, Roland Hetzer1.   

Abstract

For selected patients ventricular assist device explantation is feasible. Freedom from heart transplantation and VAD support for >15 years following VAD removal is possible even if recovery is incomplete and the underlying cause for ventricular assist device implantation was a chronic form of cardiomyopathy. Echocardiography and rightheart catheterization are necessary to assess cardiac recovery. However, to date there is no gold standard for recovery assessment. Certain echocardiography parameters appeared highly predictive for long-term (≥5 years) post-weaning cardiac function and reliable for weaning decisions.

Entities:  

Keywords:  heart failure; myocardial recovery; risk factors; survival; ventricular assist devices; ventricular function

Year:  2015        PMID: 26811833      PMCID: PMC4712030     

Source DB:  PubMed          Journal:  Heart Lung Vessel        ISSN: 2282-8419


Weaning from mechanical circulatory support After long controversy about the clinical relevance of unloading-promoted recovery from chronic heart failure (HF) and about the feasibility of elective weaning of patients from their ventricular assist device (VAD), taking into account the risk of early post-weaning HF recurrence, this topic has gained increasing interest over the past 10 years, especially after the publication of the first long-term weaning results in patients with chronic HF reversal after LVAD implantation [1,2,3]. Temporary interruptions of mechanical unloading (“off-pump trials”) are mandatory for assessment of recovery, regardless of whether the assessment is performed at rest or during exercise. Short off-pump trials allow evaluation of the heart without mechanical support under the same circumstances that will exist after VAD removal. However, whereas pulsatile VADs allow optimal assessment of heart function during complete pump stops, complete stops of axial-flow pumps lead to retrograde flow into the LV followed by reduction of the diastolic arterial pressure that, by reducing the left ventricular (LV) afterload, can generate overestimations of LV systolic function. According to our observations, the misleading retrograde bloodflow into the LV during off-pump trials is a major negative factor that can interfere with successful VAD explantation. Therefore, for such pumps, rotor-speed reduction to values resulting in close to zero flow in one cardiac cycle (3000-6000 rpm) is better than complete pump-stop [2,3,4]. Before off-pump trials, heparin must be given (60-100 IU/kg according to the prothrombin time reported as INR) to prevent thrombus formation inside the pump [5,6]. Patients with heparin-induced thrombocytopenia should receive argatroban (synthetic thrombin inhibitor) infusions (2μg/kg/min)which should be started 1h before the off-pump trials [6]. Additionally, completely stopped pulsatile devices should be allowed to pump at least once a minute, or 3 bursts of pneumatic hand-pumping should be intercalated every 15 to 60 seconds to prevent blood stagnation [2, 4, 5, 7, 8]. In patients with a biventricular assist device (BiVAD) it appeared appropriate to stop the right ventricular (RV) pump 30 seconds earlier than the LV pump; after both pumps are deactivated, evaluation of recovery can be performed [6]. Duration of individual off-pump periodscan vary between 3 and 15 minutes [4,5,6]. In patients with insufficient recovery for possible VAD removal, as already shown during the first 3 minutes, there is no reason to further extend the off-pump trial [9]. With appropriate caution, the risk of interruption of unloading is low [4, 6, 9, 10]. In weaning candidates it appears useful to conduct such trials weekly or every 2 weeks and to make the final decision for elective VAD explantation only after cardiac improvement has reached its maximum (no further improvement in at least 2 consecutive off-pump trials) [9]. During recovery it appears useful to change the working mode of the pumps, in order to intensify the unloading if the ventricle size needs further reduction, or to exert moderate load on the ventricular myocardium after maximum improvement [1, 5, 7]. In addition to the retrograde blood flow into the LV during off-pump trials, which makes the evaluation of LV function very difficult, another factor that may impair successful VAD explantation in the subgroup of patients receiving continuous VAD is the usually great difficulty in maintaining the recovered heart in a state as unmolested as possible throughout explantation surgery. During VAD removal, which was performed on the beating heart in all our weaned patients, we aimed to leave the intrathoracic part of both cannulas in place after they were reliably occluded [6, 9]. This was possible in 91% of patients with pulsatile pumps, but only in 57% of patients with continuous-flow pumps. Also, the heart-lung machine, which was used only in less than 10% of patients with pulsatile VADs, was required for all continuous-flow-pump explantations. Nevertheless, in our patients with non-ischemic chronic cardiomyopathy, there were no significant differences in the 3-year post-explant freedom from HF recurrence between patients weaned from pulsatile and those weaned from continuous flow pumps [9]. The main diagnostic methods used to assess cardiac recovery are echocardiography and heart catheterization. Cardiopulmonary exercise testing is also increasingly used by some centers [4, 8]. Echocardiography is the cornerstone for assessment of cardiac recovery allowing both selection of potential weaning candidates and decision-making in favor of or against VAD removal. After VAD implantation,repeated transthoracic echocardiography (TTE) screenings with normal VAD function are necessary for selection of potential weaning candidates, i.e. patients with signs of relevant reverse remodeling who also seem to have their contractile function improved to levels deserving further evaluation. Such potential weaning candidates are those with LV end-diastolic diameter (LVEDD)< 55mm (or 55-60mm at BSA ≥1.8m²)and fractional shortening (FS)>15%, no or ≤ grade I mitral and/or aortic regurgitation, no RV dilation and tricuspid regurgitation ≤ grade II [6]. Before the first off-pump trial, it is useful to perform stepwise pump-rate reductions under TTE monitoring, to verify whether complete interruption of unloading is possible. Thus, if incomplete unloading interruption alreadyprovokes symptoms (dizziness, sweating, etc.), complete interruption of unloading is senseless and risky. If the patient remains asymptomatic but the LVEDD increases beyond 60mm, a complete interruption is also senseless, because such a patient is not yet a weaning candidate. Echocardiographic assessment of recovery in weaning candidates is usually based on the results of repeated off-pump trials at rest [2, 4, 7, 9]. After months of intensive VAD support, even short periods of physiological loading by interruption of unloading may represent a serious challenge for a possibly incompletely-recovered ventricle. It therefore appears reasonable to avoid (at least initially) any risk of myocardial exhaustion, which might interfere with possibly still ongoing recovery. For this reason, all 119 patients who were weaned since 1995 in our center from long-term VADs primarily designed as a bridge-to-HTx or as permanent therapy underwent assessment of heart function exclusively at rest [6, 9, 10]. Although at rest the conclusive value of assessments is limited by the lack of information about inotropic reserves and cardiac adaptation to stress, our weaning results appeared not to be relevantly affected by that, insofar as the results reported by groups who also used stress echocardiography and/or exercise testing were not better [4, 9, 11]. However, stress echocardiography can provide valuable information on inotropic reserves and adaptation to stress, which might be useful for weaning decisions. According to certain authors, an absolute EF increase by 5% during dopamine infusion indicates preservation of contractile reserve [4]. A possible limitation of dobutamine stress echocardiographycan be, at least theoretically, the risk of myocardial exhaustion with negative impact on an ongoingmyocardialrecovery process. Further studies are therefore necessary to establish the real value of dobutamine stress echocardiographyfor weaning decisions. In principle, off-pump TTE should be as comprehensive as possible, including tissue Doppler and strain imaging. Table 1 shows the TTE parameters we routinely measurein our department. Transthoracic echocardiographic measurements for evaluation of cardiac recovery during off-pump trials. * RWTED = [interventricular septum thickness + posterior wall thickness] / LVEDD. ** Product of time velocity integral measured with pulsed-wave Doppler at the LV outflow tract (LVOT) and cross-sectional area of the LVOT. LV = left ventricular. However,not all these parameters can be reliably measured in all patients because of poor image quality in some patients with VAD support. Image quality was a serious limitation, especially for off-pump Doppler measurements that had to be performed during rotor-speed reduction instead of complete pump stop. Off-pump right heart catheterization (RHC) is also a cornerstone for recovery assessment [4, 9] and is necessary for final decisions in favor of or against VAD explantation. A final off-pump trial of ≥15minutes in the operation room, with repeated measurements of hemodynamic parameters under continuous echocardiographic monitoring, is indispensablebefore the start of explantation surgery [9]. RHC appeared also necessary before any preliminary decision-making in potential weaning candidates with borderline TTE data and/or relevant cardiac improvement only after >6 months of unloading and/or long history duration (>3 years) of HF [9]. In patients with axial-flow pumps, such preliminary RHC appears more reliable if off-pump measurements are preceded by occlusion of the outlet cannula with a balloon [9]. Balloon occlusion allows complete pump stops without any misleading retrograde flow into the ventricle. Normal and stable hemodynamics during off-pump RHC trials is a necessary condition for a decision in favor of VAD removal, but not sufficiently predictive for long-term post-explant cardiac stability [6, 9]. It was observed that, unlike non-recovered patients, those with unloading-promoted recovery show a significant LVEF increase in the 6-minute walking test (6MW) [4]. Non-recovered patients showed a significant correlation between heart rate (HR) and mean arterial pressure (MAP) after 6MW, suggesting that HR increase was compensatoryfor MAP reduction, whereas recovered patients showed after the 6MW no correlation between HR and MAP, suggesting that HR increase was independent of MAP change and therefore a true inotropic reserve response [4]. Harefield Hospital (UK) developed an algorithm for testing recovery that includes a 6MW test with repeated measurements afterwards to determine the inotropic reserve [4, 8]. The same group also uses cardiopulmonary exercise testing with oxygen uptake (VO2) as a parameter for weaning decisions [4]. Our weaning criteria that have evolved during nearly 20 years of weaning experience are summarized in Table 2. Main Left Ventricular Assist Device (LVAD) Explantation Criteria * *  Thesecriteria used at present at the Deutsches Herzzentrum Berlin have evolved from a weaning experience of 18 years after explantation of 100 long-term VADs, primarily designed as bridge-to-HTx or permanent therapy.6,9 **Measurements at rest, without inotropic support. Since 2005 also speckle-tracking derived 2D-strain imaging parameters (global longitudinal strain and strain rate, intra-ventricular synchrony and synergy, including pre-explant stability of all these parameters after maximum improvement and during the final off-pump trial) were taken into consideration for weaning decisions.6 Essential criteria are, in addition to a stable off-pump LVEF ≥45%, normal and stable LV size and shape, as well as high stability of hemodynamic parameters measured during the off-pump right heart catheterization (RHC) trials [6, 9]. Even in patients with good recovery, it appears useful to explant VADs only after the recovery has reached its maximum (no further improvement during the next 2-4 weeks) [9]. If a relevant LVEF reduction and/or LV enlargement occurs during the next 2-4 weeks after maximum cardiac recovery, it can be risky to explant the VAD, even if the LVEFdid not fall below 45% and theLVEDDdid not increase beyond55mm [6]. The weaning criteria used by the Harefield group include fewer TTE parameters; instead they usethe 6MW-test and the peak oxygen consumption (VO2)(optimal: >16mL/kg/min or >65% predicted)[4]. However, it is important to emphasize that they also do not explantLVADsin patients with LVEF < 45% [4, 6, 9]. For more reliable evaluation of cardiac recovery, since 2005 we have also used tissue Doppler and speckle-tracking-derived 2D-strain imaging [12, 13]. The potential usefulness of tissue Doppler wall motion velocity measurements is shown in Figure 1. Usefulness of left ventricular (LV) peak systolic wall motion velocity (Sm) measurements at the basal posterior wall by pulsed-wave tissue Doppler toassess the stability of LV contractile function during and between off-pump trials and for detection of further cardiac improvement after   left ventricularejection fraction (LVEF) has reached its peak value. At the time when LVEF reached its maximum value (55%), Sm was 9cm/s during full left ventricular assist device (LVAD) support (A) and remained stable during the off-pump trial (B). Although the LVEF remained unchanged, the Sm values with (C) and without(D) LVAD support increased during the next 3 weeks of unloading up to 11cm/s (+22%). Thereafter, there was no further Sm improvement and the LVAD was successfully explanted. The advantages of strain imaging are its ability to differentiate between active and passive movement of myocardial segments, the angle independency of deformation-velocity measurements, and the possibility to quantify intra-ventricular asynchrony and dyssynergy and to evaluate components of myocardial function, such as longitudinal myocardial shortening, thatcannot be visually assessed [12]. Unfortunately, as 2D-strain imaging is a new method, the available off-pump data are insufficient for reliable assessment of their predictive value for long-term post-explant cardiac stability. However, in borderline cases, off-pump data on deformation velocity as well as on intraventricular synchrony and synergy of contraction, including their stabilityafter maximum improvement, can be useful for weaning decisions. The major limitation of this method is its dependence on image quality. Figures 2 and 3 show the potential usefulness of 2D-strain imaging for assessment of recovery after LVAD implantation. Usefulness of speckle-tracking derived 2D-stain imaging for assessment of cardiac recovery after left ventricular assist device (LVAD) implantation. The comparison of recordingsobtained before LVAD implantation (A and B) and several months later before LVADexplantation (C and D) showed a striking improvement of systolic global longitudinal strain (myocardial longitudinal shortening) and strain-rate (velocity of myocardial longitudinal shortening) which increased during LVAD support from 4% (A) and 0.3/s (B), respectively to 17.5% (C) and 9.6/s (D), respectively. The global early diastolic strain rate (early relaxation velocity) also increased from 0.7/s (B) to 1.6/s (D) and image D also shows a normalization of early/late strain-rate ratio. Speckle-tracking derived 2D-strain and strain rate off-pump recordings showing further cardiac improvement during left ventricular assist device (LVAD) support after off-pumpleft ventricularejection fraction(LVEF) reached its maximum value of 50%.At the time of maximum LVEF improvement (images A and B) longitudinal global systolic strain (SL) and strain-rate (SrL) were slightly reduced (17% and 1/s, respectively) and the global early/late diastolic strain-rate ratio(SrLE/SrLA) was high. After few weeks of further LVAD support, although the LVEF remained unchanged, both SLandSrL increased, reaching normal values (22% and 1.25/s, respectively) and there was also improvement in intraventricular synchrony of contraction (images C and D). The 35% shortening of the time to peak strain rate (TpSr) additionally indicates improvement of left ventricular (LV) systolic function. The early relaxation velocity increase (from 1.7/s to 2.1/s) and the SrLE/SrLA ratio reduction (from 3.5 to 2.1) also indicate an improvement in LV diastolic function.

CONCLUSION

Echocardiography is the mainstay for assessment of unloading-promoted cardiac recovery. Off-pump LVEF ≥ 45% plus normal LVEDD at rest are basic weaning requirements and their stability after maximum improvement over the next 2 to 4 weeks during and between additional off-pump tests is predictable for long-term post-explant patient outcome. Tissue Doppler andstrain imaging data are useful for decision making especially if their stability after maximumremission is also considered. After maximum improvement, stable size, geometry and function also during moderate loading of theventricle by pump speed reduction for several days can be helpful for weaning decisions. Stable normal off-pump hemodynamics is necessary for weaning decision, but insufficiently predictive for long-term cardiac stability afterVAD removal. Cardio-pulmonary exercise testing (with VAD support and during off-pump trials) provides information on inotropic reserve and can be useful for weaning decisions.
  11 in total

1.  Assessment of myocardial recovery using dobutamine stress echocardiography in LVAD patients.

Authors:  T Khan; K Okerberg; A Hernandez; K Miller; T J. Myers; B Radovancevic; R M. Delgado; I Gregoric; O H. Frazier
Journal:  J Heart Lung Transplant       Date:  2001-02       Impact factor: 10.247

2.  Cardiac recovery in dilated cardiomyopathy by unloading with a left ventricular assist device.

Authors:  R Hetzer; J Müller; Y Weng; G Wallukat; S Spiegelsberger; M Loebe
Journal:  Ann Thorac Surg       Date:  1999-08       Impact factor: 4.330

3.  Long-term results in patients with idiopathic dilated cardiomyopathy after weaning from left ventricular assist devices.

Authors:  Michael Dandel; Yuguo Weng; Henryk Siniawski; Evgenij Potapov; Hans B Lehmkuhl; Roland Hetzer
Journal:  Circulation       Date:  2005-08-30       Impact factor: 29.690

4.  Hemodynamic and echocardiographic responses to acute interruption of left ventricular assist device support: relevance to assessment of myocardial recovery.

Authors:  Robert S George; Magdi H Yacoub; Giordano Tasca; Carole Webb; Christopher T Bowles; Patrick Tansley; James P Hardy; Gilles Dreyfus; Asghar Khaghani; Emma J Birks
Journal:  J Heart Lung Transplant       Date:  2007-10       Impact factor: 10.247

5.  Prediction of cardiac stability after weaning from left ventricular assist devices in patients with idiopathic dilated cardiomyopathy.

Authors:  Michael Dandel; Yuguo Weng; Henryk Siniawski; Evgenij Potapov; Thorsten Drews; Hans B Lehmkuhl; Christoph Knosalla; Roland Hetzer
Journal:  Circulation       Date:  2008-09-30       Impact factor: 29.690

Review 6.  Echocardiographic strain and strain rate imaging--clinical applications.

Authors:  Michael Dandel; Roland Hetzer
Journal:  Int J Cardiol       Date:  2008-08-29       Impact factor: 4.164

7.  Pre-explant stability of unloading-promoted cardiac improvement predicts outcome after weaning from ventricular assist devices.

Authors:  Michael Dandel; Yuguo Weng; Henryk Siniawski; Evgenij Potapov; Thomas Krabatsch; Hans B Lehmkuhl; Thorsten Drews; Christoph Knosalla; Roland Hetzer
Journal:  Circulation       Date:  2012-09-11       Impact factor: 29.690

Review 8.  Contribution of ventricular assist devices to the recovery of failing hearts: a review and the Berlin Heart Center Experience.

Authors:  Michael Dandel; Christoph Knosalla; Roland Hetzer
Journal:  Eur J Heart Fail       Date:  2013-12-20       Impact factor: 15.534

9.  Bridging-to-recovery.

Authors:  R Hetzer; J H Müller; Y Weng; R Meyer; M Dandel
Journal:  Ann Thorac Surg       Date:  2001-03       Impact factor: 4.330

10.  Heart failure reversal by ventricular unloading in patients with chronic cardiomyopathy: criteria for weaning from ventricular assist devices.

Authors:  Michael Dandel; Yuguo Weng; Henryk Siniawski; Alexander Stepanenko; Thomas Krabatsch; Evgenij Potapov; Hans B Lehmkuhl; Christoph Knosalla; Roland Hetzer
Journal:  Eur Heart J       Date:  2010-10-07       Impact factor: 29.983

View more
  6 in total

1.  Mechanical circulatory support device-heart hysteretic interaction can predict left ventricular end diastolic pressure.

Authors:  Brian Y Chang; Steven P Keller; Sonya S Bhavsar; Noam Josephy; Elazer R Edelman
Journal:  Sci Transl Med       Date:  2018-02-28       Impact factor: 17.956

Review 2.  LVAD as a Bridge to Remission from Advanced Heart Failure: Current Data and Opportunities for Improvement.

Authors:  Christos P Kyriakopoulos; Chris J Kapelios; Elizabeth L Stauder; Iosif Taleb; Rana Hamouche; Konstantinos Sideris; Antigone G Koliopoulou; Michael J Bonios; Stavros G Drakos
Journal:  J Clin Med       Date:  2022-06-20       Impact factor: 4.964

3.  Dynamic Modulation of Device-Arterial Coupling to Determine Cardiac Output and Vascular Resistance.

Authors:  Steven P Keller; Brian Y Chang; Qing Tan; Zhengyang Zhang; Ahmad El Katerji; Elazer R Edelman
Journal:  Ann Biomed Eng       Date:  2020-04-13       Impact factor: 3.934

4.  A Scalable Approach to Determine Intracardiac Pressure From Mechanical Circulatory Support Device Signals.

Authors:  Brian Y Chang; Christian Moyer; Ahmad El Katerji; Steven P Keller; Elazer R Edelman
Journal:  IEEE Trans Biomed Eng       Date:  2021-02-18       Impact factor: 4.538

5.  Joint EAPCI/ACVC expert consensus document on percutaneous ventricular assist devices.

Authors:  Alaide Chieffo; Dariusz Dudek; Christian Hassager; Alain Combes; Mario Gramegna; Sigrun Halvorsen; Kurt Huber; Vijay Kunadian; Jiri Maly; Jacob Eifer Møller; Federico Pappalardo; Giuseppe Tarantini; Guido Tavazzi; Holger Thiele; Christophe Vandenbriele; Nicolas van Mieghem; Pascal Vranckx; Nikos Werner; Susanna Price
Journal:  Eur Heart J Acute Cardiovasc Care       Date:  2021-06-30

6.  Hysteretic device characteristics indicate cardiac contractile state for guiding mechanical circulatory support device use.

Authors:  Brian Y Chang; Zhengyang Zhang; Steven P Keller; Elazer R Edelman; Kimberly Feng; Noam Josephy
Journal:  Intensive Care Med Exp       Date:  2021-12-20
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.