Literature DB >> 26807895

Comparison of Diffusion-Weighted Imaging in the Human Brain Using Readout-Segmented EPI and PROPELLER Turbo Spin Echo With Single-Shot EPI at 7 T MRI.

Ikuhiro Kida1, Takashi Ueguchi, Yuichiro Matsuoka, Kun Zhou, Alto Stemmer, David Porter.   

Abstract

OBJECTIVES: The purpose of the present study was to compare periodically rotated overlapping parallel lines with enhanced reconstruction-type turbo spin echo diffusion-weighted imaging (pTSE-DWI) and readout-segmented echo planar imaging (rsEPI-DWI) with single-shot echo planar imaging (ssEPI-DWI) in a 7 T human MR system. We evaluated the signal-to-noise ratio (SNR), image distortion, and apparent diffusion coefficient values in the human brain.
MATERIALS AND METHODS: Six healthy volunteers were included in this study. The study protocol was approved by our institutional review board. All measurements were performed at 7 T using pTSE-DWI, rsEPI-DWI, and ssEPI-DWI sequences. The spatial resolution was 1.2 × 1.2 mm in-plane with a 3-mm slice thickness. Signal-to-noise ratio was measured using 2 scans.
RESULTS: The ssEPI-DWI sequence showed significant image blurring, whereas pTSE-DWI and rsEPI-DWI sequences demonstrated high image quality with low geometrical distortion compared with reference T2-weighted, turbo spin echo images. Signal loss in ventral regions near the air-filled paranasal sinus/nasal cavity was found in ssEPI-DWI and rsEPI-DWI but not pTSE-DWI. The apparent diffusion coefficient values for ssEPI-DWI were 824 ± 17 × 10 and 749 ± 25 × 10 mm/s in the gray matter and white matter, respectively; the values obtained for pTSE-DWI were 798 ± 21 × 10 and 865 ± 40 × 10 mm/s; and the values obtained for rsEPI-DWI were 730 ± 12 × 10 and 722 ± 25 × 10 mm/s. The pTSE-DWI images showed no additional distortion comparison to the T2-weighted images, but had a lower SNR than ssEPI-DWI and rsEPI-DWI. The rsEPI-DWI sequence provided high-quality images with minor distortion and a similar SNR to ssEPI-DWI.
CONCLUSIONS: Our results suggest that the benefits of the rsEPI-DWI and pTSE-DWI sequences, in terms of SNR, image quality, and image distortion, appear to outweigh those of ssEPI-DWI. Thus, pTSE-DWI and rsEPI-DWI at 7 T have great potential use for clinical diagnoses. However, it is noteworthy that both sequences are limited by the scan time required. In addition, pTSE-DWI has limitations on the number of slices due to specific absorption rate. Overall, rsEPI-DWI is a favorable imaging sequence, taking into account the SNR and image quality at 7 T.

Entities:  

Mesh:

Year:  2016        PMID: 26807895     DOI: 10.1097/RLI.0000000000000248

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  7 in total

1.  Structural covariance and heritability of the optic tract and primary visual cortex in living human brains.

Authors:  Toshikazu Miyata; Noah C Benson; Jonathan Winawer; Hiromasa Takemura
Journal:  J Neurosci       Date:  2022-07-18       Impact factor: 6.709

2.  Assessment of Rat Sciatic Nerve Using Diffusion-Tensor Imaging With Readout-Segmented Echo Planar Imaging.

Authors:  Yueyao Chen; Zhongxian Pan; Fanqi Meng; Qian Xu; Leyu Huang; Xuejia Pu; Xuewen Yu; Yanglei Wu; Hanqing Lyu; Xiaofeng Lin
Journal:  Front Neurosci       Date:  2022-06-23       Impact factor: 5.152

3.  Performance of Single-Shot Echo-Planar Imaging in Diffusion Tensor Imaging in Rat Sciatic Nerve Compared With Readout-Segmented Echo-Planar Imaging.

Authors:  Yueyao Chen; Zhongxian Pan; Fanqi Meng; Zhujing Li; Yuanming Hu; Xuewen Yu; Jinyun Gao; Yihao Guo; Hanqing Lyu; Xiaofeng Lin
Journal:  Front Neurosci       Date:  2022-05-12       Impact factor: 5.152

4.  Single shot echo planar imaging (ssEPI) vs single shot turbo spin echo (ssTSE) DWI of the orbit in patients with ocular melanoma.

Authors:  Ekim Gumeler; Safak Parlak; Gozde Yazici; Erdem Karabulut; Hayyam Kiratli; Kader K Oguz
Journal:  Br J Radiol       Date:  2020-12-22       Impact factor: 3.039

5.  Predicting Neural Response Latency of the Human Early Visual Cortex from MRI-Based Tissue Measurements of the Optic Radiation.

Authors:  Hiromasa Takemura; Kenichi Yuasa; Kaoru Amano
Journal:  eNeuro       Date:  2020-07-02

6.  Performance of TGSE BLADE DWI compared with RESOLVE DWI in the diagnosis of cholesteatoma.

Authors:  Yaru Sheng; Rujian Hong; Yan Sha; Zhongshuai Zhang; Kun Zhou; Caixia Fu
Journal:  BMC Med Imaging       Date:  2020-04-19       Impact factor: 1.930

7.  Ultra-high field (10.5T) diffusion-weighted MRI of the macaque brain.

Authors:  Mark D Grier; Essa Yacoub; Gregor Adriany; Russell L Lagore; Noam Harel; Ru-Yuan Zhang; Christophe Lenglet; Kâmil Uğurbil; Jan Zimmermann; Sarah R Heilbronner
Journal:  Neuroimage       Date:  2022-04-13       Impact factor: 7.400

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.