Literature DB >> 26794906

Effects of Different Insertion Techniques of a Cochlear Implant Electrode on the Intracochlear Pressure.

Ingo Todt1, Arneborg Ernst, Philipp Mittmann.   

Abstract

To achieve a functional atraumatic insertion, low intracochlear pressure changes during the procedure are assumed to be important. The aim of this study was to observe intracochlear pressure changes due to different insertion techniques in a cochlear model. Cochlear implant electrode insertions were performed in an artificial cochlear model to record intracochlear pressure changes with a micropressure sensor to evaluate the maximum amplitude and frequency of pressure changes under different insertional conditions. We found statistically significant differences in the occurrence of intracochlear pressure peak changes comparing different techniques. Based on our model results, an insertion should be maximally supported to minimize micromovement-related pressure changes.
© 2016 S. Karger AG, Basel.

Mesh:

Year:  2016        PMID: 26794906     DOI: 10.1159/000442041

Source DB:  PubMed          Journal:  Audiol Neurootol        ISSN: 1420-3030            Impact factor:   1.854


  5 in total

1.  Investigation of ultra-low insertion speeds in an inelastic artificial cochlear model using custom-made cochlear implant electrodes.

Authors:  Silke Hügl; Katharina Rülander; Thomas Lenarz; Omid Majdani; Thomas S Rau
Journal:  Eur Arch Otorhinolaryngol       Date:  2018-10-09       Impact factor: 2.503

2.  Intracochlear Pressure Transients During Cochlear Implant Electrode Insertion.

Authors:  Nathaniel T Greene; Jameson K Mattingly; Renee M Banakis Hartl; Daniel J Tollin; Stephen P Cass
Journal:  Otol Neurotol       Date:  2016-12       Impact factor: 2.311

3.  Postinsertional Cable Movements of Cochlear Implant Electrodes and Their Effects on Intracochlear Pressure.

Authors:  I Todt; D Karimi; J Luger; A Ernst; P Mittmann
Journal:  Biomed Res Int       Date:  2016-11-09       Impact factor: 3.411

4.  Cochlear implant electrode sealing techniques and related intracochlear pressure changes.

Authors:  Ingo Todt; Julica Utca; Dania Karimi; Arne Ernst; Philipp Mittmann
Journal:  J Otolaryngol Head Neck Surg       Date:  2017-05-11

5.  Intracochlear Pressure Changes After Cochlea Implant Electrode Pullback-Reduction of Intracochlear Trauma.

Authors:  Gina Lauer; Julica Uçta; Lars Decker; Arneborg Ernst; Philipp Mittmann
Journal:  Laryngoscope Investig Otolaryngol       Date:  2019-07-11
  5 in total