| Literature DB >> 26779330 |
Ritvik P Mehta1, Sara L Mattson1, Brian A Kappus2, Robin L Seitzman3.
Abstract
The objective of the study was to assess the safety of the HyperSound® Audio System (HSS), a novel audio system using ultrasound technology, in normal hearing subjects under normal use conditions; we considered pre-exposure and post-exposure test design. We investigated primary and secondary outcome measures: i) temporary threshold shift (TTS), defined as >10 dB shift in pure tone air conduction thresholds and/or a decrement in distortion product otoacoustic emissions (DPOAEs) >10 dB at two or more frequencies; ii) presence of new-onset otologic symptoms after exposure. Twenty adult subjects with normal hearing underwent a pre-exposure assessment (pure tone air conduction audiometry, tympanometry, DPOAEs and otologic symptoms questionnaire) followed by exposure to a 2-h movie with sound delivered through the HSS emitter followed by a post-exposure assessment. No TTS or new-onset otological symptoms were identified. HSS demonstrates excellent safety in normal hearing subjects under normal use conditions.Entities:
Keywords: Hypersound®; directional audio; ultrasonic hearing; ultrasound safety
Year: 2015 PMID: 26779330 PMCID: PMC4698603 DOI: 10.4081/audiores.2015.136
Source DB: PubMed Journal: Audiol Res ISSN: 2039-4330
Participant baseline (pre-exposure) characteristics.
| Characteristic | N=20 |
|---|---|
| Male gender, n (%) | 8 (40%) |
| Female gender, n (%) | 12 (60%) |
| Age (years), mean±SD | 29±7 (range: 21-53) |
| PTA (dB), mean±SD | |
| Right ear | 5.9±4.5 |
| Left ear | 6.1±4.2 |
SD, standard deviation; PTA, pure tone average.
Pure tone average thresholds: pre-exposure vs post-exposure.
| Condition PTA | Pre-exposure | Post-exposure | P-value | ||||
|---|---|---|---|---|---|---|---|
| Mean±SD | Median | Range | Mean±SD | Median | Range | ||
| Right ear (dB) | 5.9±4.5 | 5 | -5 dB to 12.5 dB | 5.8±4.2 | 5 | -3.75 dB to 12.5 dB | 0.85 |
| Left ear (dB) | 6.1±4.2 | 6.25 | -3.75 to 13.75 | 6.4±4 | 6.9 | -2.5 dB to 13.75 dB | 0.33 |
PTA, pure tone average; SD, standard deviation.
*All thresholds in decibels (dB)
°P-values were obtained from the paired, 2-tailed t-test.