Literature DB >> 26774370

Three-dimensional isotropic T2-weighted fast spin-echo (VISTA) ankle MRI versus two-dimensional fast spin-echo T2-weighted sequences for the evaluation of anterior talofibular ligament injury.

H J Park1, S Y Lee1, N H Park2, M H Rho1, E C Chung1, J H Park3, S J Park3.   

Abstract

AIM: To compare the performance of axial images of the ankle joint on three-dimensional (3D) volume isotropic turbo spin echo acquisition (VISTA) with that of two-dimensional (2D) fast spin echo (FSE) T2-weighted images for the diagnosis of anterior talofibular ligament (ATFL) injury.
MATERIALS AND METHODS: This retrospective study included 101 patients who underwent both 2D FSE T2-weighted and 3D VISTA magnetic resonance imaging (MRI) of the ankle. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of both sequences were measured. The anatomical identification score and diagnostic performances of both sequences were evaluated by two radiologists. The diagnostic performances of 3D VISTA and 2D FSE images were analysed in terms of sensitivity, specificity, and accuracy for diagnosing ATFL injury. Surgically or clinically confirmed diagnoses were used as reference standards.
RESULTS: The margin sharpness scores on 3D VISTA were significantly inferior to those of 2D FSE (p<0.001). Other scores (entire length, entire width) were not significantly different between the two imaging methods. The SNRs and CNRs of 3D VISTA were significantly higher than those of 2D FSE (p<0.001). When diagnoses were classified as normal and abnormal, the specificity of the 3D VISTA images for the diagnosis of ATFL injury was 95.7%, significantly superior to 2D FSE (84.3-85.7%). There were no significant differences between 3D VISTA and 2D FSE images in sensitivity or accuracy for diagnosis (p=0.227-1.000), with the exception of accuracy by reader 1 (p=0.039).
CONCLUSION: 3D VISTA imaging has a diagnostic performance comparable to that of 2D FSE for the diagnosis of ATFL injury, although 3D VISTA is inferior to 2D FSE for the evaluation of margin sharpness. Replacing axial and coronal images with 3D VISTA can save imaging time without negatively impacting the diagnostic ability for ATFL injury.
Copyright © 2015 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 26774370     DOI: 10.1016/j.crad.2015.12.007

Source DB:  PubMed          Journal:  Clin Radiol        ISSN: 0009-9260            Impact factor:   2.350


  7 in total

1.  CAIPIRINHA-accelerated 10-min 3D TSE MRI of the ankle for the diagnosis of painful ankle conditions: Performance evaluation in 70 patients.

Authors:  Benjamin Fritz; Susanne Bensler; Gaurav K Thawait; Esther Raithel; Steven E Stern; Jan Fritz
Journal:  Eur Radiol       Date:  2018-07-23       Impact factor: 5.315

2.  Comparison of oblique coronal images in knee of three-dimensional isotropic T2-weighted turbo spin echo MRI versus two-dimensional fast spin echo T2-weighted sequences for evaluation of posterior cruciate ligament injury.

Authors:  Hee J Park; So Y Lee; Seon H Choi; Jin H Ahn; Se J Park; Jai H Park; Eugene Kim
Journal:  Br J Radiol       Date:  2016-09-21       Impact factor: 3.039

3.  Comparison of two-dimensional fast spin echo T2 weighted sequences and three-dimensional volume isotropic T2 weighted fast spin echo (VISTA) MRI in the evaluation of triangular fibrocartilage of the wrist.

Authors:  Hee Jin Park; So Yeon Lee; Kyung A Kang; Eun Young Kim; Hun Kyu Shin; Se Jin Park; Jai Hyung Park; Eugene Kim
Journal:  Br J Radiol       Date:  2018-01-05       Impact factor: 3.039

Review 4.  New Techniques in MR Imaging of the Ankle and Foot.

Authors:  Won C Bae; Thumanoon Ruangchaijatuporn; Christine B Chung
Journal:  Magn Reson Imaging Clin N Am       Date:  2017-02       Impact factor: 2.266

Review 5.  Imaging diagnosis for chronic lateral ankle ligament injury: a systemic review with meta-analysis.

Authors:  Shengxuan Cao; Chen Wang; Xin Ma; Xu Wang; Jiazhang Huang; Chao Zhang
Journal:  J Orthop Surg Res       Date:  2018-05-22       Impact factor: 2.359

6.  Evaluation of diagnostic value and T2-weighted three-dimensional isotropic turbo spin-echo (3D-SPACE) image quality in comparison with T2-weighted two-dimensional turbo spin-echo (2D-TSE) sequences in lumbar spine MR imaging.

Authors:  Jomleh Hossein; Faeghi Fariborz; Rasteh Mehrnaz; Rafiei Babak
Journal:  Eur J Radiol Open       Date:  2018-12-28

7.  Quantitative magnetic resonance imaging (MRI) analysis of anterior talofibular ligament in lateral chronic ankle instability ankles pre- and postoperatively.

Authors:  Wei Liu; Hong Li; Yinghui Hua
Journal:  BMC Musculoskelet Disord       Date:  2017-09-12       Impact factor: 2.362

  7 in total