UNLABELLED: BACKGROUnD: To determine whether the professional Omron HBP-1300 blood pressure (BP) monitor meets American Association for the Advancement of Medical Instrumentation (AAMI) accuracy standards in Chinese children and adults.Entities:
Mesh:
Year: 2016 PMID: 26758197 PMCID: PMC4711064 DOI: 10.1186/s12872-015-0177-z
Source DB: PubMed Journal: BMC Cardiovasc Disord ISSN: 1471-2261 Impact factor: 2.298
Characteristics of the study population
| Mean ± SD | Range | |
|---|---|---|
| Age (years) | 25.9 ± 15.6 | 4-72 |
| Male/female | 49:36:00 | - |
| Pulse (beats/minute) | 82 ± 15 | 51-120 |
| Arm circumference (cm) | 25.8 ± 8.3 | 15.7-47 |
| Systolic blood pressure(mmHg) | 119 ± 28 | 84-210 |
| Diastolic blood pressure(mmHg) | 72 ± 18 | 42-120 |
Cuff sizes and numbers in each size group
| Cuff range (cm) | Cuff Size label | Need 85/2n(numbers of Cuff sizes) or 85/10=9 | Numbers in this study |
|---|---|---|---|
| R 12-18 | SS | 9 | 22 |
| R 17-22 | S | 9 | 13 |
| M 22-32 | M | 9 | 25 |
| L 32-42 | L | 9 | 18 |
| XL 42-50 | XL | 9 | 9 |
Fig. 1Agreement between the test and reference methods for systolic pressure. The X axis in Fig. 1 represents the mean of systolic blood pressure of the HBP-1300 Professional Blood Pressure Monitor and the sphygmomanometer; and the Y axis represents the difference between systolic blood pressure measured by HBP-1300 Professional Blood Pressure Monitor and the systolic blood pressure measured by the sphygmomanometer
Fig. 2Agreement between the test and reference methods for diastolic pressure. The X axis in Fig. 2 represents the mean of diastolic blood pressure of the HBP-1300 Professional Blood Pressure Monitor and the sphygmomanometer; and the Y axis represents the difference between diastolic blood pressure measured by HBP-1300 Professional Blood Pressure Monitor and the diastolic blood pressure measured by the sphygmomanometer