Literature DB >> 24936124

Validation of Datascope Accutorr Plus™ using British Hypertension Society (BHS) and Association for the Advancement of Medical Instrumentation (AAMI) protocol guidelines.

Rajab Ali Khawaja1, Riaz Qureshi1, Abdul Hay Mansure1, Mohammad Emran Yahya1.   

Abstract

OBJECTIVE: To assure the accuracy and reliability of blood pressure measurement by non-invasive blood pressure monitor using Datascope Accutorr Plus™ (Paramus, NJ, USA) against mercury manometer, among adult male participants.
METHOD: Eighty participants from a family physician's office at a teaching hospital were recruited. One hundred and sixty measurements of blood pressure were performed according to BHS technique protocol. STATISTICAL ANALYSIS: Descriptive analysis was done according to the AAMI and BHS protocol guidelines. The limits of agreement between the device and the standard were plotted using the method of Bland and Altman plot.
RESULTS: The mean difference ± SD between the Datascope Acutorr Plus™ and observer was 2.7 ± 5.2 mm Hg and 1.5 ± 3.26 mm Hg for systolic and diastolic blood pressure, respectively. Datascope Acutorr Plus™ obtained A/A grading for both systolic and diastolic blood pressure.
CONCLUSION: Datascope Accutorr Plus™ (Paramus, NJ, USA) satisfies BHS and AAMI validation protocols for both systolic and diastolic BP and may be recommended for everyday use for BP monitoring at home and in clinical use for adult population.

Entities:  

Keywords:  AAMI; BHS; BP devices; Datascope Acutorr plus; Validation

Year:  2010        PMID: 24936124      PMCID: PMC4051388          DOI: 10.1016/j.jsha.2010.03.001

Source DB:  PubMed          Journal:  J Saudi Heart Assoc        ISSN: 1016-7315


  22 in total

1.  Accuracy of the Dinamap Portable Monitor, model 8100: a review of the evidence for accuracy.

Authors: 
Journal:  Blood Press Monit       Date:  1997-12       Impact factor: 1.444

2.  A validation of the Mobil O Graph (version 12) ambulatory blood pressure monitor.

Authors:  C R Jones; K Taylor; P Chowienczyk; L Poston; A H Shennan
Journal:  Blood Press Monit       Date:  2000-08       Impact factor: 1.444

3.  Accuracy of the HoMedics BPA-300, a home blood pressure monitor using the auscultatory method.

Authors:  Steven A Yarows
Journal:  Blood Press Monit       Date:  2007-10       Impact factor: 1.444

4.  Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol.

Authors:  Andrew Coleman; Paul Freeman; Stephen Steel; Andrew Shennan
Journal:  Blood Press Monit       Date:  2006-02       Impact factor: 1.444

5.  Utility of an automatic instrument for blood pressure measurement in children. The Bogalusa Heart Study.

Authors:  W A Wattigney; L S Webber; M D Lawrence; G S Berenson
Journal:  Am J Hypertens       Date:  1996-03       Impact factor: 2.689

6.  Short report: accuracy of the Dinamap portable monitor, model 8100 determined by the British Hypertension Society protocol.

Authors:  E O'Brien; F Mee; N Atkins; K O'Malley
Journal:  J Hypertens       Date:  1993-07       Impact factor: 4.844

7.  Alternative projections of mortality and disability by cause 1990-2020: Global Burden of Disease Study.

Authors:  C J Murray; A D Lopez
Journal:  Lancet       Date:  1997-05-24       Impact factor: 79.321

8.  Validation of the Welch Allyn Spot Vital Signs blood pressure device according to the ANSI/AAMI SP10: 2002. Accuracy and cost-efficiency successfully combined.

Authors:  Bruce S Alpert
Journal:  Blood Press Monit       Date:  2007-10       Impact factor: 1.444

9.  Accuracy of blood pressure measurement by the Dinamap monitor in infants and children.

Authors:  M K Park; S M Menard
Journal:  Pediatrics       Date:  1987-06       Impact factor: 7.124

10.  Accuracy of the Dinamap 1846 XT automated blood pressure monitor.

Authors:  E R Beaubien; C M Card; S E Card; H J Biem; T W Wilson
Journal:  J Hum Hypertens       Date:  2002-09       Impact factor: 3.012

View more
  9 in total

1.  Comparison of blood pressure measurements obtained in the home setting: analysis of the Health Measures at Home Study.

Authors:  Tatiana Nwankwo; Renee Gindi; Te-Ching Chen; Adena Galinsky; Ivey Miller; Ana Terry
Journal:  Blood Press Monit       Date:  2016-12       Impact factor: 1.444

2.  Validation of Omron HBP-1300 professional blood pressure monitor based on auscultation in children and adults.

Authors:  Linghui Meng; Di Zhao; Yan Pan; Wenqing Ding; Qing Wei; Hua Li; Pingjin Gao; Jie Mi
Journal:  BMC Cardiovasc Disord       Date:  2016-01-13       Impact factor: 2.298

3.  Nonadherence to antihypertensive medications and associated factors in general medicine clinics.

Authors:  Mohammed Al Ghobain; H Alhashemi; A Aljama; S Bin Salih; Z Assiri; A Alsomali; Gamal Mohamed
Journal:  Patient Prefer Adherence       Date:  2016-08-01       Impact factor: 2.711

4.  Early Pregnancy Cardiovascular Health and Subclinical Atherosclerosis.

Authors:  Laura Benschop; Sarah Schalekamp-Timmermans; Sara J C Schelling; Eric A P Steegers; Jeanine E Roeters van Lennep
Journal:  J Am Heart Assoc       Date:  2019-07-23       Impact factor: 5.501

5.  Maternal cardiovascular adaptation to twin pregnancy: a population-based prospective cohort study.

Authors:  Maria C Adank; Zoe A Broere-Brown; Romy Gonçalves; M Kamran Ikram; Vincent W V Jaddoe; Eric A P Steegers; Sarah Schalekamp-Timmermans
Journal:  BMC Pregnancy Childbirth       Date:  2020-05-29       Impact factor: 3.007

6.  Design and overview of the Origins of Alzheimer's Disease Across the Life course (ORACLE) study.

Authors:  Sander Lamballais; Maria C Adank; Rowina F Hussainali; Sarah Schalekamp-Timmermans; Meike W Vernooij; Annemarie I Luik; Eric A P Steegers; Mohammad Arfan Ikram
Journal:  Eur J Epidemiol       Date:  2020-12-16       Impact factor: 8.082

7.  Relationship of Sodium Intake With Granulocytes, Renal and Cardiovascular Outcomes in the Prospective EPIC-Norfolk Cohort.

Authors:  Eliane F E Wenstedt; Hessel Peters Sengers; S Matthijs Boekholdt; Kay-Tee Khaw; Nicholas J Wareham; Bert-Jan H van den Born; Liffert Vogt
Journal:  J Am Heart Assoc       Date:  2022-06-22       Impact factor: 6.106

8.  Cardiovascular Risk Factors Track From Mother to Child.

Authors:  Laura Benschop; Sarah Schalekamp-Timmermans; Jeanine E Roeters van Lennep; Vincent W V Jaddoe; Eric A P Steegers; M Kamran Ikram
Journal:  J Am Heart Assoc       Date:  2018-10-02       Impact factor: 5.501

9.  Blood Pressure Monitoring System Using a Two-Channel Ballistocardiogram and Convolutional Neural Networks.

Authors:  Woojoon Seok; Kwang Jin Lee; Dongrae Cho; Jongryun Roh; Sayup Kim
Journal:  Sensors (Basel)       Date:  2021-03-25       Impact factor: 3.576

  9 in total