Efficient system modeling for a small animal PET scanner with tapered DOI detectors.

Mengxi Zhang1, Jian Zhou, Yongfeng Yang, Mercedes Rodríguez-Villafuerte, Jinyi Qi.   

Abstract

A prototype small animal positron emission tomography (PET) scanner for mouse brain imaging has been developed at UC Davis. The new scanner uses tapered detector arrays with depth of interaction (DOI) measurement. In this paper, we present an efficient system model for the tapered PET scanner using matrix factorization and a virtual scanner geometry. The factored system matrix mainly consists of two components: a sinogram blurring matrix and a geometrical matrix. The geometric matrix is based on a virtual scanner geometry. The sinogram blurring matrix is estimated by matrix factorization. We investigate the performance of different virtual scanner geometries. Both simulation study and real data experiments are performed in the fully 3D mode to study the image quality under different system models. The results indicate that the proposed matrix factorization can maintain image quality while substantially reduce the image reconstruction time and system matrix storage cost. The proposed method can be also applied to other PET scanners with DOI measurement.

Entities:  

Mesh:

Year:  2015        PMID: 26682623      PMCID: PMC4874520          DOI: 10.1088/0031-9155/61/2/461

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  17 in total

1.  GATE: a simulation toolkit for PET and SPECT.

Authors:  S Jan; G Santin; D Strul; S Staelens; K Assié; D Autret; S Avner; R Barbier; M Bardiès; P M Bloomfield; D Brasse; V Breton; P Bruyndonckx; I Buvat; A F Chatziioannou; Y Choi; Y H Chung; C Comtat; D Donnarieix; L Ferrer; S J Glick; C J Groiselle; D Guez; P F Honore; S Kerhoas-Cavata; A S Kirov; V Kohli; M Koole; M Krieguer; D J van der Laan; F Lamare; G Largeron; C Lartizien; D Lazaro; M C Maas; L Maigne; F Mayet; F Melot; C Merheb; E Pennacchio; J Perez; U Pietrzyk; F R Rannou; M Rey; D R Schaart; C R Schmidtlein; L Simon; T Y Song; J M Vieira; D Visvikis; R Van de Walle; E Wieërs; C Morel
Journal:  Phys Med Biol       Date:  2004-10-07       Impact factor: 3.609

2.  High-resolution 3D Bayesian image reconstruction using the microPET small-animal scanner.

Authors:  J Qi; R M Leahy; S R Cherry; A Chatziioannou; T H Farquhar
Journal:  Phys Med Biol       Date:  1998-04       Impact factor: 3.609

3.  Improving the intrinsic spatial resolution performance of the continuous miniature crystal element (cMiCE) detector.

Authors:  Tao Ling; Tom K Lewellen; Robert S Miyaoka
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2007-10-26

4.  Modeling and incorporation of system response functions in 3-D whole body PET.

Authors:  Adam M Alessio; Paul E Kinahan; Thomas K Lewellen
Journal:  IEEE Trans Med Imaging       Date:  2006-07       Impact factor: 10.048

5.  Simulation study of spatial resolution and sensitivity for the tapered depth of interaction PET detectors for small animal imaging.

Authors:  Sara St James; Yongfeng Yang; Spencer L Bowen; Jinyi Qi; Simon R Cherry
Journal:  Phys Med Biol       Date:  2009-12-21       Impact factor: 3.609

6.  Performance evaluation of the GE healthcare eXplore VISTA dual-ring small-animal PET scanner.

Authors:  Yuchuan Wang; Jurgen Seidel; Benjamin M W Tsui; Juan J Vaquero; Martin G Pomper
Journal:  J Nucl Med       Date:  2006-11       Impact factor: 10.057

7.  Performance evaluation of a high-sensitivity large-aperture small-animal PET scanner: ClairvivoPET.

Authors:  Tetsuro Mizuta; Keishi Kitamura; Hiroshi Iwata; Yoshiyuki Yamagishi; Atsushi Ohtani; Kazumi Tanaka; Yoshihiro Inoue
Journal:  Ann Nucl Med       Date:  2008-07-04       Impact factor: 2.668

8.  Iterative image reconstruction for positron emission tomography based on a detector response function estimated from point source measurements.

Authors:  Michel S Tohme; Jinyi Qi
Journal:  Phys Med Biol       Date:  2009-05-28       Impact factor: 3.609

9.  Fast, accurate and shift-varying line projections for iterative reconstruction using the GPU.

Authors:  Guillem Pratx; Garry Chinn; Peter D Olcott; Craig S Levin
Journal:  IEEE Trans Med Imaging       Date:  2009-03       Impact factor: 10.048

10.  Characterization of a high-resolution hybrid DOI detector for a dedicated breast PET/CT scanner.

Authors:  Felipe Godinez; Abhijit J Chaudhari; Yongfeng Yang; Richard Farrell; Ramsey D Badawi
Journal:  Phys Med Biol       Date:  2012-05-11       Impact factor: 3.609

View more
  1 in total

1.  A novel depth-of-interaction rebinning strategy for ultrahigh resolution PET.

Authors:  Kyungsang Kim; Joyita Dutta; Andrew Groll; Georges El Fakhri; Ling-Jian Meng; Quanzheng Li
Journal:  Phys Med Biol       Date:  2018-08-14       Impact factor: 3.609

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.