Study of Artifact-Resistive Technology Based on a Novel Dual Photoplethysmography Method for Wearable Pulse Rate Monitors.

Congcong Zhou1, Jingjie Feng1, Jun Hu2, Xuesong Ye3,4.   

Abstract

Pulse rate is one of the major physiological parameters for monitoring of cardiovascular conditions or excise states during daily life. However it is difficult to precisely measure the exact pulse rates as photoplethysmography (PPG) is easy to be affected by motion artifacts. Instead of using accelerometers followed by algorithms such as least mean square (LMS), recursive least square (RLS) and independent component analysis (ICA) or other equipment such as complex laser systems to measure displacement directly, a novel motion artifact estimation method which had lower computational complexity and higher signal dynamic range was studied and implemented, where a differential channel following green and red light PPG channels was applied to reduce the motion artifact caused by displacement of light emitting diode (LED), photo diode (PD) and tissue deformation before the analog signal was converted to digital form. A miniaturized, battery powered wrist worn artifact-resistive pulse rates monitoring system (PRMS) was presented to verify the proposed method. Four kinds of motions were performed and the results showed that the differential channel improved the morphology of the PPG signal and appeared to be artifact resistive during motions through light intensity control and high gain-phase consistency circuit design here.

Entities:  

Keywords:  Artifact-resistive; Motion artifacts; Photoplethysmography; Pulse rates

Mesh:

Year:  2015        PMID: 26645320     DOI: 10.1007/s10916-015-0412-2

Source DB:  PubMed          Journal:  J Med Syst        ISSN: 0148-5598            Impact factor:   4.460


  19 in total

1.  Validation of a device to measure arterial pulse wave velocity by a photoplethysmographic method.

Authors:  Stavros Loukogeorgakis; Rex Dawson; Nirree Phillips; Christopher N Martyn; Stephen E Greenwald
Journal:  Physiol Meas       Date:  2002-08       Impact factor: 2.833

2.  The advantages of wearable green reflected photoplethysmography.

Authors:  Yuka Maeda; Masaki Sekine; Toshiyo Tamura
Journal:  J Med Syst       Date:  2010-05-18       Impact factor: 4.460

3.  Photon migration in the presence of a single defect: a perturbation analysis.

Authors:  S Feng; F A Zeng; B Chance
Journal:  Appl Opt       Date:  1995-07-01       Impact factor: 1.980

4.  Noninvasive measurement of heartbeat, respiration, snoring and body movements of a subject in bed via a pneumatic method.

Authors:  Kajiro Watanabe; Takashi Watanabe; Harumi Watanabe; Hisanori Ando; Takayuki Ishikawa; Keita Kobayashi
Journal:  IEEE Trans Biomed Eng       Date:  2005-12       Impact factor: 4.538

5.  Active noise cancellation using MEMS accelerometers for motion-tolerant wearable bio-sensors.

Authors:  H Harry Asada; Hong-Hui Jiang; Peter Gibbs
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2004

Review 6.  Photoplethysmography and its application in clinical physiological measurement.

Authors:  John Allen
Journal:  Physiol Meas       Date:  2007-02-20       Impact factor: 2.833

7.  Regional differences in the basal and maximal rates of blood flow in the skin.

Authors:  A B HERTZMAN; W C RANDALL
Journal:  J Appl Physiol       Date:  1948-09       Impact factor: 3.531

8.  Relations between cutaneous blood flow and blood content in the finger pad, forearm, and forehead.

Authors:  A B HERTZMAN; W C RANDALL; K E JOCHIM
Journal:  Am J Physiol       Date:  1947-07-01

9.  Photoplethysmography for non-invasive in vivo measurement of bone hemodynamics.

Authors:  Jaime Mateus; Alan R Hargens
Journal:  Physiol Meas       Date:  2012-05-04       Impact factor: 2.833

10.  Effect of probe contact pressure on the photoplethysmographic assessment of conduit artery stiffness.

Authors:  Andris Grabovskis; Zbignevs Marcinkevics; Uldis Rubins; Edgars Kviesis-Kipge
Journal:  J Biomed Opt       Date:  2013-02       Impact factor: 3.170

View more
  5 in total

1.  Angle-selective optical filter for highly sensitive reflection photoplethysmogram.

Authors:  Chan-Sol Hwang; Sung-Pyo Yang; Kyung-Won Jang; Jung-Woo Park; Ki-Hun Jeong
Journal:  Biomed Opt Express       Date:  2017-09-07       Impact factor: 3.732

2.  Validity of wrist-worn consumer products to measure heart rate and energy expenditure.

Authors:  Robert S Thiebaud; Merrill D Funk; Jacelyn C Patton; Brook L Massey; Terri E Shay; Martin G Schmidt; Nicolas Giovannitti
Journal:  Digit Health       Date:  2018-04-13

3.  Heart Rate Measures From Wrist-Worn Activity Trackers in a Laboratory and Free-Living Setting: Validation Study.

Authors:  Andre Matthias Müller; Nan Xin Wang; Jiali Yao; Chuen Seng Tan; Ivan Cherh Chiet Low; Nicole Lim; Jeremy Tan; Agnes Tan; Falk Müller-Riemenschneider
Journal:  JMIR Mhealth Uhealth       Date:  2019-10-02       Impact factor: 4.773

Review 4.  Blood pressure monitoring techniques in the natural state of multi-scenes: A review.

Authors:  Ziyi Liu; Congcong Zhou; Hongwei Wang; Yong He
Journal:  Front Med (Lausanne)       Date:  2022-08-26

5.  An optimal filter for short photoplethysmogram signals.

Authors:  Yongbo Liang; Mohamed Elgendi; Zhencheng Chen; Rabab Ward
Journal:  Sci Data       Date:  2018-05-01       Impact factor: 6.444

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.