Literature DB >> 26631083

[Problems, complications, and emergencies during pacemaker implantation. Importance of access].

Carsten W Israel1, Lucy Ekosso-Ejangue2.   

Abstract

Pacemaker implantation represents a standard procedure with a perceived 100% success rate, without mortality and with extremely rare complications. However, some pacemaker implantations may develop into a very difficult procedure or even be associated with significant complications. Good venous access is crucial and may distinguish between comfortable, successful implantation and futile implantation with severe complications (e.g., pneumo- or hematothorax, venous dissection or perforation, accidental arterial implantation, or air embolism). This review summarizes acute problems and complications during lead implantation and provides tips and hints for prevention and acute reaction during implantation. If these simple precautions are considered, the majority of acute complications during implantation of pacemaker leads can be prevented.

Entities:  

Keywords:  Intraoperative complications; Mortality; Pacemaker; Pneumothorax; Venous anomalies

Mesh:

Year:  2015        PMID: 26631083     DOI: 10.1007/s00399-015-0407-1

Source DB:  PubMed          Journal:  Herzschrittmacherther Elektrophysiol        ISSN: 0938-7412


  6 in total

Review 1.  [Strategies for the avoidance and treatment of complications during pacemaker implantation].

Authors:  C Kolb
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2007-12

2.  Short-term implantation-related complications of cardiac rhythm management device therapy: a retrospective single-centre 1-year survey.

Authors:  Sami Pakarinen; Lasse Oikarinen; Lauri Toivonen
Journal:  Europace       Date:  2010-01       Impact factor: 5.214

3.  Pneumothorax in cardiac pacing: a population-based cohort study of 28,860 Danish patients.

Authors:  Rikke Esberg Kirkfeldt; Jens Brock Johansen; Ellen Aagaard Nohr; Mogens Moller; Per Arnsbo; Jens Cosedis Nielsen
Journal:  Europace       Date:  2012-03-19       Impact factor: 5.214

4.  Association between hospital procedure volume and early complications after pacemaker implantation: results from a large, unselected, contemporary cohort of the German nationwide obligatory external quality assurance programme.

Authors:  Bernd Nowak; Karl Tasche; Linda Barnewold; Günther Heller; Boris Schmidt; Stefano Bordignon; K R Julian Chun; Alexander Fürnkranz; Rajendra H Mehta
Journal:  Europace       Date:  2015-03-01       Impact factor: 5.214

5.  Effects of increasing age onto procedural parameters in pacemaker implantation: results of an obligatory external quality control program.

Authors:  Bernd Nowak; Björn Misselwitz
Journal:  Europace       Date:  2008-11-07       Impact factor: 5.214

Review 6.  Pacemaker implantation in small hospitals: complication rates comparable to larger centres.

Authors:  Bjørn Haug; Kjærsti Kjelsberg; Knut Tore Lappegård
Journal:  Europace       Date:  2011-06-28       Impact factor: 5.214

  6 in total
  1 in total

1.  Inflammatory Pericardial Pseudocyst Secondary to Atrial Myocardial Perforation: A Rare Complication following Transvenous Pacemaker Implantation.

Authors:  Mohammad Mowaswes; Tawfik Khoury; Ziv Lahav; Ashraf Sanduka; Ayelet Shapira-Daniels; Oz M Shapira
Journal:  Case Rep Cardiol       Date:  2018-07-03