Monte Carlo dosimetry of a new (90)Y brachytherapy source.

Wu Junxiang1, You Shihu1, Huang Jing1, Long Fengxiang1, Wang Chengkai2, Wu Zhangwen1, Hou Qing1, Gou Chengjun1.   

Abstract

PURPOSE: In this study, we attempted to obtain full dosimetric data for a new (90)Y brachytherapy source developed by the College of Chemistry (Sichuan University) for use in high-dose-rate after-loading systems.
MATERIAL AND METHODS: The dosimetric data for this new source were used as required by the dose calculation formalisms proposed by the AAPM Task Group 60 and Task Group 149. The active core length of the new (90)Y source was increased to 4.7 mm compared to the value of 2.5 mm for the old (90)Sr/(90)Y source. The Monte Carlo simulation toolkit Geant4 was used to calculate these parameters. The source was located in a 30-cm-radius theoretical sphere water phantom.
RESULTS: The dosimetric data included the reference absorbed dose rate, the radial dose function in the range of 1.0 to 8.0 mm in the longitudinal axis, and the anisotropy function with a θ in the range of 0° to 90° at 5° intervals and an r in the range of 1.0 to 8.0 mm in 0.2-mm intervals. The reference absorbed dose rate for the new (90)Y source was determined to be equal to 1.6608 ± 0.0008 cGy s(-1) mCi(-1), compared to the values of 0.9063 ± 0.0005 cGy s(-1) mCi(-1) that were calculated for the old (90)Sr/(90)Y source. A polynomial function was also obtained for the radial dose function by curve fitting.
CONCLUSIONS: Dosimetric data are provided for the new (90)Y brachytherapy source. These data are meant to be used commercially in after-loading system.

Entities:  

Keywords:  90Y source; Geant4; Monte Carlo; brachytherapy; dosimetry

Year:  2015        PMID: 26622247      PMCID: PMC4663217          DOI: 10.5114/jcb.2015.55362

Source DB:  PubMed          Journal:  J Contemp Brachytherapy        ISSN: 2081-2841


Purpose

Currently, many new types of brachytherapy sources are available. Beta-emitting sources are widely utilized in brachytherapy fields. The dosimetric parameters of beta-emitting sources that have been applied in intravascular brachytherapy (IVBT) have been calculated through some experimental measurements and various simulation programs. For example, Soares et al. obtained a radial dose function and an anisotropy function from their measurements of an old 90Sr/90Y source in A150 plastic, and these authors also presented Monte Carlo results of these dosimetric parameters [1]. The dosimetric parameters of an old 90Sr/90Y source were calculated by Holmes et al. using an experimental method [2]. Furthermore, beta-emitting brachytherapy sources have been applied in radiotherapy for other localized tumors. 90Y microspheres have recently been used to treat unresectable hepatocellular cancer by Nelson et al. [3]. The 32P source model RIC-100 has been used for temporary radiation therapy of the spinal dura and other localized tumors, and dosimetric evaluations have been applied to the source with an MCNP5 Monte Carlo code by Cohen et al. [4]. Recently, a new 90Y source was developed by Wang et al. (Sichuan University) that is a beta-particle-emitting source for treatment of liver cancer. According to the American Association of Physicists in Medicine (AAPM) TG60/TG149 report recommendations [5, 6], the dosimetries at distances of one millimeter from the beta-emitting sources are poorly understood, and better understanding of the dosimetry in the millimeter range will help these sources in the development of clinical practice. All beta-emitting brachytherapy sources for use in clinical practice require obtaining dosimetric parameters based on the TG60/TG149 formalism. In addition, the TG60 and TG149 reports reviewed the physics of the IVBT and provided the dose parameter formalisms for the beta-emitting brachytherapy source. These dose parameters include the reference absorbed dose rate, radial dose function, and anisotropy function. However, the dosimetric parameters of the 90Y source have created new challenges in the field of brachytherapy dosimetry. The experimental measurements are difficult to perform accurately at these close distances due to the large dose gradients and other technical considerations. When an experimental method does not work in the spatial scale of a millimeters or submillimeters, Monte Carlo methods can provide dosimetric data with the required spatial resolution [5, 7, 8]. Monte Carlo methods play important roles in dosimetry and radiotherapy, especially regarding the performance of difficult radiation transport problems [9]. In the present work, we employed the Monte Carlo simulation code Geant4 to derive accurate calculations of the dosimetric parameters of the new 90Y source according to the dose calculation formalisms recommended by AAPM reports TG60/TG149 and the quality assurance (QA) purposes of the new 90Y source. The reference absorbed dose rate and radial dose functions were compared with the results of the old 90Sr/90Y source calculated by Wang et al. [8].

Material and methods

Brachytherapy sources

The source design, materials and sizes are presented in Figure 1. The design of the new 90Y source was provided by the College of Chemistry (Sichuan University), and the tolerances of this new source were ± 0.01 mm provided by manufacturer. The basic sizes and materials of the core and capsules used in the simulation were taken as follows: the composition of the liquid 90Y core was assumed to be H2O with a density of 1.0 g/cm3, and the radioactive material was uniformly distributed in its core. The active core was encapsulated. The capsule material was polyethylene (PE, 86% carbon and 14% hydrogen) with an effective density of 0.94 g/cm3 as shown in Table 1. The active length of the new source was 4.7 mm. At the center of the capsule was a cylinder 4.2 mm in length, 0.8 mm in outer diameter, and 0.5 mm in inner diameter, with two semispherical endings with 0.8 mm external diameters and 0.5 mm inner diameters on the top and the bottom of the cylindrical capsule. The thickness of the PE capsule was 0.15 mm at the cylindrical core.
Fig. 1

Longitudinal view of new 90Y (mm)

Table 1

Chemical composition of materials by percentage mass fraction

MaterialElementsDensity (g/cm3)Composition (% weight)
Polyethylene (PE)C0.940.8563
H0.1437
Longitudinal view of new 90Y (mm) Chemical composition of materials by percentage mass fraction The 90Y source was a pure beta emitter with a 2.7-day half-life, a maximum beta energy of 2.288 MeV, and an average beta energy of 0.934 MeV [5, 10]. The energy spectrum of the 90Y source was complicated and has been described by Cross et al. [11]; the values of the 90Y energy spectrum are presented in Table 2 [11]. Additionally, many other beta-emitting brachytherapy sources are available (90Sr, 32P, and 90Sr/90Y source). 90Sr source with a 28.5 years half-life, an average and a maximum energies of 196 keV and 546 keV, respectively [10]. The 32P source is a pure beta emitter with a 14.3 day half-life, a maximum beta energy of 1.71 MeV and an average beta energy of 0.695 MeV [12]. The 90Sr/90Y source emitted electrons with energies in a continuum that reached up to 2.2 MeV [11]. Compared to the commonly used beta-emitter brachytherapy sources, the new 90Y source had a much shorter half-life than the common beta-emitting sources with the aim of providing increased flexibility and radiation safety. Additionally, the new 90Y source had a greater energy, this high energy makes the radiation more penetrating while having a sharper slope of the depth dose. Because of these features and due to the high-purity decay of betas, 90Y seems a more suitable radionuclide for an effective treatment [10].
Table 2

Nuclear data for the 90Y source

Half-life/dElectron energy (MeV)Electrons per disintegration
2.70.00.3131
0.22880.4101
0.45760.5204
0.68640.5970
0.91520.6215
1.14410.5990
1.37290.5199
1.60170.4097
1.83050.2702
2.05930.1269
2.28800.0001
Nuclear data for the 90Y source

Dose calculation formalism

The dose calculation formalism for the beta-emitting source proposed by the AAPM report TG60 in 1999 was followed. This formalism is described in terms of the polar coordinate system as shown in Figure 2. For a beta-particle-emitting source, the dose calculation formalism is different from that for a photon-emitting source. The air kerma strength (Sk) and dose rate constant in water (Λ) were replaced by the reference absorbed dose rate D(r0, θ0), which has units of cGy s–1 mCi–1. This replacement occurred because the air kerma was only applied to photon emitting sources and did not exist for beta-emitting sources [5]. The dose at any point around the source can be expressed as follows:
Fig. 2

Polar coordinate system for the dose calculation

Polar coordinate system for the dose calculation where L is the active length of the source (L = 4.7 mm) and r is the reference distance, which was defined as 2 mm in this protocol. The reference angle θ0 was specified as p/2. GL(r, θ) is the geometry factor, gL(r) is the radial dose function (the subindex L means the use of the 2D approach instead the 1D), and F(r, θ) is the dose anisotropy function.

Monte Carlo simulation code

We used Monte Carlo Geant4 methods to study the dosimetric parameters of the new 90Y source. The Monte Carlo methods were particle of transport methods that have increasingly been used in radiation therapy, mainly in the development of the brachytherapy field. Popular Monte Carlo methods in this field include Geant4 [13], FLUKA [14], EGS [15], MCNP [16], and PENELOPE [17]. The Geant4 code has been developed and successfully applied in the high, a low energy ranges, and beta-emitting brachytherapy sources [18]. Geant4 (Geant4.9.6.P02 development version) was developed by the European Organization for Nuclear Research (CERN) and is based on the C++ language. This code exploits advanced software-engineering techniques and object-oriented technology to achieve transparency, and requires the user to write each particle and physics model for inclusion in the simulation. This software also provides interaction models for all of the electromagnetic and nuclear processes that are relevant to the transport of brachytherapy. The Geant4 physics model selected for this work was a standard electromagnetic interaction model and employed the function GetTotalEnergyDeposit to obtain the dosimetric data. The Geant4 photon and electron cross sections were based on the EPDL97 and EEDL97 cross section libraries, respectively [19, 20]. The number of electrons Ne generated in each simulation was Ne = 3 × 108, and the cutoff energy of photon and electron was set to 1 keV [21]. In the results for the new 90Y source, all of the secondary particles generated by the source particles (electrons and photons) were completed scored. In this study, we assumed that the new 90Y source was positioned at the center of a spherical liquid water phantom with a 30 cm radius, which was performed to provide dose data for water and to simulate infinite phantom conditions for distances of r < 10 mm. The absorbed dose that was used to calculate the radial dose function and anisotropy function was obtained simultaneously in cylindrical (y, z) and spherical (r, θ) coordinates, and the source axis is along the z axis of the coordinate system as shown in Figure 2. In the coordinate system, y and z are representative of the radial and axial coordinates, respectively. In the spherical coordinates, r is defined as the distance from the center of the active region of the source. When z = 0, we set up the scoring cells at y in the range from 1.0 to 8.0 mm in 0.2-mm intervals to obtain the absorbed dose D (y, z). Subsequently, we used these absorbed doses to obtain the radial dose function. In the spherical coordinate system, we set up the scoring cells at the same distances as those used in the cylindrical coordinate system. Simultaneously, we used θ ranges from 0° to 90° in 5° intervals to obtain the absorbed dose D(r, θ) and then used the anisotropy function equation to calculate the anisotropy function.

Uncertainty analysis

According to AAPM TG-43U1 recommendation, a dosimetric uncertainties (statistical uncertainty and systematic uncertainty) analysis should be performed [22]. We used Monte Carlo Geant4 absorbed dose D (y, z) over the range of 0.0 mm ≤ y ≤ 8.0 mm, 0.0 mm ≤ z ≤ 8.0 mm, and statistical uncertainty equation to obtain statistical uncertainty (type A). For the new 90Y source, the statistical uncertainty of the Monte Carlo dose D (y, z) depended on the axial position z and the radial position y. When z ≤ 4 mm, the statistical uncertainty was within 0.48% for y ≤ 8 mm. With increases in the position z (4 mm ≤ z ≤ 8 mm), the statistical uncertainty was within 2.7% for y ≤ 8 mm. Systematic uncertainty (type B) for the new 90Y source based on Monte Carlo simulation. The component of systematic uncertainty included source geometry, capsule geometry, phantom composition, and cutoff energy as shown in Table 3. 1. Tolerances of ± 0.01 mm on the new 90Y source diameter and capsule thickness were provided by the manufacturer. For +0.01 and –0.01 mm on the new 90Y source diameter, these would cause 0.007% and 0.004% variations in D (2 mm, θ0), respectively. For +0.01 and –0.01 mm on the capsule thickness, these would cause variations in D (2 mm, θ0) of 0.882% and 0.858%, respectively. 2. Systematic uncertainty in water and air phantom compositions and mass densities were calculated to result in D (2 mm, θ0) of 3.862%. 3. Given the 1 keV and 10 keV cutoff energy for D (2 mm, θ0), and the uncertainties was 0.877%.
Table 3

Systematic uncertainty for the new 90Y source

ComponentD (2 mm,θ0)
Type B (%)
Source geometry (±)0.007%
0.004%
Capsule geometry (±)0.882%
0.858%
Phantom composition3.862%
Cutoff energy0.877%
Systematic uncertainty for the new 90Y source

Results

Reference absorbed dose rate

The reference absorbed dose at point r0 = 2 mm for the new 90Y source was obtained by using the Geant4 code and it was found to be equal to 4.4887 × 10–10 Gy/electron. The reference absorbed dose rate D(r0, θ0) of the new 90Y source in water was 1.6608 cGy s–1 mCi–1 as shown in Table 4, which also illustrates the values that were calculated for the old 90Sr/90Y source by Wang et al. [8] using a EGS4 Monte Carlo code (the composition of the capsule was SS304 stainless steel and the diameter and height of the active core were 0.56 and 2.5 mm, respectively). The value of D(r0, θ0) for the new 90Y source was 45% higher than that calculated for the old 90Sr/90Y source and these percentage differences were nearly equal to the percentage differences observed for the energies, sizes, and materials for the two sources. The new 90Y source had a greater energy than the old 90Sr/90Y source and this high energy makes the radiation more penetrating.
Table 4

Absorbed dose rate at reference

SourceD(r0, θ0)Unit
New 90Y1.6608 ± 0.0008cGy s–1 mCi–1
Old 90Sr/90Y0.9063 ± 0.0005cGy s–1 mCi–1
Absorbed dose rate at reference

Radial dose functions

The two-dimensional dose distribution D(r0, θ0) was calculated for the new 90Y source in water. This dose distribution was used to derive the radial dose functions of the source. Table 5 shows the radial dose function gL(r) at the radial distances over the range of 1.0 to 8.0 mm (the subindex L means the use of the 2D approach instead the 1D with active length L = 4.7 mm). This table illustrates that, at the radial distances far from the source (r ≥ 8.0 mm), the radial dose functions had lower values.
Table 5

Comparison of the radial dose function from this work to calculated values from other author

r (mm) gL(r) (this work)Ref. 8
1.01.13681.0990
1.21.1030
1.41.0862
1.51.06001.0660
1.61.0672
1.81.0337
2.01.00001.000
2.20.9775
2.40.9521
2.50.92450.9070
2.60.9002
2.80.8750
3.00.83810.7990
3.20.8015
3.40.7628
3.60.7219
3.80.6769
4.00.63310.5640
4.20.5872
4.40.5469
4.60.5075
4.80.4695
5.00.43230.3510
6.00.26310.1890
7.00.12720.0850
8.00.04540.0300
Comparison of the radial dose function from this work to calculated values from other author In Figure 3, a fitted polynomial function of the new 90Y source and a comparison of the radial dose functions of the new 90Y and the old 90Sr/90Y source calculated according to EGS4 from Wang et al. [8], and the differences between two sources at right ordinate axis are presented. The comparison between two sources revealed the following: (a) at the radial distances of r ≤ 2.0 mm, our calculated gL(r) values agreed with the calculations of Wang et al.'s within 3.3%; (b) for 2.0 mm ≤ r ≤ 4.0 mm, differences between our calculations and those of Wang et al. were greater than 11%; and (c) for 4.0 mm ≤ r ≤ 8.0 mm, our calculations were greater than those of Wang et al. by up to 34%. These results confirm that the different active core and geometry of the source dose significantly affected the values of the radial dose functions. These differences were due to the average energies, sizes, and materials of the two sources (our new 90Y source had a greater energy than Wang et al.'s old 90Sr/90Y source according to their energy spectrum and the values of the reference absorbed dose rate; the length of the active core was 4.7 mm rather than 2.5 mm and the material of the capsule was PE, rather than stainless steel).
Fig. 3

Radial dose functions calculated for the new 90Y source and the old 90Sr/90Y source. The line is the polynomial fit of order 5 for the new 90Y source

Radial dose functions calculated for the new 90Y source and the old 90Sr/90Y source. The line is the polynomial fit of order 5 for the new 90Y source The radial dose function gL(r), also fit over the range of 1.0 mm ≤ r ≤ 8.0 mm using a fifth-order polynomial function as presented in Figure 3. The polynomial function is shown in Eq. (2). The values of the fitted parameters were as follows: R-squared (R2) 0.99992, a0 = –0.00063, a1 = 0.05057, a2 = –0.37012, a3 = 6.84190, a4 = –15.69895, a5 = 8.42267, and a6 = 1.7092. The overall accuracies of the fits were excellent, and the fits based on the polynomial exhibited maximum and mean deviations of less than 0.56% and 0.23%, respectively. g(r) = (a0r–2 + a1r–1 + a2 + a3r + a4r2 + a5r3)e–a     (2) where r is the radial distance from the center of the source for θ = 0°.

Anisotropy functions

In Table 6, the full data for the anisotropy functions F(r, θ) are presented for radial distances of r = 1.0-8.0 mm in 0.2-mm increments and at polar angles of θ= 0°-90° in 5°-increments. The anisotropy functions for the new 90Y source are shown graphically for radial distances of r = 3.0, 4.4, 5.0, 6.0, and 7.0 mm in Figure 4. This figure shows the anisotropy function of the new 90Y source for most of the points, F(r0, θ0) > 1.0.
Table 6

Anisotropy function calculated for the new 90Y source

Angle (°)Radial distance (mm)
1.01.21.41.61.82.02.22.42.62.8
00.8260.978
50.9451.018
101.0541.0691.075
151.0911.0851.1241.0811.0671.0951.108
201.1251.0891.0821.0761.0771.0591.0671.1041.118
251.0691.0651.0501.0501.0531.0611.0501.0651.1021.114
301.0411.0361.0331.0361.0411.0501.0421.0571.0931.105
351.0181.0241.0221.0271.0321.0411.0331.0481.0811.090
401.0091.0151.0151.0181.0241.0331.0251.0391.0691.077
451.0001.0041.0041.0071.0111.0201.0101.0201.0481.050
500.9971.0011.0001.0031.0060.9690.8241.0111.0371.039
550.9940.9950.9880.9891.0691.0451.0221.0211.0361.026
601.0000.9980.9890.9901.0741.0421.0181.0181.0331.024
650.9940.9910.9971.0621.0351.0321.0131.0141.0301.022
700.9930.9900.9961.0621.0281.0241.0041.0051.0211.012
750.9960.9961.0001.0701.0291.0241.0031.0041.0201.011
800.9981.0101.0201.0811.0391.0351.0161.0181.0351.027
851.0411.0661.0681.1411.0851.0781.0601.0631.0821.075
901.0001.0001.0001.0001.0001.0001.0001.0001.0001.000
Angle (°)Radial distance (mm)
3.03.23.43.63.84.04.24.44.64.8
01.0311.0651.0831.0991.1291.1491.1871.2261.2571.277
51.0581.0821.1031.1211.1471.1681.1941.2211.2461.269
101.0941.1161.1321.1491.1721.1931.2151.2411.2651.287
151.1231.1401.1541.1711.1901.2141.2311.2531.2671.289
201.1301.1431.1531.1671.1881.2061.2251.2431.2611.275
251.1271.1361.1461.1591.1771.1931.2111.2301.2441.258
301.1141.1251.1341.1451.1601.1761.1941.2091.2221.237
351.0991.1081.1171.1271.1411.1581.1731.1861.1971.206
401.0831.0911.0971.1081.1211.1341.1441.1601.1701.177
451.0531.0581.0611.0671.0761.0851.0941.1001.1041.105
501.0411.0431.0441.0471.0571.0661.0701.0771.0791.076
551.0161.0091.0031.0001.0031.0061.0091.0121.0131.011
601.0151.0081.0031.0011.0041.0081.0121.0171.0181.019
651.0141.0081.0031.0011.0041.0071.0101.0141.0151.015
701.0040.9980.9920.9910.9930.9971.0011.0041.0041.002
751.0020.9970.9920.9890.9920.9950.9991.0021.0020.998
801.0191.0131.0071.0061.0101.0131.0161.0191.0171.015
851.0671.0631.0591.0581.0641.0691.0731.0761.0751.073
901.0001.0001.0001.0001.0001.0001.0001.0001.0001.000
Angle (°)Radial distance (mm)
5.05.25.45.65.86.06.26.46.66.8
01.3141.3491.3851.4221.4841.5231.5871.6571.6981.793
51.2921.3161.3581.3751.4091.4541.5091.5651.6481.750
101.3091.3301.3591.3881.4191.4601.5071.5671.6421.718
151.3061.3271.3521.3791.4041.4451.4901.5451.5981.673
201.2911.3081.3281.3501.3791.4101.4541.5021.5561.622
251.2741.2861.3051.3201.3401.3681.4121.4531.5111.571
301.2451.2561.2681.2821.3011.3241.3621.4041.4461.498
351.2141.2201.2281.2381.2521.2731.3061.3451.3861.434
401.1841.1881.1921.1951.2031.2621.2801.2901.3151.365
451.1051.1001.1001.0981.1031.1151.1291.1451.1681.194
501.0731.0651.0661.0601.0661.0731.0851.1021.1151.142
551.0101.0081.0081.0071.0111.0191.0341.0511.0711.097
601.0201.0161.0151.0151.0181.0291.0461.0631.0861.112
651.0131.0071.0051.0031.0071.0141.0291.0461.0641.090
700.9990.9930.9900.9880.9910.9981.0131.0261.0431.067
750.9950.9900.9880.9860.9900.9961.0101.0241.0401.059
801.0121.0071.0051.0051.0081.0131.0261.0401.0541.072
851.0721.0681.0691.0681.0741.0801.0931.1071.1181.135
901.0001.0001.0001.0001.0001.0001.0001.0001.0001.000
Angle (°)Radial distance (mm)
7.07.27.47.67.88.0
01.9282.0862.2102.3892.6402.913
51.8461.9242.0702.2472.4412.711
101.8101.9202.0502.1772.3582.582
151.7551.8481.9782.0992.2822.490
201.7001.8061.8982.0152.1732.365
251.6461.7261.8181.9402.0692.249
301.5521.6381.7111.8181.9272.098
351.4901.5591.6261.7091.8051.919
401.4101.4641.5221.5971.6631.781
451.2241.2651.2951.3381.3931.472
501.1551.1821.2161.2521.2931.357
551.1271.1651.2061.2611.3231.437
601.1441.1791.2201.2751.3361.458
651.1151.1511.1831.2321.2871.400
701.0871.1181.1451.1901.2371.347
751.0791.1071.1291.1691.2131.331
801.0901.1141.1341.1711.2141.328
851.1521.1771.1941.2291.2781.416
901.0001.0001.0001.0001.0001.000
Fig. 4

Anisotropy function calculated for the new 90Y source for selected distances

Anisotropy function calculated for the new 90Y source for selected distances Anisotropy function calculated for the new 90Y source An along-away table for quality assurance (QA) purposes is presented in Table 7. For these tables, the transverse direction is equal to the z axis and the longitudinal direction is equal to the y axis. The new 90Y source is symmetric about the transverse and vertical axes. Therefore, the QA values were tabulated over the range of 0.5 mm ≤ y ≤ 8.0 mm and 0.0 mm ≤ z ≤ 8.0 mm.
Table 7

QA away-along data (cGy min–1 Bq–1) for the new 90Y source

z/mm
y/mm 0.00.51.01.52.02.53.03.54.0
0.536.832136.582435.558033.036726.171413.51736.17623.34021.9918
1.014.990914.757213.997212.561010.08446.90744.28172.66751.7099
1.58.19148.01937.51106.66955.50924.23422.93532.00791.4106
2.04.98254.84834.53344.03383.41252.79252.05641.49641.1129
2.53.19243.09492.89552.59302.23651.90241.44991.09930.8513
3.02.09882.03931.90321.71721.50411.31171.02280.79790.6358
3.51.39421.37101.26911.15221.02170.90710.71870.57120.4624
4.00.95910.91860.84840.77510.69530.62050.49890.40280.3279
4.50.62100.61450.56480.51850.46980.41840.34060.27840.2279
5.00.43090.41010.37310.34310.31350.27780.22770.18760.1538
5.50.27020.27110.24280.22320.20540.18000.14820.12300.1006
6.00.18490.17460.15340.14170.13110.11260.09320.07770.0623
6.50.10420.10750.09320.08700.08060.06690.05610.04690.0366
7.00.06770.06310.05410.05090.04740.03760.03200.02680.0202
7.50.03260.03480.02980.02820.02640.01990.01710.01440.0104
8.00.01950.01780.01510.01460.01360.00970.00840.00710.0048
z/mm
y/mm4.55.05.56.06.57.07.58.0
0.51.25190.81450.53930.35410.23130.14820.09360.0574
1.01.08430.71460.47540.29920.19380.12340.07220.0437
1.50.91750.61370.42380.26480.17110.11190.06430.0385
2.00.74820.51340.35890.22710.14780.09690.05580.0333
2.50.58710.41310.29150.18610.12250.08050.04650.0277
3.00.44750.32050.22750.14690.09760.06410.03710.0221
3.50.33110.24120.17260.11140.07430.04910.02830.0167
4.00.23800.17580.12620.08180.05470.03570.02050.0122
4.50.16620.12380.08880.05760.03850.02490.01440.0084
5.00.11270.08410.05970.03900.02600.01650.00940.0054
5.50.07390.05520.03840.02520.01670.01030.00590.0033
6.00.04620.03450.02330.01540.01020.00610.00340.0019
6.50.02750.02030.01340.00880.00580.00330.00180.0010
7.00.01520.01130.00720.00470.00300.00170.00090.0005
7.50.00780.00580.00360.00230.00150.00080.00040.0002
8.00.00370.00270.00160.00100.00060.00030.00020.0001
QA away-along data (cGy min–1 Bq–1) for the new 90Y source

Conclusions

Before any of these sources can be used in clinical practice, their dosimetric parameters must be calculated by using the Monte Carlo code, which is very difficult to perform accurately with an experimental method. Using our Monte Carlo simulation code Geant4, the dosimetric parameters for a new 90Y source in water have been calculated based on the dose calculation formalism recommended by the AAPM TG60 and TG149. The reference absorbed dose rate, radial dose functions, anisotropy functions, and away-along QA table were calculated for the new 90Y source, and the reference absorbed dose rate and radial dose functions were compared to the old 90Sr/90Y source. Differences were observed between these two sources in terms of their energies, sizes, and materials. The new 90Y source had a short half-life and high energy compared with the old 90Sr/90Y source. The parameters of this study were presented in the forms of both a graph and a table. All of the statistical uncertainties at distances for z ≤ 8.0 mm and y ≤ 8.0 mm were below 2.7%. Our purpose in this work is to obtain the dosimetric parameters for the new 90Y source to be used in commercially available afterloading system in clinical practice. At present, the new 90Y brachytherapy source has been applied in Fonics after-loading system developed in China. According to lots of preclinical experiments for the new 90Y brachytherapy source by our team, this new source is well suited for radiation therapy of the liver cancer. Furthermore, these dosimetric parameters of the new 90Y source did not previously exist.
  14 in total

1.  A Monte Carlo calculation of dosimetric parameters of 90Sr/90Y and 192Ir SS sources for intravascular brachytherapy.

Authors:  R Wang; X A Li
Journal:  Med Phys       Date:  2000-11       Impact factor: 4.071

2.  Intravascular brachytherapy physics: report of the AAPM Radiation Therapy Committee Task Group no. 60. American Association of Physicists in Medicine.

Authors:  R Nath; H Amols; C Coffey; D Duggan; S Jani; Z Li; M Schell; C Soares; J Whiting; P E Cole; I Crocker; R Schwartz
Journal:  Med Phys       Date:  1999-02       Impact factor: 4.071

3.  Characterization of a high-dose-rate 90Sr-90Y source for intravascular brachytherapy by using the Monte Carlo code PENELOPE.

Authors:  J Asenjo; J M Fernández-Varea; A Sánchez-Reyes
Journal:  Phys Med Biol       Date:  2002-03-07       Impact factor: 3.609

4.  Dose distribution to spinal structures from intrathecally administered yttrium-90.

Authors:  George Mardirossian; Michael Hall; Joseph Montebello; Patrick Stevens
Journal:  Phys Med Biol       Date:  2005-12-15       Impact factor: 3.609

5.  Experimental determination of the radial dose function of 90Sr/90Y IVBT sources.

Authors:  Shannon M Holmes; Larry A DeWerd; John A Micka
Journal:  Med Phys       Date:  2006-09       Impact factor: 4.071

6.  EGS4 dosimetry calculations for cylindrically symmetric brachytherapy sources.

Authors:  R Wang; R S Sloboda
Journal:  Med Phys       Date:  1996-08       Impact factor: 4.071

7.  Calibration and characterization of beta-particle sources for intravascular brachytherapy.

Authors:  C G Soares; D G Halpern; C K Wang
Journal:  Med Phys       Date:  1998-03       Impact factor: 4.071

8.  Dose calculation formalisms and consensus dosimetry parameters for intravascular brachytherapy dosimetry: recommendations of the AAPM Therapy Physics Committee Task Group No. 149.

Authors:  Sou-Tung Chiu-Tsao; Dennis R Schaart; Christopher G Soares; Ravinder Nath
Journal:  Med Phys       Date:  2007-11       Impact factor: 4.071

9.  Post-mortem considerations of Yttrium-90 90Y microsphere therapy procedures.

Authors:  Kevin Nelson; Paul E Vause; Polina Koropova
Journal:  Health Phys       Date:  2008-11       Impact factor: 1.316

10.  32P liquid sources--comparison of the effectiveness of postangioplasty versus poststenting intravascular brachytherapy in hypercholesterolemic rabbits. Adjunctly implanted titanium stent does not attenuate the effect of endovascular irradiation.

Authors:  Piotr Walichiewicz; Barbara Petelenz; Krzysztof Wilczek; Wojciech Jacheć; Jerzy Jochem; Andrzej Tomasik; Dariusz Lange; Jan Wodniecki
Journal:  Cardiovasc Radiat Med       Date:  2003 Apr-Jun
View more
  1 in total

1.  Monte Carlo dosimetry for a new 32P brachytherapy source using FLUKA code.

Authors:  Razieh Rajabi; Payvand Taherparvar
Journal:  J Contemp Brachytherapy       Date:  2019-02-28
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.