Calibration and characterization of beta-particle sources for intravascular brachytherapy.

C G Soares1, D G Halpern, C K Wang.   

Abstract

The calibration of a catheter-based system to be used for therapeutic radiation treatment to prevent restenosis following interventional coronary procedures is described. The primary dosimetry was performed ionometrically using an extrapolation chamber equipped with a 1-mm diameter collecting electrode to measure absorbed dose in tissue equivalent plastic at a depth of 2 mm. These results are compared with measurements with radiochromic dye film, which is also used to characterize sources for axial and trans-axial uniformity, and to determine dose distributions at various depths. A protocol for dose calculation based on that of AAPM TG43 is suggested for these sources, and examples of its use are given for the calculation of the enhancement effect on dose rate from a single seed source due to neighboring seeds. Monte Carlo calculations were also performed to validate the measured results.

Entities:  

Mesh:

Year:  1998        PMID: 9547501     DOI: 10.1118/1.598204

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Monte Carlo dosimetric parameter study of a new (32)P brachytherapy source.

Authors:  Junxiang Wu; Jing Huang; Fengxiang Long; Chengkai Wang; Zhangwen Wu; Chengjun Gou
Journal:  Br J Radiol       Date:  2016-04-04       Impact factor: 3.039

Review 2.  [Intracoronary brachytherapy with strontium/yttrium-90. Initial experiences in Germany].

Authors:  S Silber; P von Rottkay; A Gielow; A Schneider; A Bauer; H Schöfer
Journal:  Herz       Date:  1998-09       Impact factor: 1.443

3.  Monte Carlo dosimetry of a new (90)Y brachytherapy source.

Authors:  Wu Junxiang; You Shihu; Huang Jing; Long Fengxiang; Wang Chengkai; Wu Zhangwen; Hou Qing; Gou Chengjun
Journal:  J Contemp Brachytherapy       Date:  2015-10-30
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.