BACKGROUND: The Streamlined Liner of the Pharynx Airway (SLIPA), a type of supraglottic airway, has a non-inflatable cuff that softens at body temperature to fit the laryngeal structure. We investigated whether pre-warming of SLIPA to body temperature may improve insertion parameters.Entities:
Mesh:
Year: 2015 PMID: 26589142 PMCID: PMC4653944 DOI: 10.1186/s12871-015-0151-4
Source DB: PubMed Journal: BMC Anesthesiol ISSN: 1471-2253 Impact factor: 2.217
Demographic data
| Group R ( | Group W ( | P | |
|---|---|---|---|
| Age (years) | 41.3 ± 12.5 | 37.4 ± 11.6 | 0.130 |
| Sex (male/female) | 16/28 | 18/27 | 0.724 |
| Height (cm) | 163 (159–169) | 162 (159–171) | 0.583* |
| Weight (kg) | 64.2 ± 11.3 | 63.6 ± 10.8 | 0.800 |
| ASA (1/2/3/4) | |||
| Mallampati score (1/2/3/4) | 9/11/13/11 | 9/16/6/14 | 0.278 |
| Duration of anesthesia (min) | 78.7 ± 31.3 | 72.9 ± 29.1 | 0.361 |
Values are expressed as means ± standard deviations, medians (25th percentile–75th percentile), or absolute numbers
*The Mann–Whitney U-test was used, and the data were expressed as medians (25th percentile–75th percentile) because of the abnormal distribution
Fig. 1CONSORT flow diagram. The progress of patients through the trial is shown. Group R: the Streamlined Liner of the Pharynx Airway (SLIPA) was kept at room temperature; Group W: the SLIPA was pre-warmed to 37 ° C
Fig. 2Oropharyngeal leak pressure (OLP). OLP increased gradually over time in both groups. However, the mean OLP in Group W was significantly higher than that in Group R from the point of insertion until 3 min after insertion. Group R: the Streamlined Liner of the Pharynx Airway (SLIPA) was kept at room temperature; Group W: the SLIPA was pre-warmed to 37 ° C. Values are expressed as the mean ± SE. *P < 0.05
Fig. 3Insertion time. The insertion time was faster in Group W (3.60 s [3.15–4.06 s]) than in Group R. Box-and-whisker plot (median, interquartile range, and range) of the insertion time in Group R and Group W. Group R: the Streamlined Liner of the Pharynx Airway (SLIPA) was kept at room temperature; Group W: the SLIPA was pre-warmed to 37 ° C. *P < 0.001
Airway observations
| Group R ( | Group W ( | P | |
|---|---|---|---|
| Blood staining (n) | 5 | 4 | 0.699 |
| Regurgitation (n) | 0 | 0 | NA |
| Attempt (1/2; n) | 44/0 | 45/0 | NA |
| Difficulty of insertion (easy/normal/hard) | 7/28/9 | 21/21/3 | 0.004* |
Values are expressed as absolute numbers
*P < 0.05 between groups
Fig. 4Postoperative throat pain (visual analog scale [VAS]). Postoperative throat pain, determined using the VAS, was lower in Group W than in Group R. Box-and-whisker plot (median, interquartile range, and range) of the VAS in Group R and Group W. Group R: the Streamlined Liner of the Pharynx Airway (SLIPA) was kept at room temperature; Group W group: the SLIPA was pre-warmed to 37 ° C. *P = 0.006