Influence of bone density and implant drill diameter on the resulting axial force and temperature development in implant burs and artificial bone: an in vitro study.

Stephan Christian Möhlhenrich1, Mustapha Abouridouane2, Nicole Heussen3, Ali Modabber4, Fritz Klocke2, Frank Hölzle4.   

Abstract

PURPOSE: The aim of this study was to determine how the bone density affects the temperature development in artificial bone and drill.
METHODS: Ten single drills with diameters of 2.2, 2.8, 3.5, and 4.2 mm were used on four artificial bone blocks (density I-IV), with constant speed and external irrigation. Temperature measurement in blocks and drills was done by infrared camera. The resultant axial force was measured, and light microscopic examinations of the drills were performed before and after preparation.
RESULTS: The block density has a greater influence on resulting axial force than the drill diameter (D1 = 2.2 mm, 4.11 ± 0.64 N; 4.2 mm, 9.69 ± 0.78 N vs. D4 = 2.2 mm, 0.5 ± 0.18 N; 4.2 mm, 1.23 ± 0.08 N). For the narrowest drill, a decrease in bone density caused a significant temperature increase in the bone and drill. However, for the thickest drill, no thermal differences were found in the bone but were seen in the drill itself (D1 = 2.8 mm vs. D4 = 2.8 mm; bone p < 0.0001, drill p < 0.0001; D1 = 4.2 mm vs. D4 = 4.2 mm; bone p = 0.5366, drill p = 0.0411). An increase in the drill diameter in the highest bone density led to a significant thermal increase in the bone and drill. However, for the lowest bone density, thermal changes were observed only in the bone (D1 = 2.8 mm vs. D1 = 4.2 mm; bone p < 0.0001, drill p < 0.0001; D4 = 2.8 mm vs. D4 = 4.2 mm; bone p < 0.0102, drill p = 0.1784).
CONCLUSIONS: Thermal development depends on bone density with increasing density causing a temperature rise. However, this effect is reduced with increasing drill diameter. This may be important with regard to bone reactions and also in terms of tool wear.

Entities:  

Keywords:  Axial load; Drill wear; Heat generation; Implant site preparation; Infrared thermography

Mesh:

Year:  2015        PMID: 26584948     DOI: 10.1007/s10006-015-0536-z

Source DB:  PubMed          Journal:  Oral Maxillofac Surg        ISSN: 1865-1550


  32 in total

1.  An in vitro study of temperature changes in type 4 bone during implant placement: bone condensing versus bone drilling.

Authors:  Tijana Misic; Aleksa Markovic; Aleksandar Todorovic; Snjezana Colic; Scepanovic Miodrag; Biljana Milicic
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2010-12-13

2.  Influence of the surgical technique and surface roughness on the primary stability of an implant in artificial bone with a density equivalent to maxillary bone: a laboratory study.

Authors:  Afsheen Tabassum; Gert J Meijer; Johannes G C Wolke; John A Jansen
Journal:  Clin Oral Implants Res       Date:  2009-04       Impact factor: 5.977

3.  Influence of surgical technique and surface roughness on the primary stability of an implant in artificial bone with different cortical thickness: a laboratory study.

Authors:  Afsheen Tabassum; Gert J Meijer; Johannes G C Wolke; John A Jansen
Journal:  Clin Oral Implants Res       Date:  2010-02       Impact factor: 5.977

4.  Bone classification, training keys to implant success.

Authors:  C E Misch
Journal:  Dent Today       Date:  1989-05

5.  An in vitro evaluation of heat production during osteotomy preparation for dental implants with compressive osteotomes.

Authors:  Alessandro Quaranta; Sebastiano Andreana; Liana Spazzafumo; Matteo Piemontese
Journal:  Implant Dent       Date:  2013-04       Impact factor: 2.454

6.  The effect of heat on bone regeneration: an experimental study in the rabbit using the bone growth chamber.

Authors:  R A Eriksson; T Albrektsson
Journal:  J Oral Maxillofac Surg       Date:  1984-11       Impact factor: 1.895

7.  Thermal changes and drill wear in bovine bone during implant site preparation. A comparative in vitro study: twisted stainless steel and ceramic drills.

Authors:  Natália Oliveira; Fernando Alaejos-Algarra; Javier Mareque-Bueno; Eduard Ferrés-Padró; Federico Hernández-Alfaro
Journal:  Clin Oral Implants Res       Date:  2011-08-02       Impact factor: 5.977

8.  Effect of implant drill characteristics on heat generation in osteotomy sites: a pilot study.

Authors:  Hyun Jun Oh; Ulf Me Wikesjö; Ho-Seong Kang; Young Ku; Tae-Gwan Eom; Ki-Tae Koo
Journal:  Clin Oral Implants Res       Date:  2010-12-09       Impact factor: 5.977

9.  Drilling- and withdrawing-related thermal changes during implant site osteotomies.

Authors:  Georg D Strbac; Katharina Giannis; Ewald Unger; Martina Mittlböck; Christoph Vasak; Georg Watzek; Werner Zechner
Journal:  Clin Implant Dent Relat Res       Date:  2013-05-28       Impact factor: 3.932

10.  The effect of load on heat production, thermal effects and expenditure of time during implant site preparation - an experimental ex vivo comparison between piezosurgery and conventional drilling.

Authors:  Florian Stelzle; Carsten Frenkel; Max Riemann; Christian Knipfer; Philipp Stockmann; Emeka Nkenke
Journal:  Clin Oral Implants Res       Date:  2012-11-27       Impact factor: 5.977

View more
  4 in total

1.  Evaluation of accuracy in implant site preparation performed in single- or multi-step drilling procedures.

Authors:  Nadine Marheineke; Uta Scherer; Martin Rücker; Constantin von See; Björn Rahlf; Nils-Claudius Gellrich; Marcus Stoetzer
Journal:  Clin Oral Investig       Date:  2017-12-17       Impact factor: 3.573

2.  Comparison of Maximum Heat Generation during Implant Site Preparation between Single and Gradual Drilling Protocols in Artificial D1 Bone Blocks: An In Vitro Study.

Authors:  Tammam Koutiech; Omar Ahmad Heshmeh; Kamal Alkerdi; Johnny Toumi; Laith Al Sabek
Journal:  Int J Dent       Date:  2022-06-18

3.  The effects of irrigation volume to the heat generation during implant surgery.

Authors:  A Sindel; Ö Dereci; M Hatipoğlu; M-A Altay; Ö Özalp; A Öztürk
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2017-07-01

4.  Ridge augmentation-The new field of computerized guided surgery: A technical note for minimal-invasive bone splitting.

Authors:  Vasilios Alevizakos; Gergo Mitov; Marcus Schiller; Constantin von See
Journal:  Clin Case Rep       Date:  2021-03-20
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.