Literature DB >> 23442602

An in vitro evaluation of heat production during osteotomy preparation for dental implants with compressive osteotomes.

Alessandro Quaranta1, Sebastiano Andreana, Liana Spazzafumo, Matteo Piemontese.   

Abstract

BACKGROUND: To assess heat production using osteotomes under conditions simulating implant placement in D3, D4 bone.
MATERIAL AND METHODS: Implant osteotomes were tested (Winsix, Biosafin, Italy). Site preparations were performed on porcine ribs through a compressive fashion. The ribs were partially immersed in a custom-made water bath/water pump system that maintained the baseline temperature at 36 °C. Temperatures generated at different depths (2, 5, and 9 mm) during a series of 25 osteotomies were measured using 3 thermocouples connected to a digital thermometer.
RESULTS: The mean temperatures never exceeded 37 °C and were all significantly lower than those reported during preparation with standard implant drills. Statistical evaluation of the temperature during implant site preparation showed slight significant variations between the baseline values and those of the different depth preparations.
CONCLUSIONS: Within the limits of this study, the production of heat during implant preparation using osteotomes shows significant variations at different depths. However, these variations are not clinically relevant because they never increased over the values that negatively affect bone that may jeopardize osseointegration.

Entities:  

Mesh:

Year:  2013        PMID: 23442602     DOI: 10.1097/ID.0b013e318285984c

Source DB:  PubMed          Journal:  Implant Dent        ISSN: 1056-6163            Impact factor:   2.454


  3 in total

1.  Comparison of heat production and bone architecture changes in the implant site preparation with compressive osteotomes, osseodensification technique, piezoelectric devices, and standard drills: an ex vivo study on porcine ribs.

Authors:  Nishith Bhargava; Vittoria Perrotti; Vito Carlo Alberto Caponio; Victor Haruo Matsubara; Diana Patalwala; Alessandro Quaranta
Journal:  Odontology       Date:  2022-07-19       Impact factor: 2.885

2.  Comparison of Maximum Heat Generation during Implant Site Preparation between Single and Gradual Drilling Protocols in Artificial D1 Bone Blocks: An In Vitro Study.

Authors:  Tammam Koutiech; Omar Ahmad Heshmeh; Kamal Alkerdi; Johnny Toumi; Laith Al Sabek
Journal:  Int J Dent       Date:  2022-06-18

3.  Influence of bone density and implant drill diameter on the resulting axial force and temperature development in implant burs and artificial bone: an in vitro study.

Authors:  Stephan Christian Möhlhenrich; Mustapha Abouridouane; Nicole Heussen; Ali Modabber; Fritz Klocke; Frank Hölzle
Journal:  Oral Maxillofac Surg       Date:  2015-11-20