Literature DB >> 26511081

Falsely Elevated Glucose Concentrations in Peritoneal Dialysis Patients Using Icodextrin.

Kübra Dogan1, Damla Kayalp1, Gözde Ceylan1, Alper Azak2, Mehmet Senes1, Murat Duranay2, Dogan Yucel3.   

Abstract

BACKGROUND: Peritoneal dialysis (PD) is used as an alternative to hemodialysis in end-stage renal disease (ESRD). Icodextrin has been used as a hyperosmotic agent in PD. The aim of the study was to assess two different point-of-care testing (POCT) glucose strips, affected and not affected by icodextrin, with serum glucose concentrations of the patients using and not using icodextrin.
METHODS: Fifty-two chronic ambulatory peritoneal dialysis (CAPD) patients using icodextrin (Extraneal®) and 20 CAPD patients using another hyperosmotic fluid (Dianeal®) were included in the study. Duplicate capillary and serum glucose concentrations were measured with two different POCT glucose strips and central laboratory hexokinase method. Assay principles of glucose strips were based on glucose dehydrogenase-pyrroloquinoline quinone (GDH-PQQ) and a mutant variant of GDH (Mut Q-GDH). The results of both strips were compared with those of hexokinase method.
RESULTS: Regression equations between POCT and hexokinase methods in icodextrin group were y = 2.55x + 1.12 mmol/l and y = 1.057x + 0.16 mmol/l for the GDH-PQQ and Mut Q-GDH methods, respectively. The mean difference between the results of hexokinase and those of GDH-PQQ and Mut Q-GDH in icodextrin group was 3.41 ± 1.56 and 0.72 ± 0.64 mmol/l, respectively. However, the mean differences were found much lower in the control group; 0.64 mmol/l for GDH-PQQ and 0.52 mmol/l for Mut Q-GDH.
CONCLUSION: Compared to GDH-PQQ, glucose strips of Mut Q-GDH correlated better with hexokinase method in PD patients using icodextrin.
© 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  glucose; icodextrin; interference; peritoneal dialysis; point-of-care testing

Mesh:

Substances:

Year:  2015        PMID: 26511081      PMCID: PMC6807095          DOI: 10.1002/jcla.21887

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  11 in total

Review 1.  Review of glucose oxidases and glucose dehydrogenases: a bird's eye view of glucose sensing enzymes.

Authors:  Stefano Ferri; Katsuhiro Kojima; Koji Sode
Journal:  J Diabetes Sci Technol       Date:  2011-09-01

2.  The new glucose standard POCT12-A3 misses the mark.

Authors:  Jan S Krouwer
Journal:  J Diabetes Sci Technol       Date:  2013-09-01

Review 3.  Icodextrin metabolites in peritoneal dialysis.

Authors:  Elvia García-López; Bengt Lindholm
Journal:  Perit Dial Int       Date:  2009 Jul-Aug       Impact factor: 1.756

4.  Positive interference of icodextrin metabolites in some enzymatic glucose methods.

Authors:  W Janssen; G Harff; M Caers; A Schellekens
Journal:  Clin Chem       Date:  1998-11       Impact factor: 8.327

Review 5.  Technology behind commercial devices for blood glucose monitoring in diabetes management: a review.

Authors:  Sandeep Kumar Vashist; Dan Zheng; Khalid Al-Rubeaan; John H T Luong; Fwu-Shan Sheu
Journal:  Anal Chim Acta       Date:  2011-07-23       Impact factor: 6.558

6.  The danger of using inappropriate point-of-care glucose meters in patients on icodextrin dialysis.

Authors:  N J Perera; P M Stewart; P F Williams; E L Chua; D K Yue; S M Twigg
Journal:  Diabet Med       Date:  2011-10       Impact factor: 4.359

7.  Glucose meters: evaluation of the new formulation measuring strips from Roche (Accu-Chek) and Abbott (MediSense).

Authors:  G Dimeski; B W Jones; V Tilley; M N Greenslade; A W Russell
Journal:  Ann Clin Biochem       Date:  2010-07       Impact factor: 2.057

8.  False elevation of blood glucose levels measured by GDH-PQQ-based glucometers occurs during all daily dwells in peritoneal dialysis patients using icodextrin.

Authors:  Chiu-Yeh Tsai; Su-Chu Lee; Chi-Chih Hung; Jia-Jung Lee; Mei-Chuan Kuo; Shang-Jyh Hwang; Hung-Chun Chen
Journal:  Perit Dial Int       Date:  2010-02-26       Impact factor: 1.756

Review 9.  Update on peritoneal dialysis solutions.

Authors:  C W McIntyre
Journal:  Kidney Int       Date:  2007-02-14       Impact factor: 10.612

Review 10.  Icodextrin: a review of its use in peritoneal dialysis.

Authors:  James E Frampton; Greg L Plosker
Journal:  Drugs       Date:  2003       Impact factor: 9.546

View more