Literature DB >> 20592335

Glucose meters: evaluation of the new formulation measuring strips from Roche (Accu-Chek) and Abbott (MediSense).

G Dimeski1, B W Jones, V Tilley, M N Greenslade, A W Russell.   

Abstract

BACKGROUND: Both Roche and Abbott have released new glucose meter strips. They supply the entire Australian hospital market. The present study compared the performance of the new strips utilizing various specimen types (capillary, venous lithium heparin whole blood, venous lithium heparin plasma and serum) and evaluated how well they comply with the International Standards Organization (ISO) 15197 criteria.
METHODS: The study included imprecision, patient comparison and interference studies. Participants with and without diabetes were recruited to evaluate the performance of various specimen types against the Beckman DxC800 glucose method. The strips were tested for different interferences: galactose, maltose, lactose, Icodextrin, Intragam, paracetamol, sodium, ascorbic acid, variable strip storage temperature, haematocrit, haemolysis and lipaemia.
RESULTS: The imprecision of the two strips was approximately 5% or less, except for the Abbott strip at very low values (1.4 mmol/L), approximately 7%. In total, 78% and 84%, respectively, of the results from the finger prick capillary specimens with the Roche (Accu-Chek Performa meter) and Abbott (Optium Xceed meter) strips, not 95% or greater as recommended by the ISO guideline, were within the recommended limits compared with reference plasma estimation on laboratory analysers. Galactose, ascorbic acid, haematocrit and sodium on the Roche and ascorbic acid and haematocrit on the Abbott strip continue to interfere to a variable degree with the glucose measurement.
CONCLUSION: Analytically small differences exist between the glucose meter strips. The most significant analytical difference with the strips was at low glucose levels when compared with laboratory analyses and this may be of clinical importance. The impact of some of the interferences is variable between the two strips. Individuals, health-care professionals and health-care institutions should consider these data when selecting glucose meters for the management of people with diabetes mellitus.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20592335     DOI: 10.1258/acb.2010.009291

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  9 in total

Review 1.  Assessing the quality of publications evaluating the accuracy of blood glucose monitoring systems.

Authors:  Gary H Thorpe
Journal:  Diabetes Technol Ther       Date:  2013-02-15       Impact factor: 6.118

2.  Falsely Elevated Glucose Concentrations in Peritoneal Dialysis Patients Using Icodextrin.

Authors:  Kübra Dogan; Damla Kayalp; Gözde Ceylan; Alper Azak; Mehmet Senes; Murat Duranay; Dogan Yucel
Journal:  J Clin Lab Anal       Date:  2015-10-29       Impact factor: 2.352

3.  Comparative performance assessment of point-of-care testing devices for measuring glucose and ketones at the patient bedside.

Authors:  Ferruccio Ceriotti; Ewa Kaczmarek; Elena Guerra; Fabrizio Mastrantonio; Fausto Lucarelli; Francesco Valgimigli; Andrea Mosca
Journal:  J Diabetes Sci Technol       Date:  2014-12-17

4.  The association of hyperglycaemia with prevalent tuberculosis: a population-based cross-sectional study.

Authors:  Sarah Lou Bailey; Helen Ayles; Nulda Beyers; Peter Godfrey-Faussett; Monde Muyoyeta; Elizabeth du Toit; John S Yudkin; Sian Floyd
Journal:  BMC Infect Dis       Date:  2016-12-05       Impact factor: 3.090

5.  Highly specific and rapid glycan based amperometric detection of influenza viruses.

Authors:  Xikai Cui; Amrita Das; Abasaheb N Dhawane; Joyce Sweeney; Xiaohu Zhang; Vasanta Chivukula; Suri S Iyer
Journal:  Chem Sci       Date:  2017-02-14       Impact factor: 9.825

6.  A community pharmacy-based cardiovascular risk screening service implemented in Iran.

Authors:  Zahra Jahangard-Rafsanjani; Negar Hakimzadeh; Amir Sarayani; Sheyda Najafi; Kazem Heidari; Mohammad R Javadi; Molouk Hadjibabaie; Kheirollah Gholami
Journal:  Pharm Pract (Granada)       Date:  2017-06-30

Review 7.  Smartphone-Based Electrochemical Systems for Glucose Monitoring in Biofluids: A Review.

Authors:  Jie Xu; Zupeng Yan; Qingjun Liu
Journal:  Sensors (Basel)       Date:  2022-07-28       Impact factor: 3.847

8.  Analytical validation of a portable human Accu-Chek glucometer in honeybee hemolymph.

Authors:  Antoine Cournoyer; Annie Deschamps; Liza Bau-Gaudreault; Pascal Dubreuil; Marie-Odile Benoit-Biancamano
Journal:  J Vet Diagn Invest       Date:  2022-08-21       Impact factor: 1.569

Review 9.  Hybrid Technologies Combining Solid-State Sensors and Paper/Fabric Fluidics for Wearable Analytical Devices.

Authors:  Meritxell Rovira; César Fernández-Sánchez; Cecilia Jiménez-Jorquera
Journal:  Biosensors (Basel)       Date:  2021-08-28
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.