CT image artifacts from brachytherapy seed implants: a postprocessing 3D adaptive median filter.

Parminder S Basran1, Andrew Robertson, Derek Wells.   

Abstract

PURPOSE: To design a postprocessing 3D adaptive median filter that minimizes streak artifacts and improves soft-tissue contrast in postoperative CT images of brachytherapy seed implantations.
METHODS: The filter works by identifying voxels that are likely streaks and estimating more reflective voxel intensity by using voxel intensities in adjacent CT slices and applying a median filter over voxels not identified as seeds. Median values are computed over a 5 x 5 x 5 mm region of interest (ROI) within the CT volume. An acrylic phantom simulating a clinical seed implant arrangement and containing nonradioactive seeds was created. Low contrast subvolumes of tissuelike material were also embedded in the phantom. Pre- and postprocessed image quality metrics were compared using the standard deviation of ROIs between the seeds, the CT numbers of low contrast ROIs embedded within the phantom, the signal to noise ratio (SNR), and the contrast to noise ratio (CNR) of the low contrast ROIs. The method was demonstrated with a clinical postimplant CT dataset.
RESULTS: After the filter was applied, the standard deviation of CT values in streak artifact regions was significantly reduced from 76.5 to 7.2 HU. Within the observable low contrast plugs, the mean of all ROI standard deviations was significantly reduced from 60.5 to 3.9 HU, SNR significantly increased from 2.3 to 22.4, and CNR significantly increased from 0.2 to 4.1 (all P < 0.01). The mean CT in the low contrast plugs remained within 5 HU of the original values.
CONCLUSION: An efficient postprocessing filter that does not require access to projection data, which can be applied irrespective of CT scan parameters has been developed, provided the slice thickness and spacing is 3 mm or less.

Entities:  

Mesh:

Year:  2011        PMID: 21452708     DOI: 10.1118/1.3539648

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Spectral CT with monochromatic imaging and metal artifacts reduction software for artifacts reduction of ¹²⁵I radioactive seeds in liver brachytherapy.

Authors:  Qiuxia Yang; Sheng Peng; Jing Wu; Xiaohua Ban; Mingyan He; Chuanmiao Xie; Rong Zhang
Journal:  Jpn J Radiol       Date:  2015-10-11       Impact factor: 2.374

2.  The combined use of 2D scout and 3D axial CT images to accurately determine the catheter tips for high-dose-rate brachytherapy plans.

Authors:  Kun Qing; Ning J Yue; Lara Hathout; Chi Ma; Meral Reyhan; Jiahua Zhu; Ke Nie; Gilbert Monte; Irina Vergalasova
Journal:  J Appl Clin Med Phys       Date:  2021-02-27       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.