Literature DB >> 26450111

A fast, accurate and closed-form method for pose recognition of an intramedullary nail using a tracked C-arm.

H Esfandiari1, S Amiri2, D D Lichti3, C Anglin4,5.   

Abstract

PURPOSE: The main challenge of intramedullary nail (IM-nail) fixation surgery is to achieve the X-ray shot in which the distal holes of the IM-nail appear as circles (desired view); screw insertion is then performed. Although C-arm X-ray devices are routinely used in IM-nail fixation procedures, the surgeons or radiation technologists (rad-techs) usually utilize them in a trial-and-error manner. This increases both radiation exposure and surgery time. This study addresses the problem by presenting a C-arm-based IM-nail pose recognition method.
METHODS: The specific purpose of this study was to develop and validate an automated technique for identifying the current pose of the IM-nail relative to the C-arm. An accuracy assessment is performed to test the reliability of the navigation results. The algorithm starts with a sequential biplanar set of X-ray imagery (acquired from a tracked C-arm) of the distal part of an inserted IM-nail. The image-processing module then extracts features of interest, and a mathematical model incorporates them to calculate the six degree-of-freedom position and orientation parameters of the nail.
RESULTS: Translational accuracy was demonstrated to be better than 0.5 mm, rotational accuracy for roll and pitch to be better than 2° and for yaw to be better than 2.5° depending on the separation angle. Computation time was less than 0.5 s. .
CONCLUSION: An IM-nail distal locking navigation technique is introduced in this study that leads to more accurate and faster screw placement with a lower consumption of radiation dose and a minimum number of added steps to the operation.

Keywords:  C-arm; IM-nail; Pose estimation; Pose recognition; Radiation

Mesh:

Year:  2015        PMID: 26450111     DOI: 10.1007/s11548-015-1294-y

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  14 in total

1.  Inserting the distal screws in a locked femoral nail.

Authors:  C J Knudsen; G P Grobler; R E Close
Journal:  J Bone Joint Surg Br       Date:  1991-07

2.  Accuracy of single-plane fluoroscopy in determining relative position and orientation of total knee replacement components.

Authors:  Stacey Acker; Rebecca Li; Heather Murray; Paul St John; Scott Banks; Shang Mu; Urs Wyss; Kevin Deluzio
Journal:  J Biomech       Date:  2010-11-18       Impact factor: 2.712

3.  Radiation exposure to the surgeon during fluoroscopically assisted percutaneous vertebroplasty: a prospective study.

Authors:  Roger Harstall; Paul F Heini; Roberto L Mini; Rene Orler
Journal:  Spine (Phila Pa 1976)       Date:  2005-08-15       Impact factor: 3.468

4.  A robust and accurate two-stage approach for automatic recovery of distal locking holes in computer-assisted intramedullary nailing of femoral shaft fractures.

Authors:  Guoyan Zheng; Xuan Zhang; Daniel Haschtmann; Philippe Gedet; Xiao Dong; Lutz-Peter Nolte
Journal:  IEEE Trans Med Imaging       Date:  2008-02       Impact factor: 10.048

5.  A low-cost tracked C-arm (TC-arm) upgrade system for versatile quantitative intraoperative imaging.

Authors:  Shahram Amiri; David R Wilson; Bassam A Masri; Carolyn Anglin
Journal:  Int J Comput Assist Radiol Surg       Date:  2013-12-10       Impact factor: 2.924

6.  Radiation exposure to the hands and the thyroid of the surgeon during intramedullary nailing.

Authors:  L P Müller; J Suffner; K Wenda; W Mohr; P M Rommens
Journal:  Injury       Date:  1998-07       Impact factor: 2.586

7.  A computer assisted orthopaedic surgical system for distal locking of intramedullary nails.

Authors:  W J Viant; R Phillips; J G Griffiths; T O Ozanian; A M Mohsen; T J Cain; M R Karpinski; K P Sherman
Journal:  Proc Inst Mech Eng H       Date:  1997       Impact factor: 1.617

8.  The insertion of intramedullary nail locking screws without fluoroscopy: a faster and safer technique.

Authors:  Daniel S Chan; Richard B Burris; Murat Erdogan; H Claude Sagi
Journal:  J Orthop Trauma       Date:  2013-07       Impact factor: 2.512

9.  Surgical navigation based on fluoroscopy--clinical application for computer-assisted distal locking of intramedullary implants.

Authors:  N Suhm; A L Jacob; L P Nolte; P Regazzoni; P Messmer
Journal:  Comput Aided Surg       Date:  2000

10.  Deformation of femoral nails with intramedullary insertion.

Authors:  C Krettek; J Mannss; T Miclau; P Schandelmaier; I Linnemann; H Tscherne
Journal:  J Orthop Res       Date:  1998-09       Impact factor: 3.494

View more
  1 in total

1.  C-Arm Image-Based Surgical Path Planning Method for Distal Locking of Intramedullary Nails.

Authors:  Wei-En Hsu; Ching-Hsiao Yu; Chih-Ju Chang; Hung-Kang Wu; Tsong-Han Yu; Ching-Shiow Tseng
Journal:  Appl Bionics Biomech       Date:  2018-05-23       Impact factor: 1.781