Literature DB >> 26444717

Robust time-shifted spoke pulse design in the presence of large B0 variations with simultaneous reduction of through-plane dephasing, B1+ effects, and the specific absorption rate using parallel transmission.

Bastien Guérin1, Jason P Stockmann1,2, Mehran Baboli3, Angel Torrado-Carvajal4,5, Andrew V Stenger3, Lawrence L Wald1,6.   

Abstract

PURPOSE: To design parallel transmission spokes pulses with time-shifted profiles for joint mitigation of intensity variations due to B1+ effects, signal loss due to through-plane dephasing, and the specific absorption rate (SAR) at 7T.
METHODS: We derived a slice-averaged small tip angle (SA-STA) approximation of the magnetization signal at echo time that depends on the B1+ transmit profiles, the through-slice B0 gradient and the amplitude and time-shifts of the spoke waveforms. We minimize a magnitude least-squares objective based on this signal equation using a fast interior-point approach with analytical expressions of the Jacobian and Hessian.
RESULTS: Our algorithm runs in less than three minutes for the design of two-spoke pulses subject to hundreds of local SAR constraints. On a B0/B1+ head phantom, joint optimization of the channel-dependent time-shifts and spoke amplitudes allowed signal recovery in high-B0 regions at no increase of SAR. Although the method creates uniform magnetization profiles (ie, uniform intensity), the flip angle varies across the image, which makes it ill-suited to T1-weighted applications.
CONCLUSIONS: The SA-STA approach presented in this study is best suited to T2*-weighted applications with long echo times that require signal recovery around high B0 regions. Magn Reson Med 76:540-554, 2016.
© 2015 Wiley Periodicals, Inc. © 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  B0 artefact; B1+ homogeneity; parallel transmission; specific absorption rate; spokes

Mesh:

Year:  2015        PMID: 26444717      PMCID: PMC4824674          DOI: 10.1002/mrm.25902

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  32 in total

1.  Compensation of susceptibility-induced BOLD sensitivity losses in echo-planar fMRI imaging.

Authors:  R Deichmann; O Josephs; C Hutton; D R Corfield; R Turner
Journal:  Neuroimage       Date:  2002-01       Impact factor: 6.556

2.  EPI image reconstruction with correction of distortion and signal losses.

Authors:  Guoxiang Liu; Seiji Ogawa
Journal:  J Magn Reson Imaging       Date:  2006-09       Impact factor: 4.813

3.  Spatial domain method for the design of RF pulses in multicoil parallel excitation.

Authors:  William Grissom; Chun-yu Yip; Zhenghui Zhang; V Andrew Stenger; Jeffrey A Fessler; Douglas C Noll
Journal:  Magn Reson Med       Date:  2006-09       Impact factor: 4.668

4.  Reducing susceptibility artifacts in fMRI using volume-selective z-shim compensation.

Authors:  Yiping P Du; Manish Dalwani; Korey Wylie; Eric Claus; Jason R Tregellas
Journal:  Magn Reson Med       Date:  2007-02       Impact factor: 4.668

5.  MR susceptibility misregistration correction.

Authors:  T S Sumanaweera; G H Glover; T O Binford; J R Adler
Journal:  IEEE Trans Med Imaging       Date:  1993       Impact factor: 10.048

6.  Fast MRI coil analysis based on 3-D electromagnetic and RF circuit co-simulation.

Authors:  Mikhail Kozlov; Robert Turner
Journal:  J Magn Reson       Date:  2009-06-09       Impact factor: 2.229

7.  Multi-atlas and label fusion approach for patient-specific MRI based skull estimation.

Authors:  Angel Torrado-Carvajal; Joaquin L Herraiz; Juan A Hernandez-Tamames; Raul San Jose-Estepar; Yigitcan Eryaman; Yves Rozenholc; Elfar Adalsteinsson; Lawrence L Wald; Norberto Malpica
Journal:  Magn Reson Med       Date:  2015-05-18       Impact factor: 4.668

8.  A 16-channel dual-row transmit array in combination with a 31-element receive array for human brain imaging at 9.4 T.

Authors:  G Shajan; Mikhail Kozlov; Jens Hoffmann; Robert Turner; Klaus Scheffler; Rolf Pohmann
Journal:  Magn Reson Med       Date:  2014-02       Impact factor: 4.668

9.  Automated slice-specific simultaneous z-shim method for reducing B1 inhomogeneity and susceptibility-induced signal loss with parallel transmission at 3T.

Authors:  Rainer Schneider; Fernando Boada; Jens Haueisen; Josef Pfeuffer
Journal:  Magn Reson Med       Date:  2014-10-07       Impact factor: 4.668

10.  Spectral decomposition of susceptibility artifacts for spectral-spatial radiofrequency pulse design.

Authors:  Cungeng Yang; Benedikt A Poser; Weiran Deng; V Andrew Stenger
Journal:  Magn Reson Med       Date:  2012-02-14       Impact factor: 4.668

View more
  7 in total

1.  Variable slice thickness (VAST) EPI for the reduction of susceptibility artifacts in whole-brain GE-EPI at 7 Tesla.

Authors:  Sascha Brunheim; Sören Johst; Viktor Pfaffenrot; Stefan Maderwald; Harald H Quick; Benedikt A Poser
Journal:  MAGMA       Date:  2017-07-10       Impact factor: 2.310

2.  Application of the limited-memory quasi-Newton algorithm for multi-dimensional, large flip-angle RF pulses at 7T.

Authors:  Mads S Vinding; Daniel Brenner; Desmond H Y Tse; Sebastian Vellmer; Thomas Vosegaard; Dieter Suter; Tony Stöcker; Ivan I Maximov
Journal:  MAGMA       Date:  2016-08-02       Impact factor: 2.310

3.  Reducing temperature errors in transcranial MR-guided focused ultrasound using a reduced-field-of-view sequence.

Authors:  William A Grissom; Steven Allen
Journal:  Magn Reson Med       Date:  2019-09-04       Impact factor: 4.668

4.  External Dynamic InTerference Estimation and Removal (EDITER) for low field MRI.

Authors:  Sai Abitha Srinivas; Stephen F Cauley; Jason P Stockmann; Charlotte R Sappo; Christopher E Vaughn; Lawrence L Wald; William A Grissom; Clarissa Z Cooley
Journal:  Magn Reson Med       Date:  2021-09-04       Impact factor: 4.668

5.  Realistic head-shaped phantom with brain-mimicking metabolites for 7 T spectroscopy and spectroscopic imaging.

Authors:  Ghil Jona; Edna Furman-Haran; Rita Schmidt
Journal:  NMR Biomed       Date:  2020-10-04       Impact factor: 4.044

6.  Reducing SAR in 7T brain fMRI by circumventing fat suppression while removing the lipid signal through a parallel acquisition approach.

Authors:  Amir Seginer; Edna Furman-Haran; Ilan Goldberg; Rita Schmidt
Journal:  Sci Rep       Date:  2021-07-28       Impact factor: 4.379

7.  An integrated RF-receive/B0-shim array coil boosts performance of whole-brain MR spectroscopic imaging at 7 T.

Authors:  Morteza Esmaeili; Jason Stockmann; Bernhard Strasser; Nicolas Arango; Bijaya Thapa; Zhe Wang; Andre van der Kouwe; Jorg Dietrich; Daniel P Cahill; Tracy T Batchelor; Jacob White; Elfar Adalsteinsson; Lawrence Wald; Ovidiu C Andronesi
Journal:  Sci Rep       Date:  2020-09-14       Impact factor: 4.996

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.