Geometric validation of self-gating k-space-sorted 4D-MRI vs 4D-CT using a respiratory motion phantom.

Yong Yue1, Zhaoyang Fan2, Wensha Yang1, Jianing Pang2, Zixin Deng3, Elizabeth McKenzie1, Richard Tuli1, Robert Wallace1, Debiao Li3, Benedick Fraass1.   

Abstract

PURPOSE: MRI is increasingly being used for radiotherapy planning, simulation, and in-treatment-room motion monitoring. To provide more detailed temporal and spatial MR data for these tasks, we have recently developed a novel self-gated (SG) MRI technique with advantage of k-space phase sorting, high isotropic spatial resolution, and high temporal resolution. The current work describes the validation of this 4D-MRI technique using a MRI- and CT-compatible respiratory motion phantom and comparison to 4D-CT.
METHODS: The 4D-MRI sequence is based on a spoiled gradient echo-based 3D projection reconstruction sequence with self-gating for 4D-MRI at 3 T. Respiratory phase is resolved by using SG k-space lines as the motion surrogate. 4D-MRI images are reconstructed into ten temporal bins with spatial resolution 1.56 × 1.56 × 1.56 mm(3). A MRI-CT compatible phantom was designed to validate the performance of the 4D-MRI sequence and 4D-CT imaging. A spherical target (diameter 23 mm, volume 6.37 ml) filled with high-concentration gadolinium (Gd) gel is embedded into a plastic box (35 × 40 × 63 mm(3)) and stabilized with low-concentration Gd gel. The phantom, driven by an air pump, is able to produce human-type breathing patterns between 4 and 30 respiratory cycles/min. 4D-CT of the phantom has been acquired in cine mode, and reconstructed into ten phases with slice thickness 1.25 mm. The 4D images sets were imported into a treatment planning software for target contouring. The geometrical accuracy of the 4D MRI and CT images has been quantified using target volume, flattening, and eccentricity. The target motion was measured by tracking the centroids of the spheres in each individual phase. Motion ground-truth was obtained from input signals and real-time video recordings.
RESULTS: The dynamic phantom has been operated in four respiratory rate (RR) settings, 6, 10, 15, and 20/min, and was scanned with 4D-MRI and 4D-CT. 4D-CT images have target-stretching, partial-missing, and other motion artifacts in various phases, whereas the 4D-MRI images are visually free of those artifacts. Volume percentage difference for the 6.37 ml target ranged from 5.3% ± 4.3% to 10.3% ± 5.9% for 4D-CT, and 1.47 ± 0.52 to 2.12 ± 1.60 for 4D-MRI. With an increase of respiratory rate, the target volumetric and geometric deviations increase for 4D-CT images while remaining stable for the 4D-MRI images. Target motion amplitude errors at different RRs were measured with a range of 0.66-1.25 mm for 4D-CT and 0.2-0.42 mm for 4D-MRI. The results of Mann-Whitney tests indicated that 4D-MRI significantly outperforms 4D-CT in phase-based target volumetric (p = 0.027) and geometric (p < 0.001) measures. Both modalities achieve equivalent accuracy in measuring motion amplitude (p = 0.828).
CONCLUSIONS: The k-space self-gated 4D-MRI technique provides a robust method for accurately imaging phase-based target motion and geometry. Compared to 4D-CT, the current 4D-MRI technique demonstrates superior spatiotemporal resolution, and robust resistance to motion artifacts caused by fast target motion and irregular breathing patterns. The technique can be used extensively in abdominal targeting, motion gating, and toward implementing MRI-based adaptive radiotherapy.

Entities:  

Mesh:

Year:  2015        PMID: 26429253      PMCID: PMC4575318          DOI: 10.1118/1.4929552

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  28 in total

Review 1.  4-dimensional computed tomography imaging and treatment planning.

Authors:  Paul Keall
Journal:  Semin Radiat Oncol       Date:  2004-01       Impact factor: 5.934

2.  Acquiring a four-dimensional computed tomography dataset using an external respiratory signal.

Authors:  S S Vedam; P J Keall; V R Kini; H Mostafavi; H P Shukla; R Mohan
Journal:  Phys Med Biol       Date:  2003-01-07       Impact factor: 3.609

3.  MRI-based simulation of treatment plans for ion radiotherapy in the brain region.

Authors:  Christopher M Rank; Nora Hünemohr; Armin M Nagel; Matthias C Röthke; Oliver Jäkel; Steffen Greilich
Journal:  Radiother Oncol       Date:  2013-11-19       Impact factor: 6.280

4.  Respiration-based sorting of dynamic MRI to derive representative 4D-MRI for radiotherapy planning.

Authors:  Erik Tryggestad; Aaron Flammang; Sarah Han-Oh; Russell Hales; Joseph Herman; Todd McNutt; Teboh Roland; Steven M Shea; John Wong
Journal:  Med Phys       Date:  2013-05       Impact factor: 4.071

5.  The Australian magnetic resonance imaging-linac program.

Authors:  Paul J Keall; Michael Barton; Stuart Crozier
Journal:  Semin Radiat Oncol       Date:  2014-07       Impact factor: 5.934

6.  The rotating biplanar linac-magnetic resonance imaging system.

Authors:  Biagio Gino Fallone
Journal:  Semin Radiat Oncol       Date:  2014-07       Impact factor: 5.934

Review 7.  The magnetic resonance imaging-linac system.

Authors:  Jan J W Lagendijk; Bas W Raaymakers; Marco van Vulpen
Journal:  Semin Radiat Oncol       Date:  2014-07       Impact factor: 5.934

8.  Four-dimensional MRI using three-dimensional radial sampling with respiratory self-gating to characterize temporal phase-resolved respiratory motion in the abdomen.

Authors:  Zixin Deng; Jianing Pang; Wensha Yang; Yong Yue; Behzad Sharif; Richard Tuli; Debiao Li; Benedick Fraass; Zhaoyang Fan
Journal:  Magn Reson Med       Date:  2015-05-14       Impact factor: 4.668

9.  Accelerated whole-heart coronary MRA using motion-corrected sensitivity encoding with three-dimensional projection reconstruction.

Authors:  Jianing Pang; Behzad Sharif; Reza Arsanjani; Xiaoming Bi; Zhaoyang Fan; Qi Yang; Kuncheng Li; Daniel S Berman; Debiao Li
Journal:  Magn Reson Med       Date:  2014-01-16       Impact factor: 4.668

10.  ECG and navigator-free four-dimensional whole-heart coronary MRA for simultaneous visualization of cardiac anatomy and function.

Authors:  Jianing Pang; Behzad Sharif; Zhaoyang Fan; Xiaoming Bi; Reza Arsanjani; Daniel S Berman; Debiao Li
Journal:  Magn Reson Med       Date:  2014-09-12       Impact factor: 4.668

View more
  5 in total

1.  Multi-contrast four-dimensional magnetic resonance imaging (MC-4D-MRI): Development and initial evaluation in liver tumor patients.

Authors:  Lei Zhang; Fang-Fang Yin; Tian Li; Xinzhi Teng; Haonan Xiao; Wendy Harris; Lei Ren; Feng-Ming Spring Kong; Hong Ge; Ronghu Mao; Jing Cai
Journal:  Med Phys       Date:  2021-11-18       Impact factor: 4.506

2.  Computed tomography trachea volumetry in patients with scleroderma: Association with clinical and functional findings.

Authors:  Bruno Rangel Antunes Silva; Rosana Souza Rodrigues; Rogério Rufino; Cláudia Henrique Costa; Veronica Silva Vilela; Roger Abramino Levy; Alan Ranieri Medeiros Guimarães; Alysson Roncally Silva Carvalho; Agnaldo José Lopes
Journal:  PLoS One       Date:  2018-08-01       Impact factor: 3.240

3.  Novel 4D-MRI of tumor infiltrating vasculature: characterizing tumor and vessel volume motion for selective boost volume definition in pancreatic radiotherapy.

Authors:  Wensha Yang; Zhaoyang Fan; Zixin Deng; Jianing Pang; Xiaoming Bi; Benedick A Fraass; Howard Sandler; Debiao Li; Richard Tuli
Journal:  Radiat Oncol       Date:  2018-10-01       Impact factor: 3.481

4.  Design, Implementation, and Validation of a Pulsatile Heart Phantom Pump.

Authors:  Volkan Tuncay; Jan Zijlstra; Matthijs Oudkerk; Peter M A van Ooijen
Journal:  J Digit Imaging       Date:  2020-08-10       Impact factor: 4.056

5.  Evaluating differences in respiratory motion estimates during radiotherapy: a single planning 4DMRI versus daily 4DMRI.

Authors:  Duncan den Boer; Johannes K Veldman; Geertjan van Tienhoven; Arjan Bel; Zdenko van Kesteren
Journal:  Radiat Oncol       Date:  2021-09-26       Impact factor: 3.481

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.