S R Balakrishnan1, U Hashim1, Subash C B Gopinath1, P Poopalan2, H R Ramayya3, M Iqbal Omar4, R Haarindraprasad1, P Veeradasan1. 1. Biomedical Nano Diagnostics Research Group, Institute of Nano Electronic Engineering (INEE), Universiti Malaysia Perlis (UniMAP), Kangar, Perlis, Malaysia. 2. School of Microelectronic Engineering, University Malaysia Perlis (UniMAP), Kuala Perlis, Perlis, Malaysia. 3. Department of Obstetrics and Gynaecology, Hospital Tuanku Fauziah, Kangar, Perlis, Malaysia. 4. University Health Centre, Universiti Malaysia Perlis (UniMAP), Kuala Perlis, Perlis, Malaysia.
Abstract
Human chorionic gonadotropin (hCG), a glycoprotein hormone secreted from the placenta, is a key molecule that indicates pregnancy. Here, we have designed a cost-effective, label-free, in situ point-of-care (POC) immunosensor to estimate hCG using a cuneated 25 nm polysilicon nanogap electrode. A tiny chip with the dimensions of 20.5 × 12.5 mm was fabricated using conventional lithography and size expansion techniques. Furthermore, the sensing surface was functionalized by (3-aminopropyl)triethoxysilane and quantitatively measured the variations in hCG levels from clinically obtained human urine samples. The dielectric properties of the present sensor are shown with a capacitance above 40 nF for samples from pregnant women; it was lower with samples from non-pregnant women. Furthermore, it has been proven that our sensor has a wide linear range of detection, as a sensitivity of 835.88 μA mIU(-1) ml(-2) cm(-2) was attained, and the detection limit was 0.28 mIU/ml (27.78 pg/ml). The dissociation constant Kd of the specific antigen binding to the anti-hCG was calculated as 2.23 ± 0.66 mIU, and the maximum number of binding sites per antigen was Bmax = 22.54 ± 1.46 mIU. The sensing system shown here, with a narrow nanogap, is suitable for high-throughput POC diagnosis, and a single injection can obtain triplicate data or parallel analyses of different targets.
Humanchorionic gonadotropin (hCG), a glycoprotein hormone secreted from the placenta, is a key molecule that indicates pregnancy. Here, we have designed a cost-effective, label-free, in situ point-of-care (POC) immunosensor to estimate hCG using a cuneated 25 nm polysilicon nanogap electrode. A tiny chip with the dimensions of 20.5 × 12.5 mm was fabricated using conventional lithography and size expansion techniques. Furthermore, the sensing surface was functionalized by (3-aminopropyl)triethoxysilane and quantitatively measured the variations in hCG levels from clinically obtained human urine samples. The dielectric properties of the present sensor are shown with a capacitance above 40 nF for samples from pregnant women; it was lower with samples from non-pregnant women. Furthermore, it has been proven that our sensor has a wide linear range of detection, as a sensitivity of 835.88 μA mIU(-1) ml(-2) cm(-2) was attained, and the detection limit was 0.28 mIU/ml (27.78 pg/ml). The dissociation constant Kd of the specific antigen binding to the anti-hCG was calculated as 2.23 ± 0.66 mIU, and the maximum number of binding sites per antigen was Bmax = 22.54 ± 1.46 mIU. The sensing system shown here, with a narrow nanogap, is suitable for high-throughput POC diagnosis, and a single injection can obtain triplicate data or parallel analyses of different targets.
Humanchorionic gonadotropin (hCG) is a hormone that is expressed by placental trophoblasts during pregnancy and comprises α and β-subunits (hCGα & hCGβ). hCG has been widely considered as a biomarker of pregnancy and can protect against complications during prenatal care. It is also an important and critical marker in many advanced malignancies and can assist in cancer detection. Further, complications such as ectopic pregnancy, threatened abortion and pregnancy trophoblastic diseases, such as acephalocystis racemosa and chorioepithelioma, could be monitored via the detection of hCG [1-3]. This monitoring can be performed by antibody-assisted biosensors, or ‘immunosensors’, which have been developed using different types of electrodes to recognize specific targets on the sensing surface of biochips [4,5]. There are other sensing systems or strategies that have also been reported with high performance in hCG measurements [6-17]. Furthermore, different methods for hCG detection have been proposed using lab-dependent systems, such as surface plasmon resonance [6,18], enzyme-linked immunosorbent assay (ELISA) [19,20], fluoroimmunoassay [21], radioimmunoassay [22], immunochromatographic assay [23,24], resonance scattering spectral [25] and other sensing systems, including electrochemical [7,12-15,17,26,27], amperometric [16,28,29], capacitive [30] and impedimetric [31,32] systems. However, developing a lab-independent electrochemical immunosensor is necessary to facilitate the replacement of current existing lab-dependent detection systems.In most instances, lab-dependent assays require a large working space and long durations and involve the use of chemicals, which slows down the result acquisition processes. The development of independent point-of-care (POC) electrochemical immunosensors will overcome these limitations and promises a number of significant advantages, including simplicity, fast response, suitability for onsite testing, high sensitivity and time and cost savings. Consequently, the frequency of hospital visits, travel expenses and lost productivity on the job could be reduced. The success of ‘bedside analyzing systems’, such as glucose meters, has promoted the development of similar quantitative biosensing devices for pregnancy, which rely on myriad detection schemes. One of the promising alternatives to existing biosensors is capacitive based and can be used for POC applications. Capacitive biosensors are favored due to their ability to detect the analyte directly in a sample with little or no preparation [33]. The capacitive behavior of this type of sensor is based on parameters such as changes in dielectric properties, charge distribution, permittivity and conductivity that originate from the complex matrix formation on the surface of the transducing electrodes. Hence, changes in such parameters could be measured when biomolecule binding occurs on the electrode surface. In the quest for higher sensitivities and more accurate measurements, the sensing electrode shown in the present study could be used as a capacitive sensor.To develop the sensor, we used standard semiconducting materials, with silicon as the substrate and polysilicon (PS) as the sensing electrode. Polysilicon was chosen for its ability to withstand high temperatures and its high compatibility with standard semiconductor processing steps. The advantages of this material also include high sensitivity, a fast response time and high accuracy, as previously reported [34,35]. In the quest to reduce the use of expensive equipment, a conventional lithography microfabrication technique was implemented; this technique can be used to fabricate simple and non-complex nanostructures such as nanogaps and nanowires [36,37]. In this study, we developed a narrow-sized nanogap-based sensor with a single injection capability to obtain triplicate data or parallel analyses against different targets. Furthermore, because it is possible to block unused channel(s) to limit the sample usage, this system is more cost effective and easier to use. All abbreviation and symbols used in this study are described in Table 1.
Table 1
List of Symbols and Abbreviation.
PSNG
Polysilicon Nanogap
PS
Polysilicon
hCG
Human Chorionic Gonadotropin
A
Electrode area
ν
Applied voltage
lg
Gap width
le
Electrode width
lSM
Width of total surface modification
lhcg
hCG width
ρair
Air resistivity
ρe
Electrode resistivity
ρSM
Resistivity of total surface modification
PhCG
hCG resistivity
ɛT
Total permittivity
ɛe
Permittivity of electrode
ɛSM
Permittivity of total surface modification
ɛhCG
Permittivity of hCG
ZT
Total impedance
de
Dielectric width of electrode
dSM
Dielectric width of total surface modification
dhcg
Dielectric width of hCG
Materials and Methods
Materials and Reagents
Single crystal silicon wafers were used as substrates for the lab-on-chip (LOC) development throughout the experiment. The fabrication of polysilicon nanogap (PSNG) electrodes requires cleaning of the wafers using buffered oxide etchant (BOE; 6:1; J.T. Baker, Center Valley, Pennsylvania, USA) and standard cleaning 1 (RCA1) and standard cleaning 2 (RCA2) solutions. Positive photoresist (PR1-2000A), negative photoresist (NR7-6000PY) and a resist developer (RD6) for pattern transfer were purchased from Futurrex, Inc., Franklin, New Jersey, USA. Acetone, aluminum etch (80-15-3-2), hydrochloric acid (HCl; 37%), aqueous ammonia (NH4OH; 30%), hydrogen peroxide (H2O2; 30%), and sulfuric acid (H2SO4) were purchased from J.T. Baker, Center Valley, Pennsylvania, USA. The DI water used throughout the experiment was produced by an RO deionization system (13–16 MΩ-cm, SASTEC (M) Sdn. Bhd, Malaysia). Aluminum (Al), titanium (Ti) and gold (Au) metals were used to deposit thin Al, Ti and Au layers. Microfluidic fabrication was performed using the Sylgard 184 Silicone Elastomer Base and Sylgard 184 Silicone Elastomer Curing Agent (Dow Corning Corp., Midland, USA) with SU-8 negative photoresist and isopropanol (IPA) from J.T. Baker, Center Valley, Pennsylvania, USA. To achieve surface modification, iron (II) sulfate (FeSO4), sodium bicarbonate (NaHCO3) and chloroform (CHCl3) were purchased from HmBG Chemicals, Germany. (3-aminopropyl)triethoxysilane (APTES) 99% of 221.37 g/mol, bovineserum albumin (BSA) ≥98.0%, trifluoroacetic acid (TFA) 99% CF3CO2H, N-(3-dimethylaminopropyl)-N′-ethylcarbodiimide hydrochloride (EDC), di-tert-butyl dicarbonate (tBOC; ≥99%) C10H18O5, N-hydroxysuccinimide (NHS; 98%) C4H5NO3, ethylenediaminetetraacetic acid (EDTA; ≥99%) C10H16N2O8 and phosphate buffered saline (PBS; pH 7.4) were purchased from Sigma-Aldrich, St. Louis, Missouri, USA. Glucose (GLUC), carcinoembryonic antigen (CEA), uric acid (UA) and ascorbic acid (AA), also from Sigma-Aldrich, St. Louis, Missouri, USA, were used for the selectivity test. Monoclonal hCG beta antibody (hCGab; GTX42543; GeneTex, Irvine, California, USA) and hCG standards were tested with the hCG ELISA Kit; CSB-E05060h; CUSABIO Biotech, Hubei, China. All fresh 24-h urine samples for pregnant and non-pregnant women were obtained with permission from the UniMAP Health Centre. All samples were acquired with unrecorded verbal consent from patients who were informed about the research purposes, and this consent procedure was approved by the Medical Ethics Committee for the University Health Centre, Universiti Malaysia Perlis. The above-mentioned ethics committee specifically approved this study, which was conducted in accordance with the International Conference on Harmonisation–Good Clinical Practice (ICH–GCP) guidelines and the Declaration of Helsinki.
Design and fabrication of PSNG lab-on-chip (LOC)
The patterns for the PSNG electrode, contact pads and microfluidics were designed in AutoCAD for pattern transfer via mask aligner (Fig 1a). The PSNG mask has a critical dimension of 1 μm at the electrode gap. The 20.5 mm × 12.5 mm LOC immunosensor is shown in Fig 1b. The LOC consists of three standalone PSNG electrodes as the base and is bonded with microfluidic channels that serve as the fluid delivery system to the specific electrodes (Fig 1c) [38-47].
Fig 1
Design and specifications of the LOC immunosensor.
(a) The chrome mask, as seen under microscopy, for the electrodes, contacts and microfluidics used for the sensor fabrication. (b) The specifications of the immunosensor consisting of microfluidic channels for fluid delivery and chemically modified PSNG electrodes underneath for hCG detection. (c) The designed LOC immunosensor with dimensions of 20.5 × 12.5 mm.
Design and specifications of the LOC immunosensor.
(a) The chrome mask, as seen under microscopy, for the electrodes, contacts and microfluidics used for the sensor fabrication. (b) The specifications of the immunosensor consisting of microfluidic channels for fluid delivery and chemically modified PSNG electrodes underneath for hCG detection. (c) The designed LOC immunosensor with dimensions of 20.5 × 12.5 mm.A conventional lithography technique was used to fabricate the PSNG immunosensor, as reported elsewhere [37,43]. Fig 2a–2f explains the simplified fabrication process steps used to develop the PSNG electrode. Oxide was grown by wet oxidation, and polysilicon (PS) was deposited by the low-pressure chemical vapor deposition (LPCVD) technique. The deposited PS wafer was then annealed for 12 h at 600°C to allow the disordered PS molecules to rearrange after deposition. Al was first deposited as a sacrificial layer and patterned using the MIDAS Exposure System MDA-400M for the first electrode mask before PS etching, which used the reactive ion etching (RIE) technique. To obtain a second mask pattern for Ti/Au contact padding, lift-off of Ti and Au was carried out after their deposition with a thermal evaporator (Auto 306 thermal evaporator; Edwards High Vacuum International, Wilmington, MA, USA). The cuneated PS electrode was first oxidized at 900°C in a dry O2 atmosphere to form an oxide layer while simultaneously expanding the PS. Subsequent BOE etching revealed this expanded PS, and the entire oxidation and etching process was repeated until the desired nanogap (NG) was achieved (size expansion technique). The oxidation time varied from 5 to 15 min in each cycle of oxidation, and etching depended on the thickness of the polysilicon layer. The final fabricated PSNG electrode, bonded to the microfluidic channel via plasma oxidation, is shown in Fig 2g. The microfluidic layer on the LOC was fabricated using an SU-8 soft-lithography microstructure mold (Fig 2h–2j). A mixture of Sylgard 184 Silicone Elastomer Base and Sylgard 184 Silicone Elastomer Curing Agent was poured onto the SU-8 microstructure mold to form the PDMS polymer.
Fig 2
Fabrication of the PSNG LOC immunosensor using standard conventional lithography.
(a-f) PSNG electrode fabrication using electrode and contact chrome masks. (g) Final fabricated LOC. Size comparison is shown with a coin. (h-j) Soft lithography using a microfluidic chrome mask for fluid sample delivery.
Fabrication of the PSNG LOC immunosensor using standard conventional lithography.
(a-f) PSNG electrode fabrication using electrode and contact chrome masks. (g) Final fabricated LOC. Size comparison is shown with a coin. (h-j) Soft lithography using a microfluidic chrome mask for fluid sample delivery.
Surface modification on PSNG
PSNG surface hydroxylation
Because the PSNG structure is a non-biocompatible surface, it had to be modified to exhibit adhesion properties with biomolecules. The functionalization procedure is schematically shown in Fig 3. First, 10 ml of 0.1 M H2SO4 and 12.5 ml of H2O2 were diluted in water to a total volume of 100 ml. Then, 40 mg of EDTA were added, followed by 1.2 g of FeSO4 incrementally added to the solution, which reacted and heated up during the addition process. When the solution became bubbly and brown, the PSNG electrode was immersed for 30 min for hydroxylation to occur, which is known as a Fenton reaction. This process produces a–OH terminated group onto PSNG (Fig 3a).
Fig 3
Schematic illustration of chemical functionalization on the PSNG surface.
(a) Hydroxylation of the PSNG surface by Fenton reaction. (b) Application of the APTES cross-linker to establish amine terminated groups. (c) A prepared hCGab probe immobilized on the PSNG electrode. (d) hCG target binding with urine or reference samples. Signal transduction was analyzed in AC and DC measurements.
Schematic illustration of chemical functionalization on the PSNG surface.
(a) Hydroxylation of the PSNG surface by Fenton reaction. (b) Application of the APTES cross-linker to establish amine terminated groups. (c) A prepared hCGab probe immobilized on the PSNG electrode. (d) hCG target binding with urine or reference samples. Signal transduction was analyzed in AC and DC measurements.
hCGab immobilization on APTES
The amine-terminated PS surface was prepared by applying 1 μL of 10% APTES in water onto the PSNG electrode to establish a uniform layered surface (Fig 3b). The electrode was then incubated for 2 h and rinsed with ethanol to remove any unreacted APTES. hCGab (1 mg) was mixed with 1 ml DI water, 0.1 g of NaHCO3, the previously prepared mixture of 0.05 g tBOC and 1 ml chloroform. Then, 25 mM of EDAC and 50 mM of NHS were prepared in 0.01 M PBS solution. First, EDAC was added to the Ab solution and incubated for 20 min and then incubated in NHS for 10 min. Then, 1 μL of hCGab (8.6 mg/ml) solution was applied onto the PSNG electrode and incubated for 3 h at room temperature and humidity of 85.5% (Fig 3c). Subsequently, the remaining free surface amine groups on the PSNG electrode after Ab immobilization were blocked with 5% BSA in PBS for 10 min to avoid non-specific binding of hCG on the electrode. Then, the–NH2 protected by tBOC was removed by immersing the electrode in 10 mM TFA solution prior to detection with the hCG target [48].
hCG antigen binding
Clinical urine samples and hCG standard control solutions were reacted with the surface-modified PSNG electrode to analyze the signal transduction characteristics. Clinical samples consisting of pregnant and non-pregnant women’s urine were first analyzed in a clinical laboratory to estimate the hCG concentration in the urine and the presence of impurities. Furthermore, hCG standard control samples of 0, 8, 16, 40, 100, and 240 mIU/ml from the ELISA Kit were also applied to the PSNG electrode as a reference to validate the immunosensor. Before electrical characterization was conducted, 1 μL of hCG sample concentration was applied to the PSNG electrode and incubated for 3 h at room temperature and 85.5% humidity (Fig 3d).
PSNG Electrode Characterization
Optical and morphological characterization
Inspections of the PSNG during fabrication were performed using high-power microscopy (OLYMPUS-BX51) to make sure no contaminants were present. The fabricated nanogaps were also inspected and their gap size measured using scanning electron microscopy (SEM, JEOL JSM-6460LA). Chemical functionalization and Ag-Ab surface modification on 1 x 1 cm PS surface samples were inspected using field emission scanning electron microscopy (FESEM, Zeiss Ultra55) and Fourier transform infrared spectroscopy (FTIR, Spectrum 65 Perkin-Elmer). Then, for surface roughness and grain size measurement of PS, atomic force microscopy (AFM) (SPA400-SPI3800, Seiko Instruments Inc., Japan) was used. Furthermore, the luminescence properties of APTES were studied using photoluminescence (PL) (Horiba Fluorolog-3 for PL spectroscopy, HORIBA Jobin Yvon Inc., USA). The thin-film layer thickness and uniformity were observed using a Filmetrics F20-UV spectrometer.
Electrical characterization
Electrical measurements were conducted to analyze signals transduced by the Ag-Ab binding reaction from the PSNG electrode, which were probed via the Au contact pad on the LOC (Fig A in S1 File). AC measurements were conducted using an impedance spectrometer (Alpha-A High Performance Frequency Analyzer, Novocontrol Technologies, Hundsangen, Germany). A two-wire probe station was used with the impedance spectrometer to measure capacitance (C), permittivity (ε), loss tangent (tan δ) and conductivity (σ). Measurement was started at 0.1 V with a scaling factor of 1.4 and a frequency range of (1 x 100 to 1 x 106 Hz). The loss tangent (tan δ) value was continuously monitored during capacitance measurement to ensure it stayed below 100, indicating that the immunosensor was behaving as a capacitor. A Kiethley 6487 Picoammeter was used to measure current-voltage DC characteristics of the PSNG electrode, from which amperometric characteristics and sensitivity of the immunosensor could be analyzed. At each step during hCG detection, the PSNG electrodes were washed with DI water to ensure that measurement data were valid.
Results and Discussion
Several different sensing methods have been proposed to detect and discriminate hCG for pregnancy tests with various detection limits. Factors needed for a good sensor, such as sensitivity, selectivity, stability, reproducibility and simplicity, are critical and are achieved in the present study. Using the size expansion technique, a small nanogap size of 25 nm was achieved. Analyses were also conducted on the different gap sizes to measure their resistivity and the benefits of the small gap size. Analyses were performed using clinical urine samples and were correlated with hCG standard control samples for stability and selectivity. In addition, a single injection option facilitates triplicate data analyses, and the sensing surface can be regenerated.
Morphological analysis of the PSNG
SEM, FESEM, FTIR and AFM analyses
SEM analysis of the fabricated PSNG electrodes showed that the smallest gap obtained was 25 nm from the size expansion process (Fig 4a–4c). Both SEM and FESEM (in red boxes) images were acquired to determine the gap sizes. Because the conventional lithography method was used, uneven nanogap geometrical shapes were the outcome. However, the reproducibility of the desired nanogaps is higher, at 49%, with nanogaps fabricated on a wafer, making this a great success (Fig B in S1 File). Surface modification and Ag-Ab binding on PSNG were examined under FTIR and FESEM. It was clearly observed that binding occurs on PS grains, which proves that good surface roughness is important for Ag-Ab attachments (Fig 4d). This finding was further supported by FTIR spectrometry, which could analyze PSNG/APTES, PSNG/APTES/hCGab and PSNG/APTES/hCGab/hCG complexes (Fig C in S1 File). The peaks of C-O stretching and N-H bending were observed at 1049–1201 cm-1 and 1641–1676 cm-1, respectively, demonstrating low transmittance peaks after hCG target binding. Therefore, this spectrum confirms the presence of an additional layer over hCGab. The PSNG surface roughness properties were examined using AFM, as shown in Fig 4e, which showed an overall surface roughness (Sa) of 27.30 nm at a peak-valley (P-V) of 1.03 μm and an RMS value of 44.53 nm. However, with a small scan area, the surface roughness at the middle of the gap was Sa = 6.68 nm, and the P-V was 161.20 nm. This indicates that the fabricated PSNG structure exhibits a good surface-to-volume ratio that enhances the biocompatibility of the immunosensor [49].
Fig 4
SEM, FESEM and AFM morphological analyses of the PSNG electrode.
(a) SEM and FESEM (red box) image of 70 nm gap. (b) SEM and FESEM (red box) image of 60 nm gap (c) SEM and FESEM (red box) image of 25 nm gap (d) FESEM image of hCGab/hCG binding on PSNG electrode surface. (e) AFM images of electrode surface roughness before size expansion (right) and after size expansion (left); a final thickness of less than 100 nm and a bearing ratio of 6.86% were achieved.
SEM, FESEM and AFM morphological analyses of the PSNG electrode.
(a) SEM and FESEM (red box) image of 70 nm gap. (b) SEM and FESEM (red box) image of 60 nm gap (c) SEM and FESEM (red box) image of 25 nm gap (d) FESEM image of hCGab/hCG binding on PSNG electrode surface. (e) AFM images of electrode surface roughness before size expansion (right) and after size expansion (left); a final thickness of less than 100 nm and a bearing ratio of 6.86% were achieved.
Fluorometry analysis of polysilicon
The photoluminescence properties of PS were examined using bare PS wafers via PL spectroscopy, with excitation from a 325 nm xenon lamp at both visible (VIS) and ultraviolet (UV) regions. Fig 5a shows the PL spectra of PS, which shows emission peaks at 221 and 465 nm. It is known that the emission band in the UV region is caused by a collisional recombination process and that the emission band in the VIS region is caused by electron-hole generation, depending on ionized vacancies. Therefore, the luminescence properties at both the UV and VIS regions reveal that the crystallinity of PS is not affected by a reduction in the annealing process [50].
Fig 5
Morphological analysis and electrical characterization.
(a) Polysilicon crystalline quality observed at UV and VIS regions from fluorometry analysis. (b) Step-by-step oxide growth for size expansion to achieve the nanogap-sized electrode. (c) Effect of different gap sizes on current transduction for APTES/AB immobilization. (d) DC amperometric detection of urine samples and hCG standard reference samples.
Morphological analysis and electrical characterization.
(a) Polysilicon crystalline quality observed at UV and VIS regions from fluorometry analysis. (b) Step-by-step oxide growth for size expansion to achieve the nanogap-sized electrode. (c) Effect of different gap sizes on current transduction for APTES/AB immobilization. (d) DC amperometric detection of urine samples and hCG standard reference samples.
Size expansion technique by oxide growth
The size expansion technique was used to reduce the gap size of PS. The initial patterned design on the silicon substrate had a 1 μm gap. During thermal oxidation, the patterned PS structure expanded as oxygen was being supplied (Fig 5b). After oxidation, the samples were immersed in BOE solution to remove excess oxide, leaving only the expanded underlying PS structure. Oxidation time reduced from 60 to 20 min (Fig 4a–4c) as the gap size reduced. The film thickness during each oxidation process was measured using Filmetrics Spectrometer F20-UV. Regression analysis of the oxidation process showed a linear regression, R2, of 0.8829 and linearity, y = 35.906x + 898.38, was achieved, where x is growth time and y is the film thickness. With the narrow nanogap obtained on our sensor, the effect of electrode polarization could be reduced, thus resulting in a negligible noise effect [37,51].
Effect of gap size on PSNG sensitivity
The DC electrical characteristics of the PSNG electrode were examined during chemical functionalization and attachment of APTES and hCGab on the fabricated PSNG. Fig 5c illustrates the investigation of lg = 75, 60 and 25 nm gap sizes, which were applied with 1 μL of prepared 8.6 mg/ml hCGab on APTES-modified PSNG. The graph shows that the resistance drops gradually when the gap size decreases from 75 to 60 to 25 nm, and the measured resistances were 65.7, 55.7, and 22.6 μΩ, respectively, as proven by Eqs 1 and 2. Therefore, to enhance the sensitivity of the electrode, a gap size of lg = 25 nm was selected for hCG measurement because it exhibits low resistance (S1 File).
after surface modification with APTES and hCGab,It is also known that a small gap size, lg, could reduce the electrode polarization effect, which is a major source of error in nanogap electrodes. Therefore, it was suggested that lg less than 100 nm could reduce the effect because the accumulation of ions known as the Debye layer at the smallest gap showed negligible error. Hence, reduction by lg causes the Debye layer to overlap, and the potential drop could be reduced [37,52].
DC and AC measurements on PSNG electrodes—clinical sample analyses
Current-voltage characteristics
The prepared PSNG electrode functionalized with APTES followed by hCGab immobilization was tested with pregnant and non-pregnant women’s urine samples, followed by hCG standard control samples. Each modified layer on the PSNG surface exhibits ohmic characteristics that influence the electrode resistivity according to RT = (ρTlT)/A. A total of six samples were tested (n = 6) for hCG, with three samples each from pregnant and non-pregnant women. In the pregnant women’s urine samples, the gestational ages of pregnancy were 4, 9 and 12 weeks, measured as 970, 16,200 and 230,000 mIU/ml, respectively, as obtained from a clinical laboratory. These results clearly indicated that there were no non-specific interferences by albumin, epithelial or glucose in the samples (Table 2). Therefore, by testing with our PSNG immunosensor, it was observed that an increase in hCG concentrations caused a reduction in current with both urine and hCG standard control samples (Fig 5d).
Table 2
Clinical analyses of urine samples using quantitative radioimmunoassays (RIA).
Pregnant women
Non-pregnant women
Sample
1
2
3
1
2
3
Gestational age
4 weeks
9 weeks
12 weeks
-
-
-
Albumin
Nd
Nd
Nd
Nd
Nd
Nd
Epithelial
Nd
Nd
Nd
Nd
Nd
Nd
Glucose level
Nd
Nd
Nd
Nd
Nd
Nd
Blood
No
No
No
No
No
No
hCG (mIU/ml)
970 ± 3.1
16200 ± 2.4
230000 ± 1.7
Nil
Nil
Nil
Note: Nd = Not detected.
Note: Nd = Not detected.Therefore, changes in hCG current, ΔIhCG, were due to the changes in resistivity of the hCG concentration, ΔρhCG, as proven in Eq 3, while other parameters remained constant (S1 File). It can thus be inferred that there is a binding event between hCGab and hCG on the surface of the PSNG electrode, which can transduce current based on hCG concentration [53].
Capacitance-frequency characteristics
During AC detection, each modification layer on the PSNG surface was signified as a multilayer dielectric material, and both PSNG electrodes were signified as parallel plate capacitors, thus behaving as a capacitive immunosensor based on capacitance, C = ɛrɛoA/d. This distinctive property of the PSNG electrode was used in this study and was clearly reflected in the results obtained from capacitance, permittivity (Fig D in S1 File) and conductivity (Fig E in S1 File) signals upon hCG target binding on the electrodes. The capacitance and permittivity properties of PSNG increased when urine samples with higher hCG concentrations were applied, as shown in Eq 4. The changes in hCG concentration cause the change in hCG permittivity, ΔɛhCG, which eventually increases the hCG capacitance, ChCG.Therefore, those samples in which hCG has reacted with hCGab, forming a peptide bond, correspond to an increment in ChCG of the PSNG electrode observed in the range 80 to 100 nF with the pregnant women’s urine samples [54]. However, samples without hCG do not exhibit this characteristic and exhibit ChCG in the range of 30 to 40 nF. The sensor ɛhCG values ranged from 160 F/m to 185 F/m for pregnant samples. In addition, the measured capacitance is not only due to the sample capacitance but also influenced by the sample conductance, as shown in Eq 5. The sensor conductivity values ranged from 80 nS/cm to 100 nS/cm with increasing hCG levels in the pregnant urine samples. Lonappan et al. [55] and Sánchez et al. [56] have reported that an increment in hCG levels leads to an increased level of conductivity. Hence, in the sensor operation, it can be deduced that the measured capacitance values were due to the dielectric term and were also related to sample conductivity [57].
PSNG electrode analytical validation for specificity, sensitivity and stability
The obtained experimental results are highly valuable in calculating the performance of the sensor and its capabilities. In the present investigation, various analyses, such as Ag-Ab specific binding, sensitivity, selectivity, limit of detection (LOD), reproducibility and stability were carried out and are shown in Fig 6. During these tests, a non-specific blocking agent, BSA, was used to ensure that only hCG molecules bind to the hCGab probe. Therefore, one-site specific binding was analyzed, the equilibrium dissociation constant (Kd) was calculated to be 2.23 ± 0.66 mIU, and the maximum number of binding sites per antigen would be Bmax = 22.54 ± 1.46 mIU, as indicated in Fig 6a and Eq 6, where x is reaction time and y is hCG concentration.
Fig 6
Analytical performance of the PSNG LOC immunosensor.
(a) One-site specific binding to determine the protein dissociation constant Kd. (b) Linearity fitting curves for hCG standards and urine samples. (c) Selectivity test with interfering species. CEA, AA, UA, glucose, non-pregnant and pregnant samples were used. The top inset is the AC conductivity test to determine the detection limit and the bottom inset shows that the immunosensor remains stable for 22 days. (d) Reproducibility test of signal transduction on PSNG electrodes with pregnant urine samples.
Analytical performance of the PSNG LOC immunosensor.
(a) One-site specific binding to determine the protein dissociation constant Kd. (b) Linearity fitting curves for hCG standards and urine samples. (c) Selectivity test with interfering species. CEA, AA, UA, glucose, non-pregnant and pregnant samples were used. The top inset is the AC conductivity test to determine the detection limit and the bottom inset shows that the immunosensor remains stable for 22 days. (d) Reproducibility test of signal transduction on PSNG electrodes with pregnant urine samples.These results indicate a good affinity for hCG detection and the importance of sensitive PSNG electrodes in biosensing [58-60]. When a wide range of hCG concentrations were tested, sensitivity analysis determined the sensor’s ability to transduce significant currents. The linearity achieved for hCG standards was
with a correlation coefficient R2 = 0.9943, while for urine samples it was
with R2 = 0.9279 (Fig 6b). Therefore, our sensor exhibits a sensitivity of 835.88 μA mIU-1 ml-2 cm-2 at the active sensing area, with ‘A’ of 0.00262 cm2 and a gradient of 2.19 μA mIU-1 ml-2, as calculated from Eq 9. The proposed sensitivity is higher than the reported values of 0.255 μA mIU-1 ml-2 by Tan et al. [26] and 0.1032 μA mIU-1 ml-2 by Li et al. [27]. Here, the gradients of the urine sample plots are larger than standard reference samples, which could be due to a non-linear concentration range in the urine samples. Therefore, the gradients indicate normal behavior of urine samples in that they do not contain only hCG, unlike the ideal reference hCG samples. This proves that, even though our immunosensor is highly selective, the linearity depends on the type of sample. Therefore, this sensor could also measure hCG that exists in a different medium. Because the hCG concentration detected in urine samples is larger than that in the hCG standards, linearity proves that both slopes decrease as the concentration increases. This result has been further confirmed by plotting a curve that theoretically fits both the urine and standard samples.Selectivity tests were carried out by using interfering species, such as CEA, AA, UA and glucose (GLUC), with 8 mIU/ml of hCG (Fig 6c). It can be observed that the response of interfering species on our PSNG electrode is less than 10 nS/cm at the lowest concentration of hCG. In addition, the LOD of the sensor was measured at 0.28 mIU/ml (27.78 pg/ml) at 3σ from PSNG electrode conductivity at a frequency of 1 MHz. To test this, a smaller concentration of the hCG target (0.1 mIU/ml) was prepared and its minimum conductivity saturation point was observed at 5.13 nS/cm, as shown in the top inset of (Fig 6c(i)). Hence, our LOD is proven to be considerably better than that of the nanoporous gold and graphene sensor reported by Li et al., at 0.34 mIU/ml [14] and polyethylene glycol microbeads reported by Zhu et al at 0.38 mIU/mL [61], as shown in Table 3. The Fig 6c(ii) inset shows that the stability of the present LOC decreases to 98.5% after 22 days of testing with 8 mIU/ml of hCG. This is acceptable for a sensor and comparable with the 7.1% decrease after 1 month using a MWCNT modified immunoassay, as reported by Na li et al. [27], and a 14% decrease after 3 weeks using a nanoporous gold and graphene sensor, as reported by Ru Li et al. [14].
Table 3
Comparison of different hCG immunosensors.
Electrode Matrix
Label
Linear Range
Detection Limit
Reference
PSNG/APTES/ hCGab/hCG
No
8–230,000 mIU/mL
0.28 mIU/mL
This work
*GCEa/GSb/NPGc/ hCGab/BSA/hCG
No
500–40,000 mIU/mL
0.34 mIU/mL
2011 [14]
*PEGd microbeads/ hCGab/hCG
No
0.625–40 mIU/ml
0.38 mIU/mL
2015 [61]
Epitaxial graphene/ APTES/ hCGab/hCG
No
620–5620 mIU/mL
3.1 mIU/mL
2014 [13]
*GCE/ Pt-Aue alloy nanotube array/ hCGab/hCG
No
25 to 400 mIU/mL
12 mIU/mL
2010 [15]
*Graphene-based CRETf immunoassay
Yes
0.1–10 mIU/mL
0.06 mIU/mL
2014 [65]
*Note:
aGlassy Carbon Electrode,
bGraphene Sheets,
cNanoporous Gold,
dPolyethylene Glycol,
ePlatinum-Gold,
fChemiluminescence Resonance Energy Transfer.
*Note:aGlassy Carbon Electrode,bGraphene Sheets,cNanoporous Gold,dPolyethylene Glycol,ePlatinum-Gold,fChemiluminescence Resonance Energy Transfer.Reproducibility between LOCs was also tested to ensure intersensor signal accuracy. Fig 6d shows that there is little variation in capacitance with measured values of replica samples. However, it does not affect the electrode sensitivity, in which the relative standard deviations (RSDs) of 4.3, 3.3 and 4.9% changes in capacitance at 4, 9 and 12 weeks of pregnancy, respectively, indicate good reproducibility. This could be because the surface and geometrical factor of the PSNG electrode provide a good platform for biosensing [62]. Furthermore, for regeneration purposes, the PSNG sensor may also be reusable after washing it in urea to remove the bonded Ag target on immobilized APTES/hCGab [30]. This strategy enables the reuse of the sensing surface for multiple samples with tripartite sensing.
Commercial potential of POC PSNG immunosensor
Our lab-on-chip immunosensor with electrical readouts offers several advantages over immunoassays due to its reduced cost, reduced form factor and ease of signal acquisition. The costs associated with laboratory equipment, chemicals and labeling procedures for hCG diagnosis via immunoassay methods could be reduced using POC immunosensors [63-65]. Because our POC PSNG is fabricated with multiple electrodes, it could facilitate multiplex detection from a single sample, which reduces cost and diagnosis time, as demonstrated in other sensing systems [33,66,67].Recently, a study on the sensitivity and limitations of commercially available qualitative POC pregnancy test kits was reported by Greene et al. Two commonly used hCG POC tests, OSOM and QuickVue+, were examined, and the authors concluded that both devices have a detection range of 20–300 IU/ml with urine samples, which is insufficient for early pregnancy diagnosis. Furthermore, the mean gestational age for specimens corresponding to the false negative results was 4 weeks [68]. Siemens Immulite 1000 hCG Assay shows detection limit at 2.0 mIU/mL [69]. With our method, comparable sensitivity and specificity could be achieved from electrical transductions of the binding activity on PSNG, as shown in Table 4. Moreover, the gap size of 25 nm on the proposed lab-on-chip sensor is minimal compared to other commonly available lab-on-chip systems. In addition, the tripartite flow-channel minimizes the sample usage with uniform flow for parallel analyses. Overall, the PSNG electrode immunosensor used in the current study exhibited higher performance and a more facile fabrication process than other available systems, and the sensor can be readily fit with commercially available detection systems.
Table 4
Comparisons with commercially available POC pregnancy test kits.
POC devices
Analysis Method
Linear Range
Detection Limit
Reference
PSNG/APTES/ hCGab/hCG
Quantitative
8–230,000 mIU/mL
0.28 mIU/mL
This work
Siemens Immulite 1000 hCG Assay
Qualitative
2–3500 mIU/mL
2.0 mIU/mL
2013 [69]
QuickVue+
Qualitative
20–300 mIU/mL
>20 mIU/mL
2013 [68]
OSOM
Qualitative
20–300 mIU/mL
>20 mIU/mL
2013 [68]
Conclusion
The humanchorionic gonadotrophin (hCG) level in pregnant women is usually low during early pregnancy and increases drastically during the second trimester. As detection of hCG in the clinical urine sample is critical during pregnancy, we have demonstrated the detection of hCG with a surface functionalized PSNG electrode immunosensor fabricated by a conventional lithography and size expansion technique. Our lab-on-chip sensor exhibited good stability, high sensitivity (835.88 μA mIU-1 ml-2 cm-2), good reproducibility, high selectivity and a low detection limit (0.28 mIU/ml, or 27.78 pg/ml). Due to the narrow nanogap (25 nm) fabricated on this immunosensor, it shows high performance with a good signal-to-noise ratio, and it can be applied for multiplexed POC diagnosis, further reducing expensive equipment usage. Moreover, the dielectric analyses on the PSNG electrodes with clinicalhuman urine samples provide reliable measurements and stability against regeneration reagents, which favors multiple sample analyses on the tripartite channel with the option of parallel runs.
This file contains.
Fig A. (a) Electrical characterization setup for AC measurements using a frequency analyzer and (b) DC measurements using the Kiethley 6487 Picoammeter. Fig B. Reproducibility distribution of 25 nm gap on a single wafer per lithography. Of the fabricated electrodes, 49% have a gap sized of 25 nm. Fig C. FTIR spectrometry analyzed with PSNG/APTES, PSNG/APTES/hCGab and PSNG/APTES/hCGab/hCG samples. Fig D. AC capacitance and permittivity (inset) measurements for pregnant and non-pregnant women’s urine. Fig E. Conductivity measurements for pregnant and non-pregnant women’s urine.(DOCX)Click here for additional data file.
Authors: Dina N Greene; Robert L Schmidt; Sandy M Kamer; David G Grenache; Carolyn Hoke; Thomas S Lorey Journal: Clin Chim Acta Date: 2012-11-15 Impact factor: 3.786
Authors: Frances L Cate; Courtney Moffett; Ann M Gronowski; David G Grenache; Katherine E Hartmann; Alison Woodworth Journal: Clin Chim Acta Date: 2013-03-05 Impact factor: 3.786
Authors: Veeradasan Perumal; Uda Hashim; Subash C B Gopinath; R Haarindraprasad; Wei-Wen Liu; P Poopalan; S R Balakrishnan; V Thivina; A R Ruslinda Journal: PLoS One Date: 2015-12-22 Impact factor: 3.240
Authors: Veeradasan Perumal; Uda Hashim; Subash C B Gopinath; Haarindraprasad Rajintra Prasad; Liu Wei-Wen; S R Balakrishnan; Thivina Vijayakumar; Ruslinda Abdul Rahim Journal: Nanoscale Res Lett Date: 2016-01-19 Impact factor: 4.703
Authors: Alisa N Kozitsina; Tatiana S Svalova; Natalia N Malysheva; Andrei V Okhokhonin; Marina B Vidrevich; Khiena Z Brainina Journal: Biosensors (Basel) Date: 2018-04-01