Nationwide Laboratory Adherence to Myocardial Perfusion Imaging Radiation Dose Reduction Practices: A Report From the Intersocietal Accreditation Commission Data Repository.

Scott D Jerome1, Peter L Tilkemeier2, Mary B Farrell3, Leslee J Shaw4.   

Abstract

OBJECTIVES: This study sought to examine current laboratory practices for radiation effective doses for myocardial perfusion imaging (MPI) and laboratory adherence to guideline-directed radiation reduction practices.
BACKGROUND: A recent focus on radiation dose reduction for cardiovascular imaging has led to several published guidelines and consensus statements detailing performance metrics for laboratory practices. We sought to examine laboratory adherence to optimized radiation dose protocol recommendations among 5,216 submitted cases from 1,074 MPI laboratories evaluated for Intersocietal Accreditation Commission accreditation.
METHODS: Eligible imaging centers included MPI laboratories enrolled in the Intersocietal Accreditation Commission data repository of accreditation applications from 2012 to 2013. Accreditation requires submission of 3 to 5 cases for evaluation of a range of representative cases. Based on standard dosimetry for rest and stress MPI, an effective dose (in millisieverts) was calculated. Model simulations were performed to estimate guideline-directed effective doses.
RESULTS: The average effective dose was 14.9 ± 5.8 mSv (range 1.4 to 42.4 mSv). A 1-day technetium Tc 99m protocol was used in 82.9% of cases, whereas a 2-day technetium Tc 99m and dual isotope protocol was used in 7.5% of submitted cases. Only 1.5% of participating imaging centers met current guidelines for an average laboratory radiation exposure ≤9 mSv, whereas 10.1% of patient effective doses were >20.0 mSv. A model simulation replacing the radiation exposure of dual isotope MPI with that of a 1-day technetium Tc 99m protocol reduced the proportion of patients receiving an effective dose >20 mSv to only 2.7% of cases (p < 0.0001).
CONCLUSIONS: Mandatory laboratory accreditation for MPI allows for examination of current radiation dosimetry practices. Current guidelines for reduced patient-specific radiation exposure are rarely implemented, with few laboratories meeting recommendations of ≤9 mSv for 50% of patients. Increased educational efforts and the development of performance measures for laboratory accreditation may be required to meet current radiation dose-reduction standards.
Copyright © 2015 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  laboratory accreditation; myocardial perfusion imaging; performance measures; radiation exposure

Mesh:

Substances:

Year:  2015        PMID: 26363837     DOI: 10.1016/j.jcmg.2015.07.008

Source DB:  PubMed          Journal:  JACC Cardiovasc Imaging        ISSN: 1876-7591


  15 in total

1.  Myocardial perfusion SPECT just keeps getting better and better.

Authors:  E Gordon DePuey
Journal:  J Nucl Cardiol       Date:  2016-02-16       Impact factor: 5.952

Review 2.  Relationship between Calcium Score and Myocardial Scintigraphy in the Diagnosis of Coronary Disease.

Authors:  Fabio Paiva Rossini Siqueira; Claudio Tinoco Mesquita; Alair Augusto Sarmet M Damas Dos Santos; Marcelo Souto Nacif
Journal:  Arq Bras Cardiol       Date:  2016-07-18       Impact factor: 2.000

3.  Personalized Models for Injected Activity Levels in SPECT Myocardial Perfusion Imaging.

Authors:  Albert Juan Ramon; Yongyi Yang; P Hendrik Pretorius; Karen L Johnson; Michael A King; Miles N Wernick
Journal:  IEEE Trans Med Imaging       Date:  2018-12-06       Impact factor: 10.048

4.  Current status of stress myocardial perfusion imaging pharmaceuticals and radiation exposure in Japan: Results from a nationwide survey.

Authors:  Ryuto Otsuka; Narumi Kubo; Yosuke Miyazaki; Mio Kawahara; Jun Takaesu; Kazuki Fukuchi
Journal:  J Nucl Cardiol       Date:  2017-03-28       Impact factor: 5.952

5.  Time to get on the stress-only bandwagon?

Authors:  Matthew W Parker; Gary V Heller
Journal:  J Nucl Cardiol       Date:  2016-10-14       Impact factor: 5.952

6.  Evaluation of the effect of reducing administered activity on assessment of function in cardiac gated SPECT.

Authors:  Albert Juan Ramon; Yongyi Yang; Miles N Wernick; P Hendrik Pretorius; Karen L Johnson; Piotr J Slomka; Michael A King
Journal:  J Nucl Cardiol       Date:  2018-11-07       Impact factor: 5.952

7.  Which SPECT for today, which SPECT for tomorrow?

Authors:  Milena J Henzlova; W Lane Duvall
Journal:  J Nucl Cardiol       Date:  2016-04-12       Impact factor: 5.952

8.  It's about time we think about lowering radiation dose in obese patients too.

Authors:  Daniel A Kim; Mary Beth Farrell; Scott D Jerome
Journal:  J Nucl Cardiol       Date:  2016-08-19       Impact factor: 5.952

9.  Safety of coronary CT angiography and functional testing for stable chest pain in the PROMISE trial: A randomized comparison of test complications, incidental findings, and radiation dose.

Authors:  Michael T Lu; Pamela S Douglas; James E Udelson; Elizabeth Adami; Brian B Ghoshhajra; Michael H Picard; Rhonda Roberts; Kerry L Lee; Andrew J Einstein; Daniel B Mark; Eric J Velazquez; William Carter; Michael Ridner; Hussein R Al-Khalidi; Udo Hoffmann
Journal:  J Cardiovasc Comput Tomogr       Date:  2017-08-15

10.  Drivers of radiation dose reduction with myocardial perfusion imaging: A large health system experience.

Authors:  Firas J Al Badarin; John A Spertus; Timothy M Bateman; Krishna K Patel; Eric V Burgett; Kevin F Kennedy; Randall C Thompson
Journal:  J Nucl Cardiol       Date:  2019-01-31       Impact factor: 5.952