Contemporary Oxygenator Design Relative to Hemolysis.

Leonie H Venema, Ajay S Sharma, Antoine P Simons, Otto Bekers, Patrick W Weerwind.   

Abstract

Hemolysis is a well-known phenomenon during cardiovascular surgery and generally attributed to cardiopulmonary bypass, particularly when using high-resistant oxygenators. This study aimed at investigating whether transoxygenator pressure drop can be considered an independent factor of hemolysis. Additionally, intraoxygenator blood distribution and shear stress were assessed. A low-resistant (LR, n = 3), a moderate-resistant (MR, n = 3), and a high-resistant (HR, n = 3) clinically used membrane oxygenator were tested in vitro using a roller pump and freshly drawn heparinized porcine blood. Flow rates were set to 2 and 4 L/min and maximum flow compliant to the oxygenator type for 1 hour each. As a control, the oxygenator was excluded from the system. Blood samples were taken every 30 minutes for plasma-free hemoglobin assay and transoxygenator pressure was measured inline. Intraoxygenator blood distribution was assessed using an ultrasound dilution technique. Despite the relatively broad spectrum of pressure drop and resultant transoxygenator pressure drops (LR: 14-41 mmHg, MR: 29-115 mmHg, HR: 77-284 mmHg, respectively), no significant association (R2 = .074, p = .22) was found with the normalized index of hemolysis. The shear stress of each oxygenator at maximum flow rate amounted to 3.0 N/m2 (LR), 5.7 N/m2 (MR), and 8.4 N/m2 (HR), respectively. Analysis of blood flow distribution curves (kurtosis and skewness) revealed intraoxygenator blood flow distribution to become more homogeneous when blood flow rates increased. Contemporary oxygenators were shown not to be a predominant factor for red blood cell damage.

Entities:  

Mesh:

Year:  2014        PMID: 26357786      PMCID: PMC4566829     

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  14 in total

1.  Mechanical trauma in leukocytes.

Authors:  T S Dewitz; T C Hung; R R Martin; L V McIntire
Journal:  J Lab Clin Med       Date:  1977-10

2.  Improved computational fluid dynamic simulations of blood flow in membrane oxygenators from X-ray imaging.

Authors:  Cameron C Jones; James M McDonough; Patrizio Capasso; Dongfang Wang; Kyle S Rosenstein; Joseph B Zwischenberger
Journal:  Ann Biomed Eng       Date:  2013-05-15       Impact factor: 3.934

3.  Clinical evaluation of five commercially available adult oxygenators in terms of pressure drop during normothermic and hypothermic cardiopulmonary bypass.

Authors:  Bingyang Ji; Hongyue Wang; Na Miao; Jialin Xing; Wei Liu; Ruifang Liu; Cun Long
Journal:  Int J Artif Organs       Date:  2010-05       Impact factor: 1.595

4.  Hemolytic effects of energy dissipation in flowing blood.

Authors:  M Bluestein; L F Mockros
Journal:  Med Biol Eng       Date:  1969-01

5.  PI 3-kinase p110beta: a new target for antithrombotic therapy.

Authors:  Shaun P Jackson; Simone M Schoenwaelder; Isaac Goncalves; Warwick S Nesbitt; Cindy L Yap; Christine E Wright; Vijaya Kenche; Karen E Anderson; Sacha M Dopheide; Yuping Yuan; Sharelle A Sturgeon; Hishani Prabaharan; Philip E Thompson; Gregg D Smith; Peter R Shepherd; Nathalie Daniele; Suhasini Kulkarni; Belinda Abbott; Dilek Saylik; Catherine Jones; Lucy Lu; Simon Giuliano; Sascha C Hughan; James A Angus; Alan D Robertson; Hatem H Salem
Journal:  Nat Med       Date:  2005-04-17       Impact factor: 53.440

6.  Shear-induced activation of platelets.

Authors:  J M Ramstack; L Zuckerman; L F Mockros
Journal:  J Biomech       Date:  1979       Impact factor: 2.712

7.  Hemolytic characteristics of oxygenators during clinical extracorporeal membrane oxygenation.

Authors:  Shinji Kawahito; Tomohiro Maeda; Tadashi Motomura; Hiroshi Ishitoya; Tamaki Takano; Kenji Nonaka; Joerg Linneweber; Seiji Ichikawa; Masaki Kawamura; Kazuhiro Hanazaki; Julie Glueck; Yukihiko Nosé
Journal:  ASAIO J       Date:  2002 Nov-Dec       Impact factor: 2.872

8.  Clinical evaluation of nine hollow-fibre membrane oxygenators.

Authors:  P A Segers; J F Heida; I de Vries; C Maas; A J Boogaart; S Eilander
Journal:  Perfusion       Date:  2001-03       Impact factor: 1.972

9.  Haemolysis of various mammalian erythrocytes in sodium chloride, glucose and phosphate-buffer solutions.

Authors:  T Matsuzawa; Y Ikarashi
Journal:  Lab Anim       Date:  1979-10       Impact factor: 2.471

10.  Selective inhibition of the platelet phosphoinositide 3-kinase p110beta as promising new strategy for platelet protection during extracorporeal circulation.

Authors:  Andreas Straub; Hans Peter Wendel; Klaus Dietz; Daniela Schiebold; Karlheinz Peter; Simone M Schoenwaelder; Gerhard Ziemer
Journal:  Thromb Haemost       Date:  2008-03       Impact factor: 5.249

View more
  4 in total

1.  Goal-Directed Perfusion Methodology for Determining Oxygenator Performance during Clinical Cardiopulmonary Bypass.

Authors:  Alfred H Stammers; Randi Miller; Stephen G Francis; Laszlo Fuzesi; Anthony Nostro; Eric Tesdahl
Journal:  J Extra Corpor Technol       Date:  2017-06

Review 2.  In Vitro models for thrombogenicity testing of blood-recirculating medical devices.

Authors:  Deepika N Sarode; Shuvo Roy
Journal:  Expert Rev Med Devices       Date:  2019-06-10       Impact factor: 3.166

3.  Is There a Relationship between Pressure Gradients through Contemporary Oxygenators and Immune Cell Proliferation during Cardiopulmonary Bypass? A Pilot Study.

Authors:  Roger D P Stanzel; Mark Henderson
Journal:  J Extra Corpor Technol       Date:  2017-09

Review 4.  ECMO Membrane Lung Failure due to Hypertriglyceridemia: A Case Report and Review of the Literature.

Authors:  Jenelle H Badulak; Eleanor Curtis; Eileen M Bulger
Journal:  J Extra Corpor Technol       Date:  2020-09
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.