Literature DB >> 26331110

Four-year clinical follow-up of the first-in-man randomized comparison of a novel sirolimus eluting stent with abluminal biodegradable polymer and ultra-thin strut cobalt-chromium alloy: the INSPIRON-I trial.

Marcos Danillo Peixoto Oliveira1, Expedito E Ribeiro1, Carlos M Campos1, Henrique B Ribeiro1, Bruno L R Faillace1, Augusto C Lopes1, Rodrigo B Esper1, George X Meirelles1, Marco A Perin1, Alexandre Abizaid1, Pedro A Lemos1.   

Abstract

BACKGROUND: The Inspiron™ sirolimus-eluting stent (SES) is a low-dose, ultra-thin-strut cobalt-chromium stent abluminally coated with biodegradable polymers (BP). Previous results from the INSPIRON-I trial, a first-in-man study, have proven the efficacy of the novel stent in reducing neointimal proliferation. The present report aims at evaluating the long-term clinical outcomes of patients enrolled into the INSPIRON-I trial (Clinical Trials Gov. identifier: NCT01093391).
METHODS: A total of 57 patients (60 lesions) were randomly allocated in a 2:1 ratio to treatment with the Inspiron™ SES vs. its equivalent Cronus™ bare metal stent (BMS, both by Scitech Medical™, Aparecida de Goiânia, Goiás, Brazil), in four tertiary centers. The primary endpoint of the present analysis was the occurrence of major adverse cardiac events (MACE) [death, myocardial infarction (MI), target vessel revascularization (TVR) and/or target lesion revascularization (TLR)] at 4 years.
RESULTS: Baseline clinical and angiographic characteristics of both groups were similar. After 4 years, the primary endpoint occurred in 7.9% vs. 23.5% of patients in Inspiron and control groups respectively (P=0.11). The rate of death/MI was similar between the groups, but there was a significant decrease in the risk of repeat revascularization in the Inspiron group compared to the control arm TLR (0.0% vs. 23.5% respectively, P=0.02). There were no stent thromboses in the study population.
CONCLUSIONS: The novel Inspiron™ SES showed a sustained safe and effective clinical profile after 4-year of follow-up, with very low adverse events and null stent thrombosis (ST) occurrence.

Entities:  

Keywords:  Biodegradable polymer (BP); long-term outcomes; percutaneous coronary intervention (PCI); sirolimus-eluting stent (SES)

Year:  2015        PMID: 26331110      PMCID: PMC4536482          DOI: 10.3978/j.issn.2223-3652.2015.07.05

Source DB:  PubMed          Journal:  Cardiovasc Diagn Ther        ISSN: 2223-3652


  22 in total

1.  In-stent restenosis in small coronary arteries: impact of strut thickness.

Authors:  Carlo Briguori; Cristiano Sarais; Paolo Pagnotta; Francesco Liistro; Matteo Montorfano; Alaide Chieffo; Fabio Sgura; Nicola Corvaja; Remo Albiero; Goran Stankovic; Costantinos Toutoutzas; Erminio Bonizzoni; Carlo Di Mario; Antonio Colombo
Journal:  J Am Coll Cardiol       Date:  2002-08-07       Impact factor: 24.094

2.  Safety and efficacy of drug-eluting and bare metal stents: comprehensive meta-analysis of randomized trials and observational studies.

Authors:  Ajay J Kirtane; Anuj Gupta; Srinivas Iyengar; Jeffrey W Moses; Martin B Leon; Robert Applegate; Bruce Brodie; Edward Hannan; Kishore Harjai; Lisette Okkels Jensen; Seung-Jung Park; Raphael Perry; Michael Racz; Francesco Saia; Jack V Tu; Ron Waksman; Alexandra J Lansky; Roxana Mehran; Gregg W Stone
Journal:  Circulation       Date:  2009-06-15       Impact factor: 29.690

3.  Biodegradable polymer drug-eluting stents reduce the risk of stent thrombosis at 4 years in patients undergoing percutaneous coronary intervention: a pooled analysis of individual patient data from the ISAR-TEST 3, ISAR-TEST 4, and LEADERS randomized trials.

Authors:  Giulio G Stefanini; Robert A Byrne; Patrick W Serruys; Antoinette de Waha; Bernhard Meier; Steffen Massberg; Peter Jüni; Albert Schömig; Stephan Windecker; Adnan Kastrati
Journal:  Eur Heart J       Date:  2012-03-24       Impact factor: 29.983

4.  Rationale and design for the PAINT randomized trial.

Authors:  Pedro A Lemos; Bruno Moulin; Marco A Perin; Ludmilla A R R Oliveira; J Airton Arruda; Valter C Lima; Antonio A G Lima; Paulo R A Caramori; Cesar R Medeiros; Mauricio R Barbosa; Fabio S Brito; Expedito E Ribeiro; Eulógio E Martinez
Journal:  Arq Bras Cardiol       Date:  2009-12       Impact factor: 2.000

Review 5.  Clinical outcomes with bioabsorbable polymer- versus durable polymer-based drug-eluting and bare-metal stents: evidence from a comprehensive network meta-analysis.

Authors:  Tullio Palmerini; Giuseppe Biondi-Zoccai; Diego Della Riva; Andrea Mariani; Manel Sabaté; Pieter C Smits; Christoph Kaiser; Fabrizio D'Ascenzo; Giacomo Frati; Massimo Mancone; Philippe Genereux; Gregg W Stone
Journal:  J Am Coll Cardiol       Date:  2013-11-06       Impact factor: 24.094

Review 6.  Polymer coatings and delayed arterial healing following drug-eluting stent implantation.

Authors:  R A Byrne; M Joner; A Kastrati
Journal:  Minerva Cardioangiol       Date:  2009-10       Impact factor: 1.347

7.  Randomized evaluation of two drug-eluting stents with identical metallic platform and biodegradable polymer but different agents (paclitaxel or sirolimus) compared against bare stents: 1-year results of the PAINT trial.

Authors:  Pedro A Lemos; Bruno Moulin; Marco A Perin; Ludmilla A R R Oliveira; J Airton Arruda; Valter C Lima; Antonio A G Lima; Paulo R A Caramori; Cesar R Medeiros; Mauricio R Barbosa; Fabio S Brito; Expedito E Ribeiro; Eulógio E Martinez
Journal:  Catheter Cardiovasc Interv       Date:  2009-11-01       Impact factor: 2.692

Review 8.  Vascular responses to drug eluting stents: importance of delayed healing.

Authors:  Aloke V Finn; Gaku Nakazawa; Michael Joner; Frank D Kolodgie; Erik K Mont; Herman K Gold; Renu Virmani
Journal:  Arterioscler Thromb Vasc Biol       Date:  2007-05-17       Impact factor: 8.311

9.  Late clinical outcomes after implantation of drug-eluting stents coated with biodegradable polymers: 3-year follow-up of the PAINT randomised trial.

Authors:  Pedro A Lemos; Bruno Moulin; Marco A Perin; Ludmilla A R R Oliveira; J Airton Arruda; Valter C Lima; Antonio A G Lima; Paulo R A Caramori; Cesar R Medeiros; Mauricio R Barbosa; Fabio S Brito; Expedito E Ribeiro
Journal:  EuroIntervention       Date:  2012-05-15       Impact factor: 6.534

10.  Randomized trial of three rapamycin-eluting stents with different coating strategies for the reduction of coronary restenosis.

Authors:  Julinda Mehilli; Robert A Byrne; Anna Wieczorek; Raisuke Iijima; Stefanie Schulz; Olga Bruskina; Jürgen Pache; Rainer Wessely; Albert Schömig; Adnan Kastrati
Journal:  Eur Heart J       Date:  2008-06-11       Impact factor: 29.983

View more
  3 in total

1.  Intravascular imaging comparison of two metallic limus-eluting stents abluminally coated with biodegradable polymers: IVUS and OCT results of the DESTINY trial.

Authors:  J Ribamar Costa; Daniel Chamié; Alexandre A C Abizaid; Expedito Ribeiro; George C Meireles; Maurício Prudente; Carlos A Campos; Juliana P Castro; Ricardo Costa; Pedro A Lemos
Journal:  Int J Cardiovasc Imaging       Date:  2016-10-06       Impact factor: 2.357

2.  Current State of Bioabsorbable Polymer-Coated Drug-Eluting Stents.

Authors:  Abhilash Akinapelli; Jack P Chen; Kristine Roy; Joseph Donnelly; Keith Dawkins; Barbara Huibregtse; Dongming Hou
Journal:  Curr Cardiol Rev       Date:  2017

3.  One-Year Clinical Outcome of Inspiron Stent in All-Comers Population (Analysis from 790 Consecutive Patients).

Authors:  Felipe Falcão; Fabiano Cantarelli; Rodrigo Cantarelli; Flávio Mota; Manuela Navarro; Henrique Mota; Martinelly Santos; Daniel Cruz; André Sansônio; Marcelo Parente; Flávio Oliveira
Journal:  J Interv Cardiol       Date:  2020-12-17       Impact factor: 2.279

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.