Technical Note: Validation of halo modeling for proton pencil beam spot scanning using a quality assurance test pattern.

Liyong Lin1, Sheng Huang1, Minglei Kang1, Timothy D Solberg1, James E McDonough1, Christopher G Ainsley1.   

Abstract

PURPOSE: The purpose of this paper is to demonstrate the utility of a comprehensive test pattern in validating calculation models that include the halo component (low-dose tails) of proton pencil beam scanning (PBS) spots. Such a pattern has been used previously for quality assurance purposes to assess spot shape, position, and dose.
METHODS: In this study, a scintillation detector was used to measure the test pattern in air at isocenter for two proton beam energies (115 and 225 MeV) of two IBA universal nozzles (UN #1 and UN #2). Planar measurements were compared with calculated dose distributions based on the weighted superposition of location-independent (UN #1) or location-dependent (UN #2) spot profiles, previously measured using a pair-magnification method and between two nozzles.
RESULTS: Including the halo component below 1% of the central dose is shown to improve the gamma-map comparison between calculation and measurement from 94.9% to 98.4% using 2 mm/2% criteria for the 115 MeV proton beam of UN #1. In contrast, including the halo component below 1% of the central dose does not improve the gamma agreement for the 115 MeV proton beam of UN #2, due to the cutoff of the halo component at off-axis locations. When location-dependent spot profiles are used for calculation instead of spot profiles at central axis, the gamma agreement is improved from 98.0% to 99.5% using 2 mm/2% criteria. The two nozzles clearly have different characteristics, as a direct comparison of measured data shows a passing rate of 89.7% for the 115 MeV proton beam. At 225 MeV, the corresponding gamma comparisons agree better between measurement and calculation, and between measurements in the two nozzles.
CONCLUSIONS: In addition to confirming the primary component of individual PBS spot profiles, a comprehensive test pattern is useful for the validation of the halo component at off-axis locations, especially for low energy protons.

Entities:  

Mesh:

Year:  2015        PMID: 26328965     DOI: 10.1118/1.4928157

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Toward improved target conformity for two spot scanning proton therapy delivery systems using dynamic collimation.

Authors:  Alexandra Moignier; Edgar Gelover; Blake R Smith; Dongxu Wang; Ryan T Flynn; Maura L Kirk; Liyong Lin; Timothy D Solberg; Alexander Lin; Daniel E Hyer
Journal:  Med Phys       Date:  2016-03       Impact factor: 4.071

2.  Development of a storage phosphor imaging system for proton pencil beam spot profile determination.

Authors:  Jufri Setianegara; Thomas R Mazur; Yao Hao; Deshan Yang; H Harold Li
Journal:  Med Phys       Date:  2021-08-10       Impact factor: 4.506

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.