Evaluation of two HIV antibody confirmatory assays: Geenius™ HIV1/2 Confirmatory Assay and the recomLine HIV-1 & HIV-2 IgG Line Immunoassay.

I Friedrichs1, C Buus2, A Berger2, O T Keppler2, H F Rabenau2.   

Abstract

The laboratory diagnosis of an HIV infection mainly depends on the detection of HIV-specific antibodies/HIV p24 antigen whereby different algorithms for the confirmation of reactive screening assays exist. The objective of the present study was to compare the performance of two supplemental HIV antibody confirmatory assays: the Geenius™ HIV1/2 Confirmatory Assay and the recomLine HIV-1 & HIV-2 IgG Line Immunoassay. Therefore 279 serum samples previously analyzed for HIV during routine diagnostics at the Institute for Medical Virology, National Reference Center for Retroviruses, University Hospital Frankfurt, were analyzed retrospectively. 96.8% samples had concordant results in both HIV confirmatory assays, whereby the Geenius Assay showed a discrimination rate of 100% while two HIV-1 samples were not typeable with the recomLine Assay. Overall assay sensitivity was 100% in both assays and specificity was 99.0% (recomLine Assay) and 93.4% (Geenius Assay), respectively. The κ-values for both assays indicated high agreement. Overall nine samples had discordant results from which four were from acutely EBV/CMV-infected patients and one from a patient with primary HIV-1 infection during seroconversion. In conclusion, both assays are well suited for the detection, confirmation and discrimination of HIV-1- and -2-specific antibodies.
Copyright © 2015 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Confirmatory assay; Immunochromatographic assay; Line immunoassay; Serological HIV diagnosis

Mesh:

Substances:

Year:  2015        PMID: 26315319     DOI: 10.1016/j.jviromet.2015.08.015

Source DB:  PubMed          Journal:  J Virol Methods        ISSN: 0166-0934            Impact factor:   2.014


  2 in total

1.  Evaluation of the Bio-Rad Geenius HIV 1/2 Assay as Part of a Confirmatory HIV Testing Strategy for Quebec, Canada: Comparison with Western Blot and Inno-Lia Assays.

Authors:  Bouchra Serhir; Céline Desjardins; Florence Doualla-Bell; Marc Simard; Cécile Tremblay; Jean Longtin
Journal:  J Clin Microbiol       Date:  2019-05-24       Impact factor: 5.948

2.  Comparative evaluation of the Geenius HIV 1/2 Confirmatory Assay and the HIV-1 and HIV-2 Western blots in the Japanese population.

Authors:  Makiko Kondo; Koji Sudo; Takako Sano; Takuya Kawahata; Ichiro Itoda; Shinya Iwamuro; Yukihiro Yoshimura; Natsuo Tachikawa; Yoko Kojima; Haruyo Mori; Hiroshi Fujiwara; Naoki Hasegawa; Shingo Kato
Journal:  PLoS One       Date:  2018-10-31       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.