Literature DB >> 26303190

Characterization of a new transmission detector for patient individualized online plan verification and its influence on 6MV X-ray beam characteristics.

Johannes Thoelking1, Yuvaraj Sekar2, Jens Fleckenstein2, Frank Lohr2, Frederik Wenz2, Hansjoerg Wertz2.   

Abstract

PURPOSE: Online verification and 3D dose reconstruction on daily patient anatomy have the potential to improve treatment delivery, accuracy and safety. One possible implementation is to recalculate dose based on online fluence measurements with a transmission detector (TD) attached to the linac. This study provides a detailed analysis of the influence of a new TD on treatment beam characteristics.
METHODS: The influence of the new TD on surface dose was evaluated by measurements with an Advanced Markus Chamber (Adv-MC) in the build-up region. Based on Monte Carlo simulations, correction factors were determined to scale down the over-response of the Adv-MC close to the surface. To analyze the effects beyond dmax percentage depth dose (PDD), lateral profiles and transmission measurements were performed. All measurements were carried out for various field sizes and different SSDs. Additionally, 5 IMRT-plans (head & neck, prostate, thorax) and 2 manually created test cases (3×3cm(2) fields with different dose levels, sweeping gap) were measured to investigate the influence of the TD on clinical treatment plans. To investigate the performance of the TD, dose linearity as well as dose rate dependency measurements were performed.
RESULTS: With the TD inside the beam an increase in surface dose was observed depending on SSD and field size (maximum of +11%, SSD = 80cm, field size = 30×30cm(2)). Beyond dmax the influence of the TD on PDDs was below 1%. The measurements showed that the transmission factor depends slightly on the field size (0.893-0.921 for 5×5cm(2) to 30×30cm(2)). However, the evaluation of clinical IMRT-plans measured with and without the TD showed good agreement after using a single transmission factor (γ(2%/2mm) > 97%, δ±3% >95%). Furthermore, the response of TD was found to be linear and dose rate independent (maximum difference <0.5% compared to reference measurements).
CONCLUSIONS: When placed in the path of the beam, the TD introduced a slight, clinically acceptable increase of the skin dose even for larger field sizes and smaller SSDs and the influence of the detector on the dose beyond dmax as well as on clinical IMRT-plans was negligible. Since there was no dose rate dependency and the response was linear, the device is therefore suitable for clinical use. Only its absorption has to be compensated during treatment planning, either by the use of a single transmission factor or by including the TD in the incident beam model.
Copyright © 2015. Published by Elsevier GmbH.

Entities:  

Keywords:  Online-Dosisverifikation; Quality assurance; Qualitätssicherung; Transmissionsdetektor; online dose verification; transmission detector

Mesh:

Year:  2015        PMID: 26303190     DOI: 10.1016/j.zemedi.2015.08.001

Source DB:  PubMed          Journal:  Z Med Phys        ISSN: 0939-3889            Impact factor:   4.820


  8 in total

1.  Commissioning and validation of fluence-based 3D VMAT dose reconstruction system using new transmission detector.

Authors:  Yuji Nakaguchi; Takeshi Oono; Masato Maruyama; Yoshinobu Shimohigashi; Yudai Kai; Yuya Nakamura
Journal:  Radiol Phys Technol       Date:  2018-03-12

2.  Validation of Dolphin dosimetry in three dimensional patient-specific quality assurance programme.

Authors:  Niyas Puzhakkal; Abdullah Kallikuzhiyil Kochunny; Dinesh Makuny; Arun Krishnan M P; Ranjith C Poyil; Vysakh Raveendran
Journal:  Rep Pract Oncol Radiother       Date:  2019-08-12

3.  Two-dimensional solid-state array detectors: A technique for in vivo dose verification in a variable effective area.

Authors:  Kananan Utitsarn; Giordano Biasi; Nauljun Stansook; Ziyad A Alrowaili; Marco Petasecca; Martin Carolan; Vladimir L Perevertaylo; Wolfgang A Tomé; Tomas Kron; Michael L F Lerch; Anatoly B Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2019-10-14       Impact factor: 2.102

4.  DVH analysis using a transmission detector and model-based dose verification system as a comprehensive pretreatment QA tool for VMAT plans: Clinical experience and results.

Authors:  Ahamed B Mohamed Yoosuf; Salem AlShehri; Abdulrahman Alhadab; Mamdooh Alqathami
Journal:  J Appl Clin Med Phys       Date:  2019-10-11       Impact factor: 2.102

Review 5.  Novel methodologies for dosimetry audits: Adapting to advanced radiotherapy techniques.

Authors:  Marlies Pasler; Victor Hernandez; Núria Jornet; Catharine H Clark
Journal:  Phys Imaging Radiat Oncol       Date:  2018-03-19

6.  Effect of an integral quality monitor on 4-, 6-, 10-MV, and 6-MV flattening filter-free photon beams.

Authors:  Trang Hong Thi Nguyen; Haruna Yokoyama; Hironori Kojima; Naoki Isomura; Akihiro Takemura; Shinichi Ueda; Kimiya Noto
Journal:  J Appl Clin Med Phys       Date:  2020-12-03       Impact factor: 2.102

7.  Development of Optical Fiber Based Measurement System for the Verification of Entrance Dose Map in Pencil Beam Scanning Proton Beam.

Authors:  Jaeman Son; Se Byeong Lee; Youngkyung Lim; Sung Yong Park; Kwanho Cho; Myonggeun Yoon; Dongho Shin
Journal:  Sensors (Basel)       Date:  2018-01-15       Impact factor: 3.576

8.  Validation of a method for in vivo 3D dose reconstruction in SBRT using a new transmission detector.

Authors:  Yuji Nakaguchi; Takeshi Ono; Masato Maruyama; Yoshinobu Shimohigashi; Yudai Kai
Journal:  J Appl Clin Med Phys       Date:  2017-06-02       Impact factor: 2.102

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.