| Literature DB >> 26295397 |
Na Yao1,2, Zhuan Liu3,4, Ying Chen5, Yikai Zhou6, Bin Xie7.
Abstract
A novel rapid methodology for determining the chemical oxygen demand (COD) based on a thermal sensor with a flow injection analysis system was proposed and experimentally validated. The ability of this sensor to detect and monitor COD was based on the degree of enthalpy increase when sodium hypochlorite reacted with the organic content in water samples. The measurement results were correlated with COD and were compared against the conventional method using potassium dichromate. The assay required only 5-7 min rather than the 2 h required for evaluation by potassium dichromate. The linear range was 5-1000 mg/L COD, and the limit of detection was very low, 0.74 mg/L COD. Moreover, this method exhibited high tolerance to chloride ions; 0.015 mol/L chloride ions had no influence on the response. Finally, the sensor was used to detect the COD of different water samples; the results were verified by the standard dichromate method.Entities:
Keywords: chemical oxygen demand; flow injection analysis system; sodium hypochlorite; thermal sensor
Mesh:
Substances:
Year: 2015 PMID: 26295397 PMCID: PMC4570432 DOI: 10.3390/s150820501
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1Schematic of the thermal sensor instrument for COD analysis.
Figure 2(a) Effect of pH of NaClO solution on the thermal sensor for a 350 μL injection of 156.2 mg/L glycine solution, at a 1.0 mL/min flow rate and signal magnification of 100; (b) pH effect of the glycine solution in the thermal sensor when injecting 350 μL of 156.2 mg/L glycine solution, at a 1.0 mL/min flow rate and signal magnification of 100.
Figure 3Peak voltage from the reaction of glycine with different NaClO concentrations (Other conditions are as in Figure 2).
Influence of flow rate on the thermal sensor. (Other conditions were as in Figure 2)
| Flow Rate (mL/min) | ΔV (mV) | RSD (%) | ||
|---|---|---|---|---|
| 0.25 | 0 | 0 | 0 | 0 |
| 0.5 | 27.4 | 25.8 | 24.6 | 5.417 |
| 1.0 | 35.1 | 36.1 | 35.3 | 1.491 |
| 1.5 | 25.6 | 29.4 | 19.2 | 20.842 |
| 2.0 | 22.0 | 25.1 | 30.3 | 16.256 |
Figure 4Calibration curve for standard glycine samples. Data represent the average of three measurements.
Figure 5An actual sensor signal record for measurements of 50, 80, 100, and 200 mg/L COD solutions.
Figure 6Stability of the response to 100 mg/L COD using FIA- thermal sensor observed for 28 continuous days. Samples were injected six times every day during this period.
Figure 7Correlation of COD values of real samples from rivers and lakes determined by the thermal sensor and conventional method. The CODET values of real samples were inserted into the calibration equation using the standard sample method, CODET = 6.7654 + 5.3577 CODCr, R = 0.9707. Data represent the average of three measurements, n = 16.