The AdvanSure™ RV real-time PCR kit (AdvanSure; LG Life Sciences, Korea) is based on multiplex real-time PCR and can simultaneously detect 14 respiratory viruses. We compared the performance of the AdvanSure assay with the Seeplex RV 12 ACE detection kit (Seeplex; Seegene, Seoul, South Korea), a multiplex end-point PCR assay. A total of 454 consecutive respiratory specimens were tested with both AdvanSure and Seeplex assays; AdvanSure detected 153 (33.7%) positive cases and Seeplex detected 145 (31.9%) positive cases. The positive percent agreement, negative percent agreement, and kappa value for the two assays were 87.2% (95% CI, 80.3-92.1), 91.1% (95% CI, 87.2-93.9), and 0.77 (95% CI, 0.70-0.83), respectively. Compared with the Seeplex assay, the AdvanSure assay had a shorter turnaround time (3h vs. 8h) and a shorter hands-on time (<1h vs 2h). In conclusion, the AdvanSure assay demonstrated comparable performance to the Seeplex assay.Entities:
Keywords: Korea; Multiplex real-time PCR; Respiratory virus; Validation
Mesh:
Year: 2015 PMID: 26277911 PMCID: PMC7113774 DOI: 10.1016/j.jviromet.2015.08.003
Source DB: PubMed Journal: J Virol Methods ISSN: 0166-0934 Impact factor: 2.014
Detection of respiratory viruses by AdvanSure™ RV real-time RT-PCR and Seeplex® RV12 ACE assays.
| Virus | Agreement | Kappa value | ||||
|---|---|---|---|---|---|---|
| Positive | Negative | Observed kappa | 95% CI | |||
| % | 95% CI | % | 95% CI | |||
| 80.3–92.1 | 87.2–93.9 | 0.70–0.83 | ||||
| ADV | 100 | 71.7–100 | 91.1 | 97.2–99.6 | 0.83 | 0.69–0.98 |
| INF A | 100 | 59.8–100 | 98.9 | 98.6–99.9 | 0.94 | 0.82–1.00 |
| INF B | 100 | 56.1–100 | 99.8 | 98.6–99.9 | 0.93 | 0.80–1.00 |
| PIV 1 | 100 | 19.8–100 | 99.8 | 99.0–100 | 1.00 | 1.00 |
| PIV 2 | 50.0 | 2.7–97.3 | 100 | 98.2–99.9 | 0.40 | 0.00–0.94 |
| PIV 3 | 100 | 5.4–100 | 99.8 | 98.6–99.9 | 0.67 | 0.05–1.00 |
| HRV | 60.3 | 47.2–72.2 | 95.9 | 93.3–97.6 | 0.60 | 0.49–0.71 |
| RSV A | 100 | 19.8–100 | 99.6 | 98.2–99.9 | 0.66 | 0.23–1.00 |
| RSV B | 100 | 91.9–100 | 98.5 | 96.6–99.4 | 0.94 | 0.89–0.99 |
| HMPV | 100 | 19.8–100 | 100 | 99.0–100 | 1.00 | 1.00 |
| CoV 229E/NL63 | 60.0 | 17.0–92.7 | 99.8 | 98.6–99.9 | 0.66 | 0.31–1.00 |
ADV, adenovirus; INF A, influenza A virus; INF B, influenza B virus; PIV, parainfluenza virus; HRV, human rhinovirus; RSV A, respiratory syncytial virus; HMPV, human metapneumovirus; CoV, coronavirus.
Analysis of samples with discrepant results in AdvanSure™ RV real-time RT-PCR and Seeplex® RV12 ACE assays.
| Virus | Discrepant results | No. of positive results in monoplex PCR and sequencing | |
|---|---|---|---|
| AdvanSure™ RV real-time RT-PCR/Seeplex® RV12 ACE detection result | No. of samples | ||
| ADV | +/− | 5 | 5 |
| INF A | +/− | 1 | 1 |
| INF B | +/− | 1 | 1 |
| PIV 2 | +/− | 2 | 2 |
| −/+ | 1 | 0 | |
| PIV 3 | +/− | 1 | 1 |
| HRV | +/− | 16 | 10 |
| −/+ | 25 | 17 | |
| RSV A | +/− | 2 | 1 |
| RSV B | +/− | 6 | 3 |
| CoV 229E/NL63 | +/− | 1 | 0 |
| −/+ | 2 | 0 | |
ADV, adenovirus; INF, influenza virus; PIV, parainfluenza virus; HRV, human rhinovirus; RSV, respiratory syncytial virus; HMPV, human metapneumovirus; CoV, coronavirus.
Percent sensitivity, specificity, PPV and NPV of two assays for detection of respiratory viruses.
| Virus | AdvanSure™ RV real-time RT-PCR | Seeplex® RV12 ACE | ||||||
|---|---|---|---|---|---|---|---|---|
| Sensitivity (95% CI) | Specificity (95% CI) | PPV (95% CI) | NPV (95% CI) | Sensitivity (95% CI) | Specificity (95% CI) | PPV (95% CI) | NPV (95% CI) | |
| ADV | 18/18 | 436/436 | 18/18 | 436/436 | 13/18 | 436/436 | 13/13 | 436/441 |
| INF A | 9/9 | 445/445 | 9/9 | 445/445 | 8/9 | 445/445 | 8/8 | 445/446 |
| INF B | 8/8 | 446/446 | 8/8 | 446/446 | 7/8 | 446/446 | 7/7 | 446/447 |
| PIV 1 | 2/2 | 452/452 | 2/2 | 452/452 | 2/2 | 451/451 | 2/2 | 451/451 |
| PIV 2 | 3/3 | 451/451 | 3/3 | 451/451 | 1/3 | 450/451 | 1/2 | 450/452 |
| PIV 3 | 2/2 | 452/452 | 2/2 | 452/452 | 1/2 | 452/452 | 1/1 | 452/453 |
| HRV | 48/65 | 383/389 | 48/54 | 383/400 | 55/65 | 381/389 | 55/63 | 381/391 |
| RSV A | 3/3 | 450/451 | 3/4 | 450/450 | 2/3 | 451/451 | 2/2 | 451/452 |
| RSV B | 58/58 | 393/396 | 58/61 | 393/393 | 55/58 | 395/395 | 55/55 | 395/398 |
| HMPV | 2/2 | 452/452 | 2/2 | 452/452 | 2/2 | 452/452 | 2/2 | 452/452 |
| CoV | 3/3 | 450/451 | 3/4 | 450/450 | 3/3 | 449/451 | 3/5 | 449/449 |
PPV, predictive positive value; NPV, negative predictive value; ADV, adenovirus; INF, influenza virus; PIV, parainfluenza virus; HRV, human rhinovirus; RSV, respiratory syncytial virus; HMPV, human metapneumovirus; CoV, coronavirus; NE, not estimated (number of positive results was less than five).
Fig. 1Flowchart showing the turnaround time and hands-on time of Seeplex® RV12 ACE and AdvanSure™ RV assays.