D Elleri1,2, M Biagioni1,2, J M Allen1,2, K Kumareswaran2, L Leelarathna2, K Caldwell2, M Nodale2, M E Wilinska1,2, A Haidar2, P Calhoun3, C Kollman3, N C Jackson4, A M Umpleby4, C L Acerini1,2, D B Dunger1,2, R Hovorka1,2.
Abstract
AIMS: To evaluate safety, efficacy and glucose turnover during closed-loop with meal announcement using reduced prandial insulin boluses in adolescents with type 1 diabetes (T1D).Entities:
Keywords: closed-loop insulin delivery; postprandial hypoglycaemia; type 1 diabetes
Mesh:
Substances:
Year: 2015 PMID: 26257323 PMCID: PMC4832358 DOI: 10.1111/dom.12549
Source DB: PubMed Journal: Diabetes Obes Metab ISSN: 1462-8902 Impact factor: 6.577
Study outcomes based on plasma glucose between 24:00 hours on day 1 until 08:00 hours on day 3 (32 h) is considered for analysis during closed‐loop therapy with standard prandial insulin boluses and closed‐loop therapy with reduced prandial insulin boluses.
| Closed‐loop with standard prandial boluses (n = 8) | Closed‐loop with reduced prandial boluses (n = 8) |
| |
|---|---|---|---|
|
| |||
| Median (IQR) percentage of time spent in target of 3.9–10 mmol/l | 80 (65, 96) | 74 (66, 84) | 0.87 |
|
| |||
| Mean (s.d.) glucose, mmol/l | 8.4 (0.9) | 8.4 (0.9) | 0.98 |
| Median (IQR) standard deviation of glucose, mmol/l | 1.9 (1.4, 2.3) | 2.1 (1.8, 2.5) | 0.40 |
| Median (IQR) percentage of time spent in target of 3.9–8 mmol/l | 48 (30, 54) | 42 (36, 52) | 0.84 |
|
| |||
| ≤3.9 mmol/l (%) | 0.0 (0.0, 1.8) | 0.0 (0.0, 1.5) | 0.88 |
| ≤3.5 mmol/l (%) | 0.0 (0.0, 0.9) | 0.0 (0.0, 0.9) | 0.88 |
| Median (IQR) low blood glucose index, unitless | 0.2 (0.1, 0.5) | 0.2 (0.1, 0.4) | 0.93 |
|
| |||
| >10 mmol/l | 18.0 (4.1, 34.2) | 21.8 (16.3, 33.5) | 0.87 |
| >16.6 mmol/l | 0.0 (0.0, 0.0) | 0.0 (0.0, 0.0) | — |
| Median (IQR) insulin infusion, IU/h | 1.5 (1.3, 1.6) | 1.4 (1.2, 1.7) | 0.20 |
| Median (IQR) insulin concentration, pmol/l | 252 (198, 336) | 186 (171, 260) | 0.002 |
| Median (IQR) insulin boluses, IU | 25.7 (24.1, 29.4) | 18.6 (17.6, 21.4) | <0.001 |
| Median (IQR) total insulin, IU | 72.5 (63.6, 80.3) | 61.9 (55.2, 75.0) | 0.01 |
IQR, interquartile range; s.d., standard deviation.
No subject had values >16.6 mmol/l on either visit.
One subject was excluded as a result of having a much greater insulin concentration on both visits.
Figure 1Plasma glucose concentration (A), basal insulin infusion rates (B) and the plasma insulin levels (C) during closed‐loop therapy with standard prandial insulin boluses (grey line) and closed‐loop therapy with reduced prandial boluses (black line). Values are median, interquartile ranges.
Figure 3Mean plasma insulin levels on the first and second night during closed‐loop therapy with standard prandial insulin boluses (light grey boxes) and closed‐loop with reduced prandial boluses (dark grey boxes).
Figure 2(A) Total glucose appearance and (B) disposal (mean, standard error) from 08:00 hours on day 2 for 24 h during closed‐loop therapy with standard prandial boluses (grey line) and closed‐loop with reduced prandial boluses (black line).
Glucose turnover from 08:00 hours on day 2 until midnight on day 2.
| Closed‐loop with standard prandial boluses (n = 8) | Closed‐loop with reduced prandial boluses (n = 8) | p | |
|---|---|---|---|
| Insulin concentration, pmol/l | 317 (252, 435) | 224 (204, 346) | 0.009 |
| Ra_total, µmol/kg/min | 25.4 (21.0, 29.2) | 26.3 (21.9, 28.0) | 0.19 |
| Rd, µmol/kg/min | 25.2 (21.2, 28.8) | 25.8 (21.0, 26.9) | 0.46 |
| Insulin metabolic clearance of insulin, ml/kg | 15.7 (8.3, 17.5) | 16.6 (9.2, 20.9) | 0.16 |
| Insulin sensitivity, nmol/kg/min/pmol/l | 58.1 (36.3, 74.0) | 81.2 (52.2, 94.0) | 0.22 |
Values are median (interquartile range). Ra_total, total glucose appearance; Rd, glucose disposal.
One subject was excluded as a result of having a much greater insulin concentration on both visits.