Biomechanical Testing of Spinal Segment Fixed by Arcofix System on the Swine Spine.

Martin Kelbl1, Jan Kocis1, Radek Vesely1, Zdenek Florian2, Tomas Navrat2, Petr Vosynek2.   

Abstract

STUDY
DESIGN: An in vitro biomechanical study.
PURPOSE: To evaluate the mechanical properties of the spinal segment in the intact, injured, and stabilized state after fixation by an Arcofix implant. OVERVIEW OF LITERATURE: Several types of thoracolumbar spine injury necessitates anterior instrumentation. The Arcofix plate represents the latest generation of angular stablity systems. The biomechanical properties of these implants have not been sufficiently studied yet.
METHODS: A total of ten porcine specimens (levels Th12-L3) were prepared. The tests were performed for intact, injured, and implanted specimens. In each state, the specimen was subjected to a tension load of a prescribed force, and subsequently, twisted by a given angle. The force load was 200 N. The torsion load had a deformation character, i.e., the control variable was the twisting angle and the measured variable was the moment of a couple. The amplitude of the load alternating cycle was 3°. Another parameter that was evaluated was the area of the hysteresis loop. The area corresponds to the deformation energy which is dissipated during the cycle.
RESULTS: A statistically significant difference was found between the intact and injured states as well as between the injured and implanted specimens. The statistical evaluation also showed a statistically different value of the hysteresis loop area. In the case of instability, the area decreased to 33% of the physiological value. For the implanted sample, the area increased to 170% of the physiological value.
CONCLUSIONS: The Arcofix implant with its parameters appears to be suitable and sufficiently stable for the treatment of the anterior column of the spine.

Entities:  

Keywords:  Anterior stabilization; Biomechanics; Spine stabilization; Thoracolumbar fracture

Year:  2015        PMID: 26240706      PMCID: PMC4522437          DOI: 10.4184/asj.2015.9.4.503

Source DB:  PubMed          Journal:  Asian Spine J        ISSN: 1976-1902


  12 in total

1.  In vitro biomechanical analysis of three anterior thoracolumbar implants.

Authors:  P W Hitchon; V K Goel; T N Rogge; J C Torner; A P Dooris; J S Drake; S J Yang; K Totoribe
Journal:  J Neurosurg       Date:  2000-10       Impact factor: 5.115

2.  Biomechanical analysis of anterior instrumentation for lumbar corpectomy.

Authors:  Frances D Faro; Klane K White; Jae S Ahn; Richard S Oka; Andrew T Mahar; Maneesh Bawa; Christine L Farnsworth; Steven R Garfin; Peter O Newton
Journal:  Spine (Phila Pa 1976)       Date:  2003-11-15       Impact factor: 3.468

3.  Clinical spinal instability and low back pain.

Authors:  Manohar M Panjabi
Journal:  J Electromyogr Kinesiol       Date:  2003-08       Impact factor: 2.368

4.  Comparative biomechanical testing of anterior and posterior stabilization procedures.

Authors:  Christian H Flamme; Christof Hurschler; Caroline Heymann; Nadine von der Heide
Journal:  Spine (Phila Pa 1976)       Date:  2005-07-01       Impact factor: 3.468

5.  Angular stable anterior plating following thoracolumbar corpectomy reveals superior segmental stability compared to conventional polyaxial plate fixation.

Authors:  Alexander C Disch; Christian Knop; Klaus D Schaser; Michael Blauth; Werner Schmoelz
Journal:  Spine (Phila Pa 1976)       Date:  2008-06-01       Impact factor: 3.468

6.  In vitro biomechanical characteristics of the spine: a comparison between human and porcine spinal segments.

Authors:  Iris Busscher; Albert J van der Veen; Jaap H van Dieën; Idsart Kingma; Gijsbertus J Verkerke; Albert G Veldhuizen
Journal:  Spine (Phila Pa 1976)       Date:  2010-01-15       Impact factor: 3.468

7.  Testing criteria for spinal implants: recommendations for the standardization of in vitro stability testing of spinal implants.

Authors:  H J Wilke; K Wenger; L Claes
Journal:  Eur Spine J       Date:  1998       Impact factor: 3.134

8.  Biomechanical testing of spinal segment fixed by thoracolumbar spine locking plate on the swine lumbar spine.

Authors:  Jan Kocis; Tomas Navrat; Zdenek Florian; Peter Wendsche
Journal:  Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub       Date:  2010-12       Impact factor: 1.245

9.  Effects of dorsal versus ventral shear loads on the rotational stability of the thoracic spine: a biomechanical porcine and human cadaveric study.

Authors:  Jan-Willem M Kouwenhoven; Theo H Smit; Albert J van der Veen; Idsart Kingma; Jaap H van Dieën; René M Castelein
Journal:  Spine (Phila Pa 1976)       Date:  2007-11-01       Impact factor: 3.468

10.  Basic Science Symposium III: Animal Models for Orthopaedic Implant Evaluation.

Authors:  Matthew J Allen; Anthony Simon Turner; Koichi Sairyo; Lisa Ferrara
Journal:  SAS J       Date:  2008-12-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.