| Literature DB >> 26219005 |
Timothy P Szczykutowicz1, Robert K Bour, Nicholas Rubert, Gary Wendt, Myron Pozniak, Frank N Ranallo.
Abstract
This article explains a method for creating CT protocols for a wide range of patient body sizes and clinical indications, using detailed tube current information from a small set of commonly used protocols. Analytical expressions were created relating CT technical acquisition parameters which can be used to create new CT protocols on a given scanner or customize protocols from one scanner to another. Plots of mA as a function of patient size for specific anatomical regions were generated and used to identify the tube output needs for patients as a function of size for a single master protocol. Tube output data were obtained from the DICOM header of clinical images from our PACS and patient size was measured from CT localizer radiographs under IRB approval. This master protocol was then used to create 11 additional master protocols. The 12 master protocols were further combined to create 39 single and multiphase clinical protocols. Radiologist acceptance rate of exams scanned using the clinical protocols was monitored for 12,857 patients to analyze the effectiveness of the presented protocol management methods using a two-tailed Fisher's exact test. A single routine adult abdominal protocol was used as the master protocol to create 11 additional master abdominal protocols of varying dose and beam energy. Situations in which the maximum tube current would have been exceeded are presented, and the trade-offs between increasing the effective tube output via 1) decreasing pitch, 2) increasing the scan time, or 3) increasing the kV are discussed. Out of 12 master protocols customized across three different scanners, only one had a statistically significant acceptance rate that differed from the scanner it was customized from. The difference, however, was only 1% and was judged to be negligible. All other master protocols differed in acceptance rate insignificantly between scanners. The methodology described in this paper allows a small set of master protocols to be adapted among different clinical indications on a single scanner and among different CT scanners.Entities:
Mesh:
Year: 2015 PMID: 26219005 PMCID: PMC5690004 DOI: 10.1120/jacmp.v16i4.5412
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Depiction of an example of how noticing a protocol is not reaching the maximum tube current can lead to a change of parameters (higher pitch and/or faster rotation time) and, thus, a reduction in scan time can be made without compromising image quality.
Figure 2Example box and whisker plot used to assess a protocol by analyzing what mA values the AEC system was using. The median is shown with a horizontal line, the extent of the 90th and 10th percentiles are shown with a thick line, the extent of all points outside the 90th and 10th percentiles but not considered outliers are shown with a thin line, and outliers are individually shown with a cross. The data were taken from a routine abdominal/pelvis scan. Each box and whisker represents the mA distribution from a single patient exam. The horizontal axis value was computed by taking the sum of the average of each patient's lateral and anterior/posterior width measured from CT localizer radiograph images.
Example of customizing a master protocol within a single scanner. The routine abdomen–pelvis, trauma, and precontrast protocols refer to master protocols B4, B1, and B7 listed in Table 4, respectively
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|
| routine abdomen–pelvis medium adult | 120 |
| 0.4 | 0.516 | 15.5 | ‐ | ‐ |
| trauma abdomen–pelvis medium adult | 120 |
| 0.6 | 0.516 | 10 | 1 | 1 |
| precontrast abdomen–pelvis medium adult | 120 |
| 0.4 | 0.516 | 22 | 1 | 1 |
a This is the range of mA values allowed to be used by the scanner for AEC which, on a GE scanner, the user can adjust.
Figure 3Example plot used to obtain reference minimum and maximum mA values.
Master protocols in use for the abdominal section at our institution. Following the protocol name is a descriptor describing the type of imaging phase. T denotes a trauma phase, A denotes an arterial phase, P denotes a parenchymal or portal venous phase, R denotes a routine abdominal phase, V denotes a venous phase (different scan delay compared to P), LA denotes a late arterial phase, and D denotes a delayed phase
|
|
|
|
|
|---|---|---|---|
| B1 | 120 | 10 | trauma abdomen–pelvis T, trauma cystogram T |
| B2 | 120 | 11.5 | transjugular intrahepatic portosystemic shunting evaluation A, pancreas neoplasm screening A |
| B3 | 120 | 13 | urography P, adrenal gland (adenoma) R, pancreas neoplasm screening V, kidney tumor D |
| B4 | 120 | 15.5 | routine abdomen/pelvis R, Pelvis R, Rule out hernia R, flank pain R, biphasic liver LA, adrenal gland (adenoma) A/D, kidney tumor A, trauma abdomen–pelvis D, cystogram R, peritoneogram A, preinferior vena cava filter removal A, transjugular intrahepatic portosystemic shunting evaluation P |
| B5 | 120 | 18 | urography R, gastric varices R/P, small bowel enterography V, portosystemic shunt evaluation A |
| B6 | 120 | 20.5 | pancreas neoplasm screening LA, pancreas transplant LA, kidney tumor LA, renal donor LA, obscure gastrointestinal bleed LA, body perfusion LA |
| B7 | 120 | 22 | Low dose renal stone and flank pain R, triphasic liver LA |
| B8 | 100 | 14 | pancreas neoplasm pre‐operation A, obscure gastrointestinal bleed A, triphasic liver A, liver donor work‐up A/D |
| B9 | 100 | 15.5 | pancreas neoplasm pre‐operation P |
| B10 | 100 | 18.5 | biphasic liver A, liver cholangiocarcinoma D, liver donor work‐up V, renal donor P |
| B11 | 100 | 22 | triphasic liver LA/V, liver donor work‐up LA, obscure gastrointestinal bleed D, mesenteric ischemia V |
| B12 | 100 | 23.5 | flank pain D |
a kV for a medium sized patient; small patients are imaged at 20 kV lower and large patients at 20 kV higher.
b NI at 5 mm image thickness for a medium sized patient.
c Master protocol from which detailed mA distribution data were collected and used to create the other master protocols using Eq. (1).
d Delayed biliary contrast‐only scan.
Figure 4Flowchart detailing the proper use of mA data and (1), (2) in creating protocols on the same scanner at different CNR levels and customizing these protocols to a different scanner.
Calculation of using iodine contrast values taken from Yu et al. and experimentally measured changes in image noise as a function of kV at equal mAs. The contrast change was measured in a 25 cm water phantom and the image noise as a function of kV measurements were performed using a 22 cm water phantom
|
|
|
|
|
|---|---|---|---|
|
|
| 1.32 | 2.17 |
|
|
| 1.27 | 1.63 |
|
|
| 1.21 | 1.39 |
aRatio of HU values at the lower kV to the higher kV.
Square of the ratio of image noise values at the lower kV to the higher kV measured with the same mAs.
Example of customizing a master protocol from a scanner with an iterative denoising option to a scanner without iterative denoising. In addition, the new scanner has maximum mA limits of 500 at 120 kV and 400 at 100 kV. Scan times () were computed assuming , and
|
|
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|---|---|
| Original protocol | 100 | 700 | 0.4 | 0.984 | 20 | ‐ | ‐ | ‐ | 4.6 |
| Higher kV | 120 | 700 | 0.4 | 0.984 | 16 | 1 | 0.8 | 1.25 | 1.6 |
| Longer t | 100 | 389 | 0.9 | 0.984 | 20 | 1 | 1 | 1.25 | 10.3 |
| Longer P | 100 | 458 | 0.4 | 0.516 | 20 | 1 | 1 | 1.25 | 8.7 |
| Combination | 120 | 467 | 0.6 | 0.984 | 16 | 1 | 0.8 | 1.25 | 6.9 |
a In order to keep the CNR level constant, the NI must change to compensate for the change in image contrast due to the kV change.
b Assuming the use of an iterative denoiser provides a dose reduction of 20%.
Note, the value of used here does not match the value in Table 2. If the value in Table 2 were used, would be 0.99. However, iodine CNR is not a sufficient index of image quality for nonangiographic exams. In this example, we decreased in order to accommodate noniodinated structures. All NI values assume a 5 mm slice thickness.
Figure 5Box and whisker plot of the mA as a function of patient size for the (top) low dose B7 master protocol and (bottom) higher dose B1 master protocols from a LightSpeed VCT scanner for the medium patient size range. The median is shown with a horizontal line, the extent of the 90th and 10th percentiles are shown with a thick line, the extent of all points outside the 90th and 10th percentiles but not considered outliers are shown with a thin line, and outliers are individually shown with a cross. The minimum/maximum mA values for the top and bottom graphs were 15/200 and 50/640, respectively.
Figure 6Box and whisker plot of the mA as a function of patient size for the routine abdominal pelvis protocol from three different CT scanners for the medium patient size range. The LightSpeed VCT, Optima 660, and Discovery HD 750 systems are shown from top to bottom, respectively. The median is shown with a horizontal line, the extent of the 90th and 10th percentiles are shown with a thick line, the extent of all points outside the 90th and 10th percentiles but not considered outliers are shown with a thin line, and outliers are individually shown with a cross. The minimum/maximum mA values for the top, bottom, and middle graphs were 30/400, 40/320 and 50/500, respectively.
Example of customizing a protocol to a different scanning platform
|
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|---|
| Routine abdomen–pelvis master protocol (B4/LightSpeed VCT) | 120 |
| 0.4 | 0.516 | 15.5 | ‐ | ‐ | ‐ |
| B4 Optima 660 | 120 |
| 0.5 | 0.516 | 15.5 | 1 | 1 | 1 |
| B4 Discovery HD 750 | 120 |
| 0.4 | 0.516 | 18 | 1 | 1 | 1 |
| B4 LightSpeed 16 | 120 |
| 0.5 | 0.516 | 14 | 1 | 1 | 1.25 |
a This is the range of mA values to be used by the scanner for the AEC option.
b The AEC function was reported by the manufacturer to have changed for this model, making a NI of 18 essentially equivalent to 15.5 for this patient size.
c Iterative denoising was not an option for this scanning platform, which required the mA value to be modulated by and additionally the NI parameter to be reduced to compensate for the increase in image noise. The NI was reduced by the factor .
Quality assurance data for the master protocols as a function of scanner model. The values listed are the percent acceptance rate. The sample sizes for the various master protocols ranged in number from a high of over 5,000 patient exams (B4) to single digits, and in some cases no master protocol were scanned for some scanner models. The two‐tailed Fisher's exact test p‐value comparing each customized master protocol's acceptance rate with the acceptance rate of the reference scanner (LightSpeed VCT) are shown in parenthesis. B2 and B9 were excluded from the analysis due to an insufficient amount of data
|
|
|
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|---|---|---|
| LightSpeed VCT | 98 | 96 | 97 | 97 | 89 | 100 | 96 | 96 | 100 | 98 |
| Optima 660 | 95 (0.158) | 94 (0.7545) | 96 | 96 (1.000) | 95 (0.414) | 95 (1.000) | 95 (1.000) | 94 (0.745) | 94 (1.000) | 97 (0.757) |
| Discovery HD 750 | 98 (0.609) | 100 (0.834) | 98 (0.658) | 100 (1.000) | 100 (0.252) | 100 (1.000) | 100 (0.218) | 96 (1.000) | 100 (1.000) | 98 (0.598) |
a A p‐value less than 0.005.