Guclu Aykut1, Gerke Veenstra, Claudia Scorcella, Can Ince, Christiaan Boerma.
Abstract
BACKGROUND: Orthogonal polarized spectral (OPS) and sidestream dark field (SDF) imaging video microscope devices were introduced for observation of the microcirculation but, due to technical limitations, have remained as research tools. Recently, a novel handheld microscope based on incident dark field illumination (IDF) has been introduced for clinical use. The Cytocam-IDF imaging device consists of a pen-like probe incorporating IDF illumination with a set of high-resolution lenses projecting images on to a computer controlled image sensor synchronized with very short pulsed illumination light. This study was performed to validate Cytocam-IDF imaging by comparison to SDF imaging in volunteers.Entities:
Year: 2015 PMID: 26215807 PMCID: PMC4512989 DOI: 10.1186/s40635-015-0040-7
Source DB: PubMed Journal: Intensive Care Med Exp ISSN: 2197-425X
Figure 1Smaller SDF image in larger Cytocam-IDF image. This figure shows the field of view of SDF and Cytocam-IDF imaging superimposed on each other showing the larger field of view offered by the larger image sensor used by the later technique.
Figure 2Histogram points taken for analyses; square capillary; circle venule; left Cytocam-IDF imaging; right SDF-imaging.
Baseline characteristics
|
|
|
|---|---|
| Age [years] | 33 [27.5 to 46.0] |
| Gender, male [ | 8 |
| Systolic blood pressure [mmHg] | 130 [119 to 141] |
| Diastolic blood pressure [mmHg] | 74 [66 to 87] |
| Mean arterial blood pressure [mmHg] | 92 [84 to 102] |
| Heartrate [beats per minute] | 70 [65 to 77] |
| Weight [kg] | 73 [63 to 78] |
| Height [cm] | 173 [170 to 180] |
Figure 3Boxplots of TVD and PVD.
Microcirculatory parameters
|
|
|
| |
|---|---|---|---|
| MFI small | 3.0 [3.0 to 3.0] | 3.0 [2.96 to 3.00] | 0.289 |
| MFI large | 3.0 [3.0 to 3.0] | 3.0 [2.96 to 3.00] | 0.494 |
| TVD [mm/mm2] | 21.60 ± 4.30 | 16.35 ± 2,78 | <0.0001* |
| PPV [%] | 100 [99 to 100] | 99 [99 to 100] | 0.368 |
| PVD [mm/mm2] | 21.50 ± 4.38 | 16.24 ± 2.81 | <0.0001* |
MFI small: <20 um; MFI large: >20 um.
*p < 0.05 is the cutoff value for statistical significance.
Results from contrast and sharpness analysis
|
|
|
|
| |
|---|---|---|---|---|
| Capillaries | SDF | 0.04 ± 0.01 | 0.78 ± 0.06 | 0.03 ± 0.01 |
| Cytocam | 0.99 ± 0.03 | 0.90 ± 0.05 | 0.09 ± 0.03 | |
|
|
|
| ||
| Venules | SDF | 0.07 ± 0.03 | 0.82 ± 0.07 | 0.06 ± 0.03 |
| Cytocam | 0.12 ± 0.03 | 0.88 ± 0.05 | 0,11 ± 0.03 | |
|
|
|
|
*p < 0.05 is the cutoff value for statistical significance.