| Literature DB >> 26210400 |
Alistair Curd1, Alexa Cleasby2, Katarzyna Makowska2, Andrew York3, Hari Shroff3, Michelle Peckham2.
Abstract
A challenge in biological imaging is to capture high-resolution images at fast frame rates in live cells. The "instant structured illumination microscope" (iSIM) is a system designed for this purpose. Similarly to standard structured illumination microscopy (SIM), an iSIM provides a twofold improvement over widefield microscopy, in x, y and z, but also allows much faster image acquisition, with real-time display of super-resolution images. The assembly of an iSIM is reasonably complex, involving the combination and alignment of many optical components, including three micro-optics arrays (two lenslet arrays and an array of pinholes, all with a pitch of 222 μm) and a double-sided scanning mirror. In addition, a number of electronic components must be correctly controlled. Construction of the system is therefore not trivial, but is highly desirable, particularly for live-cell imaging. We report, and provide instructions for, the construction of an iSIM, including minor modifications to a previous design in both hardware and software. The final instrument allows us to rapidly acquire fluorescence images at rates faster than 100 fps, with approximately twofold improvement in resolution in both x-y and z; sub-diffractive biological features have an apparent size (full width at half maximum) of 145 nm (lateral) and 320 nm (axial), using a 1.49 NA objective and 488 nm excitation.Entities:
Keywords: Construction; Fluorescence microscopy; Instant structured illumination microscope; Super-resolution
Mesh:
Year: 2015 PMID: 26210400 PMCID: PMC4641873 DOI: 10.1016/j.ymeth.2015.07.012
Source DB: PubMed Journal: Methods ISSN: 1046-2023 Impact factor: 3.608
Fig. 1The layout of the implementation of iSIM at UoL, designed to replicate the function of the original [14]. (A) The overall set-up for the iSIM. Plane mirrors 3 and 4 form a periscope to raise the beam height to match the axis of the sample stage. The beamlets generated by lenslet array 1 are focussed at the sample, and scanned across it by the scanning mirror, which rotates about the y-axis as indicated. The beamlets also come to a focus at the intermediate image plane shown. (B) The positions of bypass mirrors implemented when imaging in brightfield. The majority of the hardware was fixed to a metric anti-vibration table (Table 1). 1″ diameter pedestal posts and fork clamps were used to attach most components to the table, with component height adjustable using spacers or ½″ diameter post holders and posts (Table 1).
List of components used in the iSIM.
| Component | Manufacturer | Notes |
|---|---|---|
| Optical table: M-RS2000-48-8 | Newport Spectra-Physics Ltd. | Table was electrically earthed |
| Legs: S-2000A-423.5 | ||
| Pedestal posts, holders: ESK021A/M, ESK01/M, ESK03/M | Thorlabs | Except where described, 1″ and 2″ diameter optics were mounted in these lens mounts |
| Lens mounts: LMR1, LMR2 | ||
| Sample stage: RAMM-BASIC-DV frame, MIM-FC-FOCUS-K (for objective and folding mirror), PZ2300 piezoelectric | Applied Scientific Instrumentation | These components make up the basic frame, motorised stage insert, and housing for objective and folding mirror |
| UK distributors: | A tube lens may be supplied by default with the RAMM frame, but needs to be removed to allow use of the Edmund Optics tube lens listed below | |
| sCMOS camera: pco.edge 4.2 (air cooled) | PCO | The camera was mounted on a translation stage (Thorlabs, PT1B/M) to set its |
| UK distributors: | ||
| Excitation lasers: (561 nm and 488 nm wavelengths) Genesis MX488-1000 STM Genesis MX561-500 STM | Coherent | Both have a maximum power 1.1 W), both clamped to the optical table |
| Dichroic mirror 1: LPD01-400-RU-25 | Semrock | For combination of excitation laser paths |
| UK Distributors | ||
| Acousto-optic tunable filter (AOTF): AOTFnC-400.650-TN, with 8 channel controller MDS8C-B66-22-74.158) | AA Opto-electronic | The AOTF deflects the beams into and out of the system path as required |
| UK distributors: Photon Lines Ltd. | M3 bolts and washers were used with M3-M6 thread adaptors (Thorlabs, AE3M6M) to fasten the AOTF to pillar pedestal posts | |
| Power supply for AOTF: IPS 303DD and 25-pin D-Sub connector (e.g. Amphenol, L77DB25SST; L717DB25PST (with gender changer) | Isotech | The D-Sub connector is required to connect the AOTF to the control channels of the breakout box and to the power supply |
| Supplied by Radio Shack | NB, the power supply must supply more than the maximum specified current (0.9 A) during the warm-up period (a current limit at 1.5 A was sufficient) | |
| Half-wave plates: WPH10M-488 and AHWP05M-600 | Thorlabs | Mounted in rotation mounts (Thorlabs, RSP1C) |
| To maximise diffraction efficiency of the AOTF for the 488 nm and 561 nm lasers | ||
| Beam expander: ∗64-837, | ∗Edmund Optics | The shorter focal length lens was mounted in a v-clamp (Thorlabs, VC3) |
| ∗∗Thorlabs | ||
| Scan lens × 4: | Special Optics | Each mounted in a v-clamp (Thorlabs, VC3, with extension post MS1R/M) |
| Tube lens: 49-289-INK, | Edmund Optics | This was mounted in a tip-tilt mount (Thorlabs KM200) |
| Objective lenses: APON60XOTIRF, NA 1.49, apochromatic TIRF lens | Olympus | |
| UPLSAPO60XS, NA 1.3, plan, super-apochromatic | ||
| UPLSAPO60XW, NA 1.2, plan, super-apochromatic | ||
| Relay lens x 2: AC508-300-A-ML, | Thorlabs | |
| Dichroic mirror 2: 488-561 DM | Iridian Spectral Technologies | To separate emission and excitation light |
| UK distributors: Laser Lines | Filter is 6 mm thick to reduce curvature of the reflecting surface introduced by clamping in its mount | |
| Mount used was KM100C (Thorlabs) | ||
| Filter wheel: Lambda 10-B | Sutter Instrument Company | To house filters for further rejection of excitation light and possible bandpass separation of emission channels |
| UK distributors: Photometrics | ||
| Notch filters: NF03-488E-25 and NF03-561E-25 | Semrock | For use in the filter wheel, for further rejection of excitation light |
| UK Distributors | ||
| Scanning mirror: SPO9086, coated on both sides | Sierra Precision Optics | The scanning mirror was directly shipped to Nutfield Technologies for mounting |
| Mounting and control for Scanning Mirror: QS-12 based Single Axis Scan Set N-2071, with connector cables C-PWR-FL-36 and C-CMD-FL-36 and mounting block OFH-QS12-15 | Nutfield Technologies | The mounting block requires substantial modification for use in the iSIM (sawing in half and tapping threaded holes), since it is designed for two-axis scanning. An alternative should be considered |
| Power supply (±15 V) for Galvo scanning mirror control circuit board: 32212C (±15 V) | Calex | Power supply requires a safety cover, insulating putty or other protection from the open contacts where the electrical mains is connected |
| It was necessary to earth the COM output of the power supply to achieve symmetry of the supply voltages relative to ground | ||
| Lenslet arrays: APO-Q-P222-F1.86 ( | Advanced Microoptic Systems | 1 mm thick, comprising a square grid (pitch: 222 μm) of microlenses. Anti-reflection coated (400–650 nm) |
| APO-Q-P222-F0.93 ( | These were mounted in rotation, tip-tilt and translation mounts (Thorlabs: K6XS), and fixed to a further | |
| Pinhole array: 40 μm diameter pinholes, square array with pitch 222 μm, chrome on 0.090-inch-thick quartz) | Photosciences | Mounted in the same way as the lenslet arrays |
| Broadband AR coated back & front. (We also purchased an equivalent pinhole array with 50 μm diameter pinholes that might be more suitable for thin specimens) | ||
| Compensator plate: 0.025″ thick PW1-2025-UV | Melles Griot | |
| UK distributor: | ||
| Plane mirrors: 1″ diameter × 7, KM100-E02, 2″ diameter × 5, KM200-E02, Periscope assembly, RS99/M | Thorlabs | These parts include tip-tilt mounts for mounting |
| Brightfield source | Office Depot | A standard anglepoise desk lamp was used during construction |
| Analogue output card: PCI-6733, with breakout box BNC-2110 | National Instruments uk.ni.com/ | Electronic control and image acquisition (from a Python script – see Sections |
| This is different from the control arrangement of the previously published iSIM | ||
| Computer for control and acquisition: Intel Core i7-3820 CPU, 64 GB RAM, 64-bit Windows 7 | Stone | |
| Shelf unit PTA278 | Thorlabs | Overbench and underbench storage, with powerstrips |
Components used in iSIM alignment.
| Component | Manufacturer | Notes |
|---|---|---|
| Neutral density filters: | Thorlabs | Used to attenuate the beams for some stages of alignment (e.g. when focussing beams onto a sensor) |
| NE10A-A (OD 1.0) | ||
| NE30A-A (OD 3.0) | ||
| Shearing interferometer: SI254, with additional shear plates SI035P and SI100P | Thorlabs | For testing beam collimation |
| Collimated diode laser: CPS532, with power supply LDS5-EC | Thorlabs | Provides an auxiliary beam for testing system alignment. Mounted in pitch/yaw mount KAD11F (Thorlabs) |
| Optical power meter kit: PM130D | Thorlabs | Kit includes an optical power sensor |
| Image acquisition software: CamWare | PCO | This is installed with the PCO Edge 4.2 camera; useful for some alignment checks and procedures |
| Colour CMOS camera: DCC1645C | Thorlabs | Auxiliary camera, useful for detecting intermediate images in the optical system (with accompanying software) |
Fig. 2Use of an auxiliary laser beam, mirror and camera, together with the excitation beam, to align the foci of lenslet array 1 to the pinhole array. (See Fig. 1 for context.)
Fig. 3The graphical user interface for iSIM control. (A) Selection and control of the excitation laser and laser power, and number of scanning mirror sweeps per exposure; snapshot button. (B) z-Stack settings for the start and the end of the z-stack, and the step size required, together with the “Acquire Z-stack” acquisition button. (C) Timelapse settings controlling the number of images required, and the delay between the images, together with the “Acquire timelapse” acquisition button. Below this panel is a check box “Snap if stage moves” which can be checked when scanning the sample, for continuously previewing the sample.
Fig. 4iSIM images of cells co-stained for F-actin and microtubules. (A, B) Maximum intensity projections in x–y from 80 z-slices taken over a 4 μm depth (scale bars 5 μm) using a prepared slide (F-14781, Life Technologies), in which F-actin was stained with Texas Red-X phalloidin (red) and α-tubulin was stained with a primary mouse antibody to bovine α-tubulin and a secondary BODIPY FL goat anti-mouse antibody (green). (C, D) Raw (C) and deconvolved (D) iSIM images of a microtubule (scale bar 500 nm, pixel size 56 nm), from the region marked (yellow rectangle) in the combined image. The apparent microtubule width (FWHM in x–y) in this case was 188 nm (raw) and 138 nm (deconvolved), measured at the yellow line.
Performance of the iSIM system at Leeds (UoL), in comparison with the previously published iSIM.
| Original iSIM | UoL iSIM (NA = 1.49) | |||
|---|---|---|---|---|
| Raw | Deconvolved | Raw | Deconvolved | |
| Lateral FWHM (nm) | 213 ± 26 | 145 ± 14 | 212 ± 25 | 152 ± 13 |
| 216 ± 19 | 145 ± 9 | |||
| Axial FWHM (nm) | 511 ± 24 | 356 ± 37 | 513 ± 33 | 320 ± 16 |
Mean ± standard deviation.
Width obtained from slice of best focus at the microtubule.
Width obtained from maximum intensity projection of all slices onto the x–y plane.
Fig. 5Widefield, iSIM and deconvolved images captured using the iSIM at Leeds, with a 60× water immersion objective lens (Table 1). Resolution increases when changing from widefield fluorescence mode (no lenslet or pinhole arrays in place) to iSIM, and further following image deconvolution. Scale bar 10 μm. Sample is BPH1 cells stained for F-actin (red), and non-muscle myosin 2A (green).