Gert Jan Pelgrim1, Marco Das2, Ulrike Haberland3, Cees Slump4, Astri Handayani1, Sjoerd van Tuijl5, Marco Stijnen5, Ernst Klotz3, Matthijs Oudkerk6, Joachim E Wildberger2, Rozemarijn Vliegenthart1.
Abstract
OBJECTIVE: To test the feasibility of a CT-compatible, ex vivo, perfused porcine heart model for myocardial perfusion CT imaging.Entities:
Mesh:
Year: 2015 PMID: 26185756 PMCID: PMC4491382 DOI: 10.1155/2015/412716
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1From the venous reservoir (VR), the blood first passed the cardiac pump CP. Then, the blood was pumped through the blood filter (BF) into the oxygenator (OX) and, from there, through the aorta into the coronary arteries. The setup was placed on the scanner table; all parts which could possibly interfere with the signal were placed outside the field of view.
Arterial blood flow (L/min) and blood pressure (mm/Hg) in 5 different stenosis grades.
| Normal perfusion | FFR 0.70 | FFR 0.50 | FFR 0.30 | Total occlusion | |
|---|---|---|---|---|---|
| Arterial blood flow (SD) | 1.0 L/min | 1.0 L/min | 1.0 L/min | 0.93 L/min | 0.90 L/min |
|
| |||||
| Arterial blood pressure (SD) | 82.7 mm/Hg | 80.7 mm/Hg | 86.7 mm/Hg | 88.7 mm/Hg | 94 mm/Hg |
SD: standard deviation.
Diameter of stenosis divided by the artery diameter before stenosis, and the area stenosis divided by the area before stenosis, for the FFR-based stenosis grades.
| FFR 0.70 | FFR 0.50 | FFR 0.30 | |
|---|---|---|---|
| Mean diameter on CTA | 74% | 47% | 31% |
| Area stenosis on CTA | 49% | 22% | 12% |
Figure 2Example of images of the Cx artery with an FFR of 0.30. (a) The computed tomography angiography (CTA) image in curved multiplanar reformat shows the luminal narrowing. (b) The resulting perfusion defect in segment 5 as visible on maximum intensity projection (MIP) image.
Figure 3Image (a) shows the inflow of CM bolus through the inflow tube. Images (b)–(i) show the contrast inflow in the myocardium at short-axis cross-sections from basal to apical.
Figure 4Blood flow map of the porcine heart at complete Cx occlusion: red color meaning higher/normal flow and blue indicating reduced flow.
Median blood flow (L/min) in normal and defected segments for multiple stenosis grades.
| No stenosis | FFR 0.70 | FFR 0.50 | FFR 0.30 | Total occlusion | |
|---|---|---|---|---|---|
| Calculated normal | 151 mL/100 mL/min | 173 mL/100 mL/min | 162 mL/100 mL/min | 124 mL/100 mL/min | 108 mL/100 mL/min |
|
| |||||
| Calculated defect | 133 mL/100 mL/min | 157 mL/100 mL/min | 121 mL/100 mL/min | 74 mL/100 mL/min | 34 mL/100 mL/min |
Minimum and maximum blood flow for each group are shown in brackets.